PWS Micro Automatic Sampler

Frequently Asked Questions

  • A
    1. Product Basics
      Q1: About PWS?
      The PWS is an AQUAS automatic water-quality sampler. It can be used for automatic sampling in water-quality, wastewater, and surface-water applications, and can be integrated with sensors, data logging, communications, and cloud platforms.

      Q2: What is the primary purpose of the PWS?
      The PWS automatically collects water samples based on time, alarms, or sensor-value conditions, supporting laboratory analysis, exceedance sampling, event-based sampling, and on-site water-quality monitoring.

      Q3: Where is the PWS suitable for use?
      The PWS is suitable for water-quality sampling in sewer systems, stormwater runoff, municipal and industrial wastewater, rivers, streams, irrigation channels, and similar applications.

      Q4: Which sensors can the PWS be used with?
      The PWS can be used with MAX multiparameter water-quality analyzers and various SMR digital sensors, such as level, flow, rainfall, or water-quality sensors, to trigger sampling or log data according to site requirements.

      Q5: Does the PWS support alarm-triggered sampling?
      Yes. The PWS can trigger water sampling based on sensor or I/O alarm conditions and starts sampling immediately when an alarm occurs.

      Q6: Does the PWS support scheduled sampling?
      Yes. The PWS can be configured for fixed-interval sampling, such as every 2, 5, 10, or 30 minutes, or every 1, 2, 6, 12, or 24 hours.

      Q7: Does the PWS support manual sampling?
      Yes. In AQCFG, the PWS can be operated in manual mode. Sampling starts immediately after clicking Start.

      Q8: Is the PWS a single-bottle sampler only?
      This model is a single-bottle sampler. After alarm sampling, the sample bottle will contain collected water. The operator must confirm that the bottle has been replaced and press the reset button before the next alarm-triggered sampling cycle can begin.

      Q9: Can the PWS be used with a cloud platform?
      Yes. The PWS can be integrated with the AQWEB cloud platform for data upload, historical data query, trend charts, alarms, and remote monitoring.

      Q10: Does the PWS support 4G / GPRS communication?
      Yes. Depending on the model, the PWS can use GPRS, 3G, 4G, Wi-Fi, Ethernet, and other communication methods to transmit data back to a server.

      Q11: Does the PWS support FTP upload?
      Yes. The PWS can be configured in FTP mode to upload data to an FTP server. CCS / FTP mode can also be selected for simultaneous upload.

      Q12: Does the PWS support SMS alarms?
      Yes. The PWS can store up to 10 mobile phone numbers and can send SMS notifications when an alarm starts or ends.

      Q13: Does the PWS support GPS?
      GPS can be supported. Coordinates can be entered manually, or an optional GPS module can be used to search for and record coordinates.

      Q14: Does the PWS support Bluetooth connection?
      Bluetooth is optional. If the Bluetooth module is installed, the computer can manage Bluetooth devices and connect to the PWS by entering the password.

  • B
    1. Hardware and Accessories
      Q1: What are the common PWS accessories?
      Common PWS accessories include CAB06, CAB12, CAB13, CAB01-A, PAN03 solar panel, BAT11-2 rechargeable lithium battery, digital sensors, notebook computer, inlet tubing, and outlet tubing.

      Q2: What is CAB06 used for?
      CAB06 is used to connect the PWS to a computer for setup and connection through USB / RS485 and AQCFG software.

      Q3: What is CAB12 used for?
      CAB12 is used to connect digital sensors, such as SMR sensors.

      Q4: What is CAB13 used for?
      CAB13 is used to connect external power equipment, such as the BAT11-2 rechargeable lithium battery or PAN03 solar panel.

      Q5: What is PAN03?
      PAN03 is a solar panel that can be used with the PWS for outdoor or off-grid applications.

      Q6: What is BAT11-2?
      BAT11-2 is a rechargeable lithium battery that can be used as one of the power sources for the PWS.

      Q7: What tubing ports does the PWS have?
      The PWS has an inlet and an outlet. The outlet is connected to the sample bottle to deliver the collected water sample into the bottle.

      Q8: Can the PWS be connected to an external power source?
      Yes. Depending on the configuration, the PWS can use DC power, batteries, or solar-related power modules.

      Q9: What should be considered when using batteries with the PWS?
      Select the correct battery type and actual power-supply mode. If a primary battery is replaced, press the battery reset or zero button so that the system can correctly calculate the remaining battery capacity.

      Q10: What are the risks of using an incorrect battery?
      The manual warns that using an incorrect battery type or a non-original charger may create an explosion risk. Used batteries should be handled according to the manufacturer's instructions.

  • C
    1. Sampling Functions
      Q1: Does the length of the PWS sampling tube affect sampling time?
      Yes. According to the manual, the sampling speed is approximately 3.7 seconds per 1 m of sampling tube. If the tube length differs, the required sampling time must be recalculated.

      Q2: How long does it take to fill a 1000 cc sample bottle?
      The manual states that a 1000 cc sample bottle requires approximately 240 seconds to fill.

      Q3: What is the approximate PWS drain time?
      The manual states that the drain time is approximately 20 seconds, allowing water in the internal drain tubing to be discharged.

      Q4: What conditions are the sampling times in the manual based on?
      The manual states that 240 seconds to fill 1000 cc and 20 seconds for drainage are based on the default 3.5 m sampling tube. If the tube length differs, the time must be recalculated.

      Q5: What sampling modes does the PWS provide?
      The PWS provides manual mode, timed mode, and alarm mode.

      Q6: How is sampling performed in manual mode?
      In the AQCFG sampler settings, select manual mode and click Start to begin sampling immediately.

      Q7: How is sampling performed in timed mode?
      In timed mode, set the sampling interval. The PWS will automatically sample at the specified time interval.

      Q8: Which intervals can be selected for timed sampling?
      Available intervals include 2, 5, 10, and 30 minutes, as well as 1, 2, 6, 12, and 24 hours.

      Q9: How does alarm-mode sampling work?
      Alarm mode starts sampling immediately when a sensor or I/O alarm occurs according to the configured conditions.

      Q10: Why must the reset button be pressed after alarm sampling?
      Because the PWS is a single-bottle sampler, the sample bottle contains water after alarm sampling. After replacing the bottle, the reset button must be pressed before the system allows the next alarm-triggered sampling cycle.

      Q11: What should be manually checked after sampling?
      Confirm that the sample bottle has been replaced, the tubing is normal, the sample water has entered the bottle, and the reset operation has been completed.

      Q12: What is the purpose of the drain function?
      The drain function discharges water remaining in the internal drain tubing, reducing the risk that residual water will affect the next sample.

  • D
    1. AQCFG Software Installation and Connection
      Q1: Which software is used to configure the PWS?
      The PWS uses AQUAS AQCFG software for device connection, system setup, sampling setup, sensor setup, communication setup, data import/export, and troubleshooting.

      Q2: How is AQCFG installed?
      Insert the CD, select the AQCFG folder, click the Setup icon, follow the on-screen Next steps, and finally click Install to complete the installation.

      Q3: What must be installed before connecting the PWS with CAB06?
      Install the USB Driver first. Open the USB Driver folder and run CDM20824_Setup. After installation, connect CAB06; USB Serial Port should appear in Windows Device Manager.

      Q4: Which port should the PWS be connected to for configuration?
      Use CAB06 to connect the notebook computer to Port 0 of the PWS for configuration.

      Q5: Can the PWS be connected via Bluetooth?
      Yes, but the Bluetooth module must be selected. When managing Bluetooth devices on the PC, enter password 0000 to connect to the PWS.

      Q6: What should be configured before searching for the PWS in AQCFG?
      Click Setup and configure the COM Port, baud rate, and ID range to be scanned.

      Q7: What is the default baud rate for PWS connection with AQCFG?
      The commonly used default baud rate for PWS connection with AQCFG is 115200 bps.

      Q8: What is the common AQCFG connection format?
      The common format is 115200 bps, N, 8, 1, meaning no parity, 8 data bits, and 1 stop bit.

      Q9: How do you enter the setup page after the PWS is found?
      Select PWS RTU and press LINK. The system will open the setup page.

      Q10: Is a password required to enter the PWS setup page?
      Yes. The manual screen shows that a password is required. The displayed default password is 0000, while the text also mentions an initial password of 1234. For the actual shipment version, follow the device screen or AQUAS setup data.

      Q11: What should be checked if AQCFG cannot find the PWS?
      Check that the USB Driver is installed, CAB06 is connected to Port 0, the COM Port is correct, the baud rate is correct, the ID range includes the device ID, and the device is powered.

      Q12: What should be done first after AQCFG connects?
      Before entering host system settings, click the System On button so that the PWS system is enabled before configuration.

  • E
    1. Host System Settings
      Q1: Why must System On be pressed before configuring the PWS?
      After System On is pressed, the PWS enables system functions and allows host settings, data logging, sampling, and communication-related settings to be configured.

      Q2: What information can be viewed on the system page?
      The system page shows model, serial number, location, area, sub-area, system time, battery information, communication speed, and Modbus ID.

      Q3: Can the model field be modified?
      No. The model is created during production to indicate the host hardware configuration and cannot be modified by the user.

      Q4: What is the serial number?
      The serial number is the unique device ID information for each PWS and is used for equipment identification and management.

      Q5: What is the location field used for?
      The location field can be customized by the user and may contain up to 20 letters, making it easier to manage and identify equipment in AQWEB.

      Q6: What are the area and sub-area fields used for?
      The area and sub-area fields can be customized by the user to help classify and manage different monitoring stations or installation locations on the platform.

      Q7: How is the PWS system time synchronized?
      Select no synchronization, PC / Server synchronization, or GPS synchronization. If PC / Server is selected, set the time zone and press Confirm.

      Q8: Why must the battery type be set correctly?
      Battery type affects how the system interprets battery voltage and capacity. Incorrect settings may lead to inaccurate battery capacity estimates.

      Q9: What should be done after replacing a primary battery?
      After replacing a primary battery, press Zero or Reset so that the system recalculates the battery capacity.

      Q10: What is the default COM connection speed for the PWS?
      The host system settings in the manual state that the default connection speed through the COM Port is 115200 bps.

      Q11: Can the Modbus ID of the PWS be changed?
      Yes. The system page can display and set the Modbus ID currently available for scanning and connection.

  • F
    1. 4G / GPRS / FTP Communication
      Q1: Which data return methods does the PWS support?
      The PWS can return configuration parameters, real-time values, alarms, and logged data to a server through GPRS, 3G, 4G, Wi-Fi, or Ethernet modules.

      Q2: What should be done before enabling 4G / GPRS?
      Before configuration, select Enable and then press Confirm.

      Q3: How is communication signal strength evaluated?
      AQCFG displays signal strength from 0% to 100% and a dBm value. If the dBm value is greater than -93 dBm, the signal is too weak and may not connect to the server.

      Q4: What does 4G / GPRS status Standby mean?
      Standby means the module is off or in standby mode.

      Q5: What does Try Connecting mean?
      Standby -> Try Connecting means the device is establishing a connection.

      Q6: What does Connected -> Data Communication mean?
      It means the connection is successful and data is being transmitted.

      Q7: How should the APN be configured?
      The APN, username, and password should be configured according to information provided by the telecom operator. They can also be obtained by selecting the country and operator from the network provider menu on the right side of the software.

      Q8: What must be configured in CCS mode?
      CCS mode requires the server IP and server port. The manual lists the default server port as 502.

      Q9: What is the AQWEB IP address?
      In CCS mode, the manual lists the AQWEB IP as 118.163.120.141. The test screen also shows test server IP 118.163.120.143 / Port 502. For actual projects, follow the shipment or platform configuration data.

      Q10: Can the PWS be configured with a backup server?
      Yes. If PWS data must be sent to a second server at the same time, enable the backup server and enter the second server IP and Port.

      Q11: What information is required for FTP mode?
      FTP mode requires the FTP server IP, Port, username, and password. The default FTP Port is 21.

      Q12: What is CCS / FTP mode?
      Select CCS / FTP mode when both CCS and FTP servers need to receive PWS uploaded data.

      Q13: How do you confirm communication is normal after setup?
      Click Connection Test to confirm that data can be uploaded to the server successfully.

      Q14: What should be checked first if 4G cannot connect?
      Check that the SIM card is activated, the APN is correct, the signal is sufficient, the server IP and Port are correct, the module is enabled, and no firewall or network restriction is blocking communication.

  • G
    1. AQWEB Cloud Platform
      Q1: Can the PWS upload data to AQWEB?
      Yes. The PWS can upload data through AQWEB Server to test and confirm that the data return is normal.

      Q2: How do you log in to the AQWEB test server?
      The manual illustrates opening a web browser and entering the AQWEB server address. The test account and password must be obtained from the sales representative.

      Q3: What should be verified in AQWEB?
      Use the tree view on the left to search for the test device and confirm that returned data, system information, operation functions, historical data, and trend charts are correct and complete.

      Q4: What should be checked in AQWEB system information?
      Check whether device model, serial number, location, area, system time, battery status, firmware version, and other information are correct.

      Q5: What should be checked in AQWEB operation functions?
      Confirm whether operation schedules, alarm control, logging intervals, communication settings, and data upload intervals match site requirements.

      Q6: Where can AQWEB historical data be viewed?
      Historical data can be viewed in Data Report or Data Table to confirm whether sensing data has been fully uploaded.

      Q7: Where can AQWEB trend charts be viewed?
      Trend charts can be viewed in Statistics Analysis or Analysis to confirm that uploaded data is correct and continuous.

      Q8: Which document should be used for detailed AQWEB operation?
      The manual notes that detailed AQWEB Server information should be referenced in the relevant AQWEB user manual.

  • H
    1. GPS and SMS
      Q1: What can PWS GPS do?
      GPS can obtain device coordinates and store them in the record, allowing the platform to display location and track the device.

      Q2: Can the location be configured without a GPS module?
      Yes. If no GPS module is selected, coordinates can be entered manually.

      Q3: What does GPS status off mean?
      off means the GPS module is turned off.

      Q4: What does GPS status Searching mean?
      Searching means the GPS module is enabled and searching for satellite positioning.

      Q5: What does GPS status Position mean?
      Position means positioning is complete.

      Q6: How is the GPS signal displayed?
      AQCFG displays GPS module signal strength from 0% to 100%.

      Q7: Can GPS be logged on a schedule?
      Yes. After selecting Timed, GPS coordinates will be logged together with other I/O data.

      Q8: When is GPS continuous mode suitable?
      Continuous mode keeps the GPS module enabled. The manual recommends using it when external power is available.

      Q9: What is GPS search mode?
      Search mode manually enables the GPS function. After one search is completed, the GPS function automatically turns off whether positioning succeeds or not.

      Q10: What is the shortest interval for GPS timed mode?
      The manual states that the shortest interval for GPS timed activation mode is 5 minutes.

      Q11: How many SMS phone numbers can the PWS set?
      Up to 10 mobile phone numbers can be configured.

      Q12: When can SMS messages be sent?
      Each mobile number can be configured to receive SMS messages when an alarm starts and when an alarm ends.

      Q13: Why can't alarm-end SMS be selected directly?
      The manual notes that Alarm Start must be selected first before Alarm End can be selected.

      Q14: How is the SMS function tested?
      Press SMS Test. The host will send a test SMS message to the first mobile phone number.

  • I
    1. RS485 Sensor Connection
      Q1: What can RS485 Port 1 on the PWS do?
      Port 1 can connect SMR digital sensors and scan and configure them through AQCFG.

      Q2: How are SMR sensors scanned?
      Connect the SMR sensor to PWS Port 1. After setting the baud rate and data format, press Confirm, then click Scan to search for connected SMR sensors.

      Q3: What are the commonly used settings when scanning Port 1 on the PWS?
      The manual mentions 115200 bps, N, 8, 1 for scanning SMR sensors from the PWS host. The level-meter example states that Port 1 default is 19200.N81. Actual settings must be adjusted according to the SMR sensor configuration.

      Q4: What are the SMR sensor ID restrictions?
      Each SMR ID must be unique and within the range 1 to 8.

      Q5: What happens if an SMR ID is outside the range 1 to 8?
      Sensors outside this range will not be read by the PWS.

      Q6: How do you enter settings after an SMR sensor is found?
      Click the device in the tree list, or quickly double-click the SMR sensor in the list to open the sensor settings page.

      Q7: Can the PWS connect to an SMR02 level sensor?
      Yes. The manual uses the SMR02 level sensor as an example to show how to connect it to PWS Port 1 and configure it.

      Q8: What can be configured after connecting an SMR02?
      Settings include measured liquid density, water-level height unit, logging method, water-level compensation, and zero calibration.

      Q9: How is SMR02 level compensation performed?
      Select Enable and press Confirm. Remove the level meter from the water and place it in air. Select Pressure-1 as the compensation source and click Zero Calibration. When the status field shows OK, the process is complete.

      Q10: How are SMR02 readings logged?
      Timed or alarm logging can be selected so that level data is recorded according to the configured interval or alarm conditions.

  • J
    1. I/O, Alarms, and P2P
      Q1: What can the PWS I/O module do?
      The I/O module can be used for sampler settings, digital inputs, analog inputs, alarm triggering, and sampling control.

      Q2: What modes are available in I/O sampler settings?
      Sampler settings include manual and timed modes. Manual mode starts sampling immediately when Start is clicked, while timed mode samples according to the selected time interval.

      Q3: Which sources can trigger alarm sampling?
      Sampling parameters can be configured based on Port 1 RS485 sensor alarms, Port 2 I/O alarms, and signals received through P2P communication.

      Q4: Which devices can be selected for Port 1 alarms?
      Port 1 allows alarm selection for sensor IDs 1 to 8, triggering sampling based on measurement alarm conditions.

      Q5: What is a Port 2 alarm?
      Port 2 is the I/O Port setting and can trigger sampling or logging based on digital or analog input alarm conditions.

      Q6: What types of alarm settings does the PWS include?
      Alarm settings include system alarms, digital sensor alarms, digital and analog input alarms, and barometric pressure alarms.

      Q7: What items are included in system alarms?
      System alarms may include low battery voltage, communication errors, and satellite positioning errors.

      Q8: Which alarm levels can be configured for ports?
      Four alarm levels can be configured: HH, HI, LO, and LL, with corresponding values.

      Q9: Are records stored when an alarm occurs?
      Yes. When an alarm starts or ends, the system records the alarm data in memory.

      Q10: What is P2P communication?
      P2P communication receives data according to the receiving register channel configured by the transmitting end. After the transmitter sends a value, the receiver displays it in the value field.

      Q11: Can P2P be used for sampling control?
      Yes. The PWS can configure sampling parameters based on signals from the transmitting host in P2P communication.

  • K
    1. Data Logging and Import/Export
      Q1: What data can the PWS record?
      The PWS can record sensor values, GPS coordinates, alarm data, sampling events, system status, communication data, and more.

      Q2: Can the PWS export historical data?
      Yes. Log records can be viewed and exported from the Import / Export page in AQCFG.

      Q3: What can be viewed on the AQCFG Import / Export page?
      Record and alarm data can be viewed, searched by time range, displayed, and exported.

      Q4: Can the PWS display real-time data?
      Yes. The real-time data page can display real-time values and real-time trends for all I/O.

      Q5: What can be configured in the real-time trend chart?
      Name, color, value, unit, and display interval can be selected. The chart also supports zoom in, zoom out, plot, stop, and clear.

      Q6: How long can the X-axis of the real-time trend chart be set?
      The manual states that the display interval can be set from 1 minute to 1 day.

      Q7: Can data upload be scheduled at fixed intervals?
      Yes. Data communication can be configured for timed upload, using 00:00 each day as the starting time for data return.

      Q8: Can data upload be scheduled?
      Yes. If Schedule is selected, the system will return data to the server only at the selected time points.

      Q9: Can data be uploaded immediately when an alarm occurs?
      Yes. If immediate alarm upload is enabled, the host immediately sends data back to the server when an alarm record occurs.

      Q10: Will the system retry if data upload fails?
      Yes. The retry count can be configured in operation settings. If data upload fails, the system will attempt to upload again.

  • L
    1. Firmware Update and Maintenance
      Q1: Can the PWS firmware be updated?
      Yes. The AQCFG system firmware page can load a .aqs firmware file and perform the update.

      Q2: What should be noted before updating firmware?
      During firmware update, make sure only one device is connected to avoid upload failure or an unrecoverable device condition.

      Q3: Can the firmware update process be interrupted?
      No. Once the update starts, it must not be interrupted. Wait until the update is complete.

      Q4: What should be done after firmware update?
      The manual states that after firmware update is complete, go to the Settings page under Import / Export and click Restore to reset the PWS to factory settings. After reconnecting and powering on, go to the Records page and delete data to ensure proper host operation.

      Q5: Can the PWS be disassembled for self-repair?
      It is not recommended. The safety instructions state that unless explicitly specified, do not attempt to disassemble or modify the product or components. Internal inspection, modification, and repair must be performed by AQUAS or authorized qualified specialists.

      Q6: What should be done if the PWS cable is damaged?
      If the sensor or device cable is damaged, contact AQUAS Customer Service. Do not cut, fold, damage, or replace the power cable by yourself.

      Q7: What should be checked during routine PWS maintenance?
      Regularly check the sampling tubing, inlet, outlet, sample bottle, battery, communication status, sensor connection, data upload, and alarm records.

      Q8: Does the sampling tubing need cleaning?
      Yes. If pollutants remain in the sampling tubing or the tubing is blocked, sampling volume and sample representativeness may be affected. Periodic cleaning or replacement is recommended according to site water quality.

      Q9: Does the sample bottle need to be replaced?
      Yes. Especially after alarm sampling, the operator must confirm that the sample bottle has been replaced and press the reset button before the next alarm sampling can occur.

      Q10: What should be considered when using the PWS outdoors?
      Pay attention to waterproofing, dust protection, battery condition, cable fixation, antenna or communication-module signal, and avoid external pulling or damage to tubing and cables.

  • M
    1. Troubleshooting
      Q1: What should be done if the PWS cannot connect to the computer?
      Check that the USB Driver is installed, CAB06 is connected to PWS Port 0, Device Manager shows USB Serial Port, the COM Port is correct, and the baud rate is correct.

      Q2: What should be done if AQCFG cannot scan the PWS?
      Check that the PWS is powered, the connection cable is correct, the COM Port is correct, the ID range is correct, the baud rate is correct, and then scan again.

      Q3: What should be done if Bluetooth cannot connect?
      Confirm that the device has the optional Bluetooth module, Bluetooth is enabled, the computer can find the PWS, and password 0000 is entered for pairing.

      Q4: What should be done if the PWS cannot upload to AQWEB?
      Check that the 4G / GPRS module is enabled, the APN is correct, the SIM card is usable, the server IP and Port are correct, and perform a connection test.

      Q5: What should be done if the 4G signal is too weak?
      If the dBm value is greater than -93 dBm, the signal is too weak. Adjust the installation position, antenna position, or use a telecom plan with better coverage.

      Q6: What should be done if FTP cannot upload?
      Confirm that the FTP Server IP, Port 21, username, and password are correct, and click Connection Test to verify.

      Q7: What should be done if no water enters the sample bottle?
      Check that the inlet is placed in water, the sampling tube is not blocked, the sampling time is sufficient, the outlet is correctly connected to the sample bottle, and the pump is running.

      Q8: What should be done if the sample volume is insufficient?
      Check the sampling tube length, sampling time, sample bottle capacity, and whether there is air, blockage, or leakage in the tube. Longer sampling tubes require longer sampling time.

      Q9: Why does the PWS not sample again after an alarm?
      Because the PWS is a single-bottle sampler. After alarm sampling, the sample bottle must be manually replaced and the reset button pressed before the next alarm-triggered sampling cycle can occur.

      Q10: What should be done if an SMR sensor cannot be scanned?
      Check that the SMR sensor is connected to Port 1, communication settings are consistent, the SMR ID is within 1 to 8 and not duplicated, and sensor power and wiring are normal.

      Q11: What should be done if the level-meter reading is abnormal?
      Check that SMR02 settings, measured liquid density, level-height unit, and logging method are correct. Perform water-level compensation and zero calibration if necessary.

      Q12: What should be done if SMS messages are not received?
      Confirm that the SIM card supports SMS, the phone number is correct, Alarm Start and Alarm End selections are correct, and test the function using SMS Test first.

      Q13: What should be done if GPS cannot position?
      Check whether a GPS module is installed, GPS is enabled, the site is not blocked, and signal strength is sufficient. If no GPS module is installed, use manual coordinate entry.

      Q14: What should be done if data is not recorded?
      Check that the system is turned on, operation settings are set to continuous or scheduled operation, the logging interval is configured, and sensors and I/O are enabled.

      Q15: What should be done if uploaded data is incomplete?
      Check the communication signal, server settings, upload interval, alarm upload settings, and retry count. Also confirm historical data in AQWEB or the Import / Export page.