Features
- All-in-One Compact Housing, Built-in Transmitter and Sensors
- Robust IP68 Water Submersible Protection, Directly Installed in the Field, No Cabinet Required
- Plug & Play, On- line Realtime Measurement
- Ultra Low Power Consumption, Ideal for Outdoor Applications
- 1500 N Tensible Strength Kevlar Reinforced Cable
- Temperature and Chloride Compensation
- Surge Protection for Power and RS485 Communication
- RS485 Digital Communication, Minimize Cabling and Engineering Cost
- Software Configuration, Calibration and Data Monitoring
- Standard Modbus RTU Protocol, Direct Connected with PLC, HMI
- Very Stable Electrodes, without the Need of Frequent Calibration
- Metal Membrane of Electrode to Prevent Scratch
- Electrode is over 12 Months Long Period of Life
- Auto Cleaning wiper, Less Maintenance
- Onboard Memory Allowing Users Easily Calibrate and Configure Sensor at Lab and Distribute to Various Fields Sites
- AQCFG Software Tool for Data Monitoring, Calibration , Configuration and Diagnosis
- IECEx/ATEX Ex ia IIB T5 Ga Explosive Proof Certification
Applications
Drinking water, water treatment, wastewater, fishery
Specifications
Measurement Method
The ion selective electrode continuous measurement of nitrate is carried out with specific membrane that is selective of ammoniumion. That membrane creates charge seperation between the water surrounded electrode and internal filling solution. The charge is measured as a voltage that changes the concentration in the water adjacent the membrane. Compared with the traditional colorimetric method, there are many advantages such as fast response time, more accurate, easier and no reagents used.
Installation
Submersible, flow through, pipe insertion
SMR23 Ammonium (NH4-N ) Analyzer
General |
|
Output Signal |
Output Signal: RS485 (Modbus RTU protocol), 19,200 bps, 8 data bits, no parity, 1 stop bit; 4~20 mA |
Data Resolution |
16 bits (0.001% FS) |
Surge Protection |
4000 VDC |
Power |
5~12 V DC, 20 mA (SMR23-1); 300 mA (SMR23-2) |
Protection |
Polarity, Overload, Short circuit |
Safety |
CE, FCC |
Turbidity |
|
Type |
Ion selective electrode |
Measurement range |
0.1~100/1,000 mg/L, auto range |
Accuracy |
±5% measured value ±0.5 mg/L |
Resolution |
0.01 mg/L |
Repeatability |
±3% measured value ±0.5 mg/L |
Operation pH |
pH 4~8. |
Operating pressure |
Max. 0.2 Kgf/cm² |
Operating Temperature |
0 ~ 40 °C |
Process Flow Rate |
0.1~4 m/s |
Response time |
2 min (90%) |
Protection |
IP68 |
Connection |
1”-11 PT, 5-pin M16 plug |
Housing |
POM |
Dimension |
Φ 47 x 279 mm |
Weight |
Nitrate analyzer: approx. 1.5 Kg ; cable: 80 g/m |
Potassium |
|
Measurement range |
0.1~1,000 mg/L |
Accuracy |
±5% measured valie ±0.5 mg/L |
Resolution |
0.01 mg/L |
Repeatability |
±3% measured valie ±0.5 mg/L |
Temperature |
|
Sensor |
Pt1000 |
Measurement range |
0~40 °C |
Accuracy |
±0.1 °C |
Resolution |
0.01 °C |
Repeatability |
0.1 °C |