H2O Water Quality Spectrometer
Frequently Asked Questions
-
Q1
Product Basics
A- Product Basics
Q1: What is the H2O analyzer?
The H2O Full-Spectrum Water Quality Analyzer uses UV-Vis ultraviolet and visible full-spectrum absorbance measurement to analyze multiple water quality parameters in water.
Q2: Which water quality parameters can H2O measure?
H2O can measure or calculate water quality parameters such as nitrate nitrogen, nitrite nitrogen, SAC254, COD, CODMn, TOC, BOD, DOC, turbidity, suspended solids (SS), and color.
Q3: What applications is H2O suitable for?
H2O is suitable for rivers, lakes, reservoirs, groundwater, drinking water, aquaculture, irrigation, industrial processes, wastewater treatment, and other applications.
Q4: Does H2O require reagents?
No. H2O uses the UV-Vis optical measurement principle and does not require sample pretreatment or chemical reagents.
Q5: Can H2O perform continuous online monitoring?
Yes. H2O can be installed directly in water or in a flow-through cell, and real-time water quality data can be continuously read via RS485 / Modbus RTU.
Q6: Does H2O have an automatic cleaning function?
Yes. H2O has a built-in automatic wiper that helps reduce the impact of biofouling or contamination on the optical lens. The wiper can also be manually activated in AQCFG.
Q7: Does H2O support RS485?
Yes. H2O uses RS485 communication and the Modbus RTU protocol.
Q8: Which AQUAS hosts can H2O connect to?
H2O can be used with AQUAS data loggers or controllers such as PRO, ECO, HMI, ARK, FLO, and ONE. The actual connection method should be confirmed based on the on-site host and wiring configuration.
Q9: Can H2O be connected to a computer for configuration?
Yes. H2O can be connected to a computer through CAB06 / RS485, and AQCFG software can be used for device search, configuration, calibration, and data reading.
Q10: Can H2O be used with AQWEB?
Yes. H2O can be integrated into the AQWEB / AQOPC platform through a data logger or controller for data collection, remote display, alarms, trend charts, and data analysis.
- Product Basics
-
Q2
Measurement Principle and Technical Features
A- Measurement Principle and Technical Features
Q1: What is the measurement principle of H2O?
H2O uses UV-Vis ultraviolet and visible full-spectrum absorbance measurement technology to analyze multiple water quality parameters based on the absorption characteristics of water samples at different wavelengths.
Q2: What is the spectral range of H2O?
The spectral diagram shows that H2O analyzes samples across an ultraviolet and visible light range of approximately 200 to 800 nm.
Q3: Why can H2O measure multiple parameters at the same time?
Different pollutants or water quality components have different absorption characteristics at different wavelengths. H2O uses full-spectrum data and algorithms to simultaneously analyze multiple water quality parameters.
Q4: Why can H2O measure COD?
Some organic compounds in water absorb ultraviolet light. H2O uses full-spectrum absorbance signals and calibration models to calculate parameters such as COD or CODMn.
Q5: Can H2O measure nitrate nitrogen and nitrite nitrogen?
Yes. H2O can measure NO3-N and NO2-N related parameters using ultraviolet absorbance characteristics.
Q6: Are H2O measurement results affected by turbidity?
Yes. Turbidity, suspended solids, color, and other interfering substances may affect spectral absorbance results. Therefore, H2O provides compensation settings and sample water calibration functions.
Q7: What is the purpose of the H2O compensation function?
The compensation function can apply multiplier, ratio, or offset compensation to parameters such as TSS, NO3, and NO2, helping reduce differences between field sample readings and laboratory analysis values.
Q8: Is H2O suitable for reagent-free water quality monitoring?
Yes. A key advantage of H2O is that it requires no reagents and no sample pretreatment, reducing maintenance and consumable costs.
- Measurement Principle and Technical Features
-
Q3
Specifications and Measurement Ranges
A- Specifications and Measurement Ranges
Q1: What optical path length options are available for H2O?
H2O is available with 1 mm, 5 mm, or 20 mm optical path lengths depending on the model.
Q2: Does the optical path length affect the measurement range?
Yes. Different optical path lengths correspond to different concentration ranges. A shorter path length is generally suitable for higher concentration samples, while a longer path length is suitable for lower concentration samples.
Q3: What is the COD measurement range of H2O?
Depending on the optical path length, the CODeq measurement range may include 0 to 200 mg/L, 0 to 800 mg/L, or 0 to 2,500 mg/L.
Q4: What is the CODMn measurement range of H2O?
Depending on the optical path length, the CODMn measurement range may include 0 to 50 mg/L, 0 to 200 mg/L, or 0 to 500 mg/L.
Q5: What is the BOD measurement range of H2O?
Depending on the optical path length, the BODeq measurement range may include 0 to 50 mg/L, 0 to 200 mg/L, or 0 to 1,000 mg/L.
Q6: What is the DOC measurement range of H2O?
Depending on the optical path length, the DOCeq measurement range may include 0 to 50 mg/L, 0 to 200 mg/L, or 0 to 1,000 mg/L.
Q7: What is the turbidity measurement range of H2O?
Depending on the optical path length, the turbidity measurement range may include 0 to 100 NTU, 0 to 500 NTU, or 0 to 2,500 NTU.
Q8: What is the SS measurement range of H2O?
Depending on the optical path length, the suspended solids (SS) measurement range may include 0 to 200 mg/L, 0 to 1,000 mg/L, or 0 to 5,000 mg/L.
Q9: What is the color measurement range of H2O?
The color measurement range may include 0 to 3,500 Hazen or 0 to 15,000 Hazen. The actual range should be confirmed based on the optical path length specification.
Q10: What is the resolution of H2O?
Most water quality parameters have a resolution of 0.01, and the color resolution is 0.1.
Q11: What is the protection rating of H2O?
H2O has an IP68 protection rating and can be used in submerged applications.
Q12: What is the maximum transmission distance of H2O?
The RS485 communication transmission distance can reach approximately 1.2 km. The actual distance may be affected by cable type, site conditions, and interference.
- Specifications and Measurement Ranges
-
Q4
Installation
A- Installation
Q1: What installation methods are available for H2O?
H2O can be installed vertically, horizontally, at the water surface, or in a flow-through cell.
Q2: Can H2O be directly immersed in water?
Yes. H2O is available in an immersion version and can be installed directly in water for in-situ monitoring.
Q3: Can H2O be installed in a flow-through cell?
Yes. H2O supports flow-through installation and can be used with piping or a flow-through cell for sample measurement.
Q4: Should wiring layout be considered during H2O installation?
Yes. Avoid bundling the sensor signal cable with the power cable to reduce interference and ensure signal stability.
Q5: Does H2O require an external power supply?
Yes. H2O requires an external +12 VDC power supply for normal operation.
Q6: Can H2O be powered only by USB?
No. USB power is not recommended and is not sufficient for normal operation. H2O requires an external +12 VDC power supply; USB / CAB06 is mainly used for communication and configuration.
Q7: Why should the lens position be considered during H2O installation?
H2O is an optical measurement device. If the optical lens is blocked by sludge, bubbles, biofilm, or debris, readings may become unstable or inaccurate.
Q8: What type of flow environment is suitable for H2O installation?
It is recommended to install H2O in a location with representative water samples, stable flow conditions, and a lower likelihood of sludge accumulation or bubble attachment.
- Installation
-
Q5
Communication and Wiring
A- Communication and Wiring
Q1: What communication interface does H2O use?
H2O uses an RS485 digital communication interface.
Q2: What communication protocol does H2O use?
H2O uses the Modbus RTU communication protocol.
Q3: What are the default communication settings for H2O?
The screen and settings description indicate that the commonly used communication settings are 19200 bps, None Parity, 8 Data Bits, and 1 Stop Bit.
Q4: Which baud rates does H2O support?
H2O supports 19200 and 9600 bps.
Q5: Which communication formats does H2O support?
H2O supports communication data formats such as N81, O81, and E81.
Q6: Can the H2O ID be duplicated?
No. The ID must not be duplicated, otherwise communication conflicts or scan failures may occur.
Q7: What is the H2O connector pin assignment?
The H2O 5-pin connector pin assignment is as follows:
PIN 1 Brown: GND
PIN 2 White: RS485-
PIN 3 Blue: RS485+
PIN 4 Black: +12V DC
PIN 5 Gray: Not used
Q8: Can H2O be connected via RS232?
Yes, but an RS232 / RS485 converter is required, and a separate 12 VDC power supply must be provided.
Q9: Can H2O be directly connected to a PLC RS485 port?
Yes. An RS485 port wiring diagram is provided, but 12 VDC power must still be supplied, and D+, D-, GND, and communication parameters must be confirmed.
Q10: What should be noted when connecting H2O to PRO / ECO / HMI / ARK?
H2O can be connected to devices such as PRO, ECO, HMI, and ARK via Port 1 RS485, but an external 12 VDC power supply is still required.
- Communication and Wiring
-
Q6
AQCFG Software and Connection
A- AQCFG Software and Connection
Q1: Which software is used to configure H2O?
H2O uses AQUAS AQCFG software for device search, connection, basic settings, calibration, wiper control, firmware updates, and data export.
Q2: How is AQCFG installed?
After inserting the CD or obtaining the installer, select the AQCFG folder, click the Setup icon, follow the Next steps, and click Install to complete installation.
Q3: What must be installed before connecting H2O?
The USB driver must be installed first by running CDM20824_Setup. After installation, a new Serial COM Port will appear in Windows Device Manager.
Q4: How is H2O connected to a computer?
Use a PC / notebook and connect to the M12 5-pin connector at the top of H2O through the CAB06-B RS485 cable, while providing an external 12 VDC power supply.
Q5: How does AQCFG search for H2O?
After opening AQCFG, click Setup to select the COM Port, baud rate, and ID range. Close the setup page and click Scan to search for the device.
Q6: How do I enter the settings after H2O is scanned?
Select the device to be configured and click Connect, or double-click the device to enter the basic settings window.
Q7: What do the green and red communication status indicators in AQCFG mean?
Green indicates connected; red indicates disconnected.
Q8: Can AQCFG switch languages?
Yes. The language can be selected in the settings, and AQCFG must be restarted after the change.
Q9: Can AQCFG set decimal places?
Yes. Data Format can be used to select the number of decimal places displayed for values.
Q10: Can AQCFG set a login password?
Yes. The Password function can be used to set the AQCFG login password.
- AQCFG Software and Connection
-
Q7
Basic Settings and Real-Time Readings
A- Basic Settings and Real-Time Readings
Q1: What information is shown on the H2O basic settings page?
The basic settings page displays the name, model, serial number, location, area, sub-area, GPS longitude and latitude, altitude, and real-time values for each water quality parameter.
Q2: Which real-time readings can H2O display?
The screen can display real-time parameters such as COD, BOD, CODMn, NO3, NO2, NH4-N, TN, TP, TSS, and temperature. The actual displayed items depend on the model and settings.
Q3: Where are location, area, and sub-area set?
Location, area, and sub-area are usually set in the data logger to facilitate AQWEB management and monitoring station identification.
Q4: Where do GPS longitude, latitude, and altitude come from?
Longitude, latitude, and altitude are GPS location information from the data logger. If the system does not provide GPS, they may need to be entered manually or may not display valid coordinates.
Q5: What does the wiper button on the basic page do?
After clicking Start, the wiper will clean the H2O optical lens once.
Q6: What should be done after changing H2O communication parameters?
After changing communication parameters, click Confirm to save the settings. If the COM Port or communication parameters are changed, monitoring and related settings must be restarted.
Q7: Can the pipe diameter / optical path parameter be set arbitrarily?
No. The pipe diameter represents the actual optical path parameter and must correspond to the physical optical path length, such as 1 mm, 5 mm, or 20 mm. Incorrect settings will affect measurement results.
- Basic Settings and Real-Time Readings
-
Q8
Calibration
A- Calibration
Q1: When does H2O require calibration?
Calibration can be performed when readings drift, when field sample readings differ significantly from laboratory analysis values, after maintenance to verify accuracy, or during periodic maintenance.
Q2: What steps are included in H2O calibration?
H2O calibration includes DI water calibration, sample water calibration, absorbance confirmation, and compensation settings.
Q3: How do I enter the H2O calibration page?
On the AQCFG advanced settings page, click the calibration parameter and then click Start to enter the H2O calibration page.
Q4: How is DI water calibration performed?
Place H2O in DI water with conductivity below 1 µS/cm and almost no chemical substances. Click DI Water to perform DI water spectrum calibration. The calibration is complete when the completion window appears.
Q5: What DI water conditions are required for DI water calibration?
The DI water conductivity must be below 1 µS/cm and contain almost no chemical substances.
Q6: What should be prepared before sample water calibration?
Laboratory analysis of the sample water must be completed first to obtain actual data values such as COD, NO3, NO2, and TSS.
Q7: How is sample water calibration performed?
Place H2O in the sample water and click Sample to perform sample water spectrum calibration. Add and save the sample water data, and the system will take a sample and complete the sample water spectrum data.
Q8: How many sample water records can be created for calibration?
Up to 5 sample water records can be added to the sample table.
Q9: What is the absorbance function?
The absorbance function displays the absorbance values of DI water and sample water on the spectrum. A single sample record or all sample records can also be selected to view spectral absorbance.
Q10: When should the compensation function be used?
When TSS, NO3, NO2, or other concentration parameters require correction, the compensation function can be used to set multiplier, ratio, or offset adjustments.
Q11: Is compensation required again after sample data is modified?
Yes. Compensation is calculated based on absorbance data. If sample data is added, modified, or deleted, absorbance must be run again before performing the compensation procedure.
Q12: Can the calibration procedure be canceled?
Yes. Click Cancel to exit the calibration procedure and return to the main screen.
Q13: Can H2O restore factory calibration values?
Yes. On the advanced calibration page, click Restore Factory Values to restore the factory calibration values.
- Calibration
-
Q9
Correction Values and Data Export
A- Correction Values and Data Export
Q1: What are correction value settings?
Correction value settings list the compensation relationship for each concentration parameter. The default slope multiplier is 1 and the zero offset is 0. Users can fine-tune individual parameters.
Q2: Which items can be adjusted by correction values?
Parameters such as COD, BOD, CODMn, NO3, NO2, NH4-N, TN, TP, Suspended Solids, and Temperature can be adjusted, depending on the software screen and model.
Q3: What is the correction value calculation formula?
Formula: Target value = (Actual value - Zero value) / Slope value.
Q4: How should SS be corrected if the reading is 0 but the laboratory value is 9 mg/L?
If the slope value is 1, the target value 9 = [0 - zero value] / 1. Therefore, the zero value should be entered as -9, and correction should then be enabled.
Q5: What should be done after setting correction values?
After entering the correction value, click Start. The system will apply compensation adjustment to that parameter. After completion, verify the result with actual sample measurement.
Q6: What is the purpose of the data export function?
The data export function is mainly used for maintenance support. It exports the current sample water database data for subsequent sample parameter analysis.
Q7: What files are generated during data export?
Exported data includes TXT files such as DIWater, Sample, AU, Compensate, and Source.c.
Q8: What should be done before exporting data?
The system will prompt the user to place H2O in the sample water and enter concentration data for related parameters such as COD, NO3, and TSS. After confirmation, save the data and perform spectrum analysis.
- Correction Values and Data Export
-
Q10
Modbus Readings
A- Modbus Readings
Q1: Which Modbus Function Codes does H2O support?
The Modbus table shows that 3xxxx read-only registers support Function 4, and Function 0x03 is also used to read water quality parameters.
Q2: How can Channel 1 to Channel 15 be read at once?
Read Reference 38193, Address 0x2000. The data length is 30 words, corresponding to Float data for Channel 1 to Channel 15.
Q3: How are Channel units read?
Read Reference 38194, Address 0x2001. The data length is 8 words, corresponding to units for Channel 1 to Channel 15.
Q4: How is the calibration date and time read?
Read Reference 34097, Address 0x1000. The data length is 3 words, in YYMMDDHHmmSS format.
Q5: How is the model read?
Read Reference 34098, Address 0x1001. The data length is 20 words, in Ansi-C String format.
Q6: How is the firmware version read?
Read Reference 34099, Address 0x1002. The data length is 2 words.
Q7: What is the Modbus address for SAC254?
SAC254 corresponds to Function 0x03, Reference 81 / 82, Address 0x50 / 0x51, with data type FLOAT32 H / L.
Q8: What is the Modbus address for COD?
COD corresponds to Function 0x03, Reference 83 / 84, Address 0x52 / 0x53, with data type FLOAT32 H / L.
Q9: What is the Modbus address for TOC?
TOC corresponds to Function 0x03, Reference 85 / 86, Address 0x54 / 0x55, with data type FLOAT32 H / L.
Q10: What is the Modbus address for BOD?
BOD corresponds to Function 0x03, Reference 87 / 88, Address 0x56 / 0x57, with data type FLOAT32 H / L.
Q11: What is the Modbus address for DOC?
DOC corresponds to Function 0x03, Reference 89 / 90, Address 0x58 / 0x59, with data type FLOAT32 H / L.
Q12: What is the Modbus address for CODMn?
CODMn corresponds to Function 0x03, Reference 91 / 92, Address 0x5A / 0x5B, with data type FLOAT32 H / L.
Q13: What is the Modbus address for NO3?
NO3 corresponds to Function 0x03, Reference 93 / 94, Address 0x5C / 0x5D, with data type FLOAT32 H / L.
Q14: What is the Modbus address for NO2?
NO2 corresponds to Function 0x03, Reference 95 / 96, Address 0x5E / 0x5F, with data type FLOAT32 H / L.
- Modbus Readings
-
Q11
Firmware Update
A- Firmware Update
Q1: Can H2O firmware be updated?
Yes. The AQCFG advanced page provides firmware update functions. Load the correct .aqs firmware file to perform the update.
Q2: How can the H2O firmware version be checked?
The current firmware version and version date can be checked on the AQCFG advanced page.
Q3: What are the H2O firmware update steps?
Click Open to select the correct .aqs file, then click Download to start the update. When the AQUAS INC DISCLAIMS window appears, select YES to begin installation.
Q4: What should be noted before updating firmware?
Before updating, make sure only one device is connected to avoid upload failure or an unrecoverable device condition.
Q5: Can the firmware update be interrupted?
No. Once the update starts, it must not be interrupted. Wait until the update is complete.
- Firmware Update
-
Q12
Maintenance
A- Maintenance
Q1: Does H2O require regular maintenance?
Yes. As an optical water quality analyzer, H2O requires regular cleaning of the sensor and optical glass, as well as periodic calibration to maintain measurement accuracy.
Q2: How often should H2O be cleaned?
Cleaning is recommended approximately once per week. The actual frequency can be adjusted according to water quality conditions.
Q3: What should be noted when cleaning the H2O optical glass?
Avoid scratching the glass surface, as scratches or contamination can affect optical measurement results.
Q4: How often does H2O require calibration?
Calibration is recommended approximately every 6 to 12 months. The actual interval can be adjusted according to water quality, reading drift, and on-site maintenance requirements.
Q5: How often should H2O be inspected?
Inspection is recommended approximately every 3 months.
Q6: Which cleaning agents can be used to clean the front housing?
Isopropanol, dilute acids up to 3%, dilute alkalis up to 5%, esters, hydrocarbons, ketones, and household cleaners may be used.
Q7: Which cleaning methods should not be used?
Do not use concentrated inorganic acids or alkalis, benzyl alcohol, dichloromethane, or high-pressure steam, as these may damage the optical surface.
Q8: Can the wiper be manually activated?
Yes. In AQCFG, click the wiper function to manually perform one wiper cleaning cycle.
- Maintenance
-
Q13
Troubleshooting
A- Troubleshooting
Q1: What should I do if the system cannot read H2O values during connection?
Check whether the system connection settings, including ID, baud rate, and data format, match H2O, and use AQCFG to confirm the H2O parameters in advance.
Q2: Should the ID be checked if values cannot be read?
Yes. Confirm whether H2O is using a duplicate ID with another device in the system, as duplicate IDs can cause communication abnormalities.
Q3: Should the power supply be checked if values cannot be read?
Yes. H2O requires an external +12 VDC power supply for normal operation.
Q4: Should connectors be checked if values cannot be read?
Yes. Make sure all connectors are securely connected, including RS485 and power wiring.
Q5: What should I do if H2O readings are unstable?
First clean the optical glass and check whether bubbles are present on the glass surface.
Q6: What should I do if H2O readings do not change?
Confirm that the H2O connectors are securely connected, and check whether communication and power supply are normal.
Q7: What should I do if H2O measurement values are inaccurate?
First clean the sensor glass surface, then recalibrate.
Q8: Do bubbles affect H2O readings?
Yes. Bubbles on the glass surface may cause unstable readings or abnormal measurements.
Q9: What problems can a dirty lens cause?
A dirty lens can affect spectral absorbance measurement, resulting in reading deviation, unstable readings, or inaccurate measurement results.
Q10: What are the most common reasons H2O cannot operate normally?
Common causes include no 12 VDC power supply, incorrect RS485 wiring, duplicate ID, mismatched communication parameters, dirty optical glass, or bubbles on the optical surface.
- Troubleshooting