ONE Micro Data Logger

Frequently Asked Questions

  • A
    1. Product Basics
      Q1: About ONE?
      ONE is AQUAS’s compact and cost-effective data logger, designed for applications such as water level, water quality, flow, rainfall, and manhole cover monitoring.

      Q2: What types of sites is ONE suitable for?
      ONE is suitable for groundwater, sewer systems, mining sites, underground aquifers, flooding or stormwater locations, municipal water supply, construction and mining applications, and meteorological monitoring environments.

      Q3: How many digital sensors can ONE connect to?
      ONE has a built-in RS485 serial port and can connect to up to 4 digital sensors.

      Q4: Does ONE require an external power supply?
      Not necessarily. ONE supports battery operation, and its ultra-low-power design allows long-term operation without external power. Depending on the model, it can also be configured with external power, an internal battery, an external battery, or solar power.

      Q5: How long can ONE operate?
      According to the manual, ONE can operate continuously for more than 5 years on battery power. With a primary lithium battery, some configurations can operate for up to 10 years. Actual operating time depends on the logging interval, upload frequency, communication method, number of sensors, I/O modules, and site conditions.

      Q6: Is ONE waterproof?
      Yes. ONE uses an integrated fully sealed design with a rugged IP68 submersible enclosure, allowing direct installation in groundwater monitoring wells, wet areas, flooded environments, or outdoor sites.

      Q7: Does ONE require a control cabinet?
      In most cases, no. ONE has a fully sealed IP68 enclosure and can be installed directly in outdoor or underground monitoring sites, reducing the need for control cabinets, surge protectors, and signal converters.

      Q8: Can ONE be used with AQWEB?
      Yes. ONE can upload data to the AQWEB cloud management platform, enabling remote display, control, alarm overview, trend charts, statistical data analysis, and report queries.

      Q9: Does ONE support alarm functions?
      Yes. ONE can set alarm conditions based on water level, water quality, flow, rainfall, manhole cover switches, or other sensor states. When an abnormal condition occurs, it can log the alarm and send notifications.

      Q10: Does ONE support SMS alarms?
      Yes. ONE can store up to 10 mobile phone numbers and can send SMS messages when an alarm occurs or when the alarm ends.

      Q11: Does ONE support GPS?
      GPS is an optional feature. Coordinates can be entered manually, or the GPS module can be used to search for coordinates and store the location record.

      Q12: Can ONE log data offline?
      Yes. ONE includes data logging functionality. If communication is unavailable, it can store data locally and automatically transmit the buffered data after communication is restored.

  • B
    1. Product Features and Applications
      Q1: What are the main advantages of ONE?
      ONE offers ultra-low power consumption, a rugged IP68 waterproof enclosure, multi-parameter monitoring, multiple communication options, cloud data upload, alarm notification, battery-powered operation, and quick installation.

      Q2: Is ONE suitable for groundwater monitoring?
      Yes. ONE can be installed directly in groundwater monitoring wells and used with water level, water quality, or other SMR sensors for long-term data logging and upload.

      Q3: Is ONE suitable for sewer or manhole monitoring?
      Yes. ONE can be used for sewer monitoring, manhole cover switch monitoring, water level, flow, and water quality monitoring, and can send alarms when abnormal conditions occur.

      Q4: Can ONE be used for flooding or stormwater monitoring?
      Yes. With its IP68 waterproof enclosure and low-power communication capability, ONE is suitable for flooded, stormwater, outdoor, or humid environments.

      Q5: Does ONE support multi-parameter monitoring?
      Yes. Through RS485 sensors, ONE can monitor different parameters such as water level, water quality, flow, rainfall, and manhole cover switch status.

      Q6: Can ONE be integrated with PLC or SCADA systems?
      Yes. ONE supports communication protocols such as Modbus TCP, FTP, MQTT, RESTful, and HART, and can integrate with monitoring systems through OPC Server or database interfaces.

      Q7: Does ONE support database integration?
      Yes. The manual states that data can be linked with databases such as SQL and MySQL, and can also be integrated through AQWEB or AQOPC.

      Q8: Does ONE support mobile or web operation?
      Yes. ONE can support data viewing and remote operation through cloud platforms, mobile phones, or mobile devices. Actual functions depend on the communication module and platform configuration.

  • C
    1. Hardware, Connections, and Installation
      Q1: What are the main ports on ONE?
      ONE mainly includes Port 0 for system communication, Port 1 for RS485 sensors, Port 2 for I/O, and connectors related to antennas or external power.

      Q2: What is Port 0 used for?
      Port 0 is the system communication port. It can connect to a PC via CAB06 for AQCFG software configuration, operating mode setup, alarm settings, control parameters, and data upload/download.

      Q3: What is Port 1 used for?
      Port 1 is the RS485 sensor port and can connect to SMR digital sensors or Modbus instruments.

      Q4: What is Port 2 used for?
      Port 2 is the I/O port and supports two DI channels for status input or counter input.

      Q5: What sensors can ONE connect to?
      ONE can connect to SMR sensors for water level, pH, ORP, conductivity, dissolved oxygen, turbidity, salinity, suspended solids, ammonia nitrogen, nitrate nitrogen, oil in water, chemical oxygen demand, and other parameters.

      Q6: What are the dimensions of ONE?
      The dimensional drawing in the manual shows that ONE is approximately 85 mm × 171 mm × 52 mm.

      Q7: What is the weight of ONE?
      The technical specifications in the manual list ONE at approximately 600 g.

      Q8: What is the enclosure material of ONE?
      The ONE enclosure is made of ABS.

      Q9: What installation methods does ONE support?
      ONE can be wall-mounted, pole-mounted, ground-mounted, or installed underground depending on site conditions.

      Q10: Can ONE be installed with a solar panel?
      Yes. ONE can be used with a solar panel and rechargeable lithium battery, making it suitable for sites without mains power.

      Q11: What should be noted when replacing or installing the SIM card?
      Remove the screw protection cap on the top of the unit, loosen the four screws, and open the top cover. Disconnect the internal battery cable first to prevent short circuits before installing the SIM card. After installation, make sure no cables are pinched and secure the top cover again.

      Q12: Can ONE be disassembled by the user?
      The manual provides disassembly steps for SIM card replacement. However, general internal inspection, modification, and repair should still be performed by AQUAS or authorized qualified specialists to prevent waterproofing failure or equipment damage.

  • D
    1. Power Supply and Battery
      Q1: What power supply options does ONE support?
      ONE can support external power, an internal primary battery, an internal rechargeable battery, an external battery box, a solar panel, and external battery configurations.

      Q2: What internal battery types are available for ONE?
      The manual lists BAT05-1 as a 3.6 VDC 13 Ah rechargeable lithium battery, and BAT10 as a 3.6 VDC 13 Ah primary lithium battery.

      Q3: What external battery options are available for ONE?
      The manual lists BAT11-2 as a 3.6 VDC 240 Ah rechargeable lithium battery box.

      Q4: How long can ONE operate with a primary battery?
      The manual specifications state that the BAT10 primary lithium battery can provide a service life of up to 10 years. Actual life should be evaluated based on logging and upload frequency.

      Q5: How long can ONE operate with a rechargeable battery?
      The manual lists the 3.6 VDC 13 Ah rechargeable lithium battery as having a life of approximately 300 charge/discharge cycles. When used with a solar panel and external battery, outdoor operating time can be extended.

      Q6: What should be done after replacing a primary battery?
      After replacing a primary battery, press “Reset” on the AQCFG system page so the system can correctly calculate the battery level.

      Q7: Why is it important to select the correct power mode?
      The power mode affects battery voltage, capacity, and remaining runtime estimation. Incorrect settings may cause inaccurate battery level display.

      Q8: Is there a low-battery alarm?
      Yes. ONE supports low-battery alarms. Conditions can be configured on the alarm page for logging or notification.

      Q9: What are the risks of using an incorrect battery?
      The manual warns that using an incorrect battery type or a non-original charger may create an explosion risk. Used batteries must be handled according to the manufacturer’s instructions.

  • E
    1. AQCFG Software Installation and Connection
      Q1: Which software is used to configure ONE?
      ONE uses AQUAS AQCFG software for device search, connection, system settings, sensor settings, communication settings, alarm settings, data export, and firmware updates.

      Q2: How is AQCFG installed?
      Insert the CD or USB drive, select the AQCFG folder, and click the Setup icon to start installation. Follow the prompts by clicking Next, then click Install to complete the installation.

      Q3: How does ONE connect to a computer?
      Use the CAB06 communication cable to connect ONE Port 0 to the PC, then open AQCFG to search for and configure the device.

      Q4: What must be installed before using CAB06?
      The USB driver must be installed first. Open the USB Driver folder and run CDM20824_Setup. After installation, a new Serial COM Port will appear in Device Manager.

      Q5: What should be configured before searching for ONE in AQCFG?
      Click “Settings” and select the COM Port, baud rate, and Modbus ID scan range, then close the settings page when complete.

      Q6: What is the default AQCFG baud rate for ONE?
      The manual’s quick operation and connection instructions indicate a default baud rate of 115200 bps.

      Q7: How do you enter the system after AQCFG finds ONE?
      Click “Link/Connect,” or double-click the ONE device to enter the system settings page.

      Q8: Does AQCFG automatically display the USB COM Port?
      Yes. According to the manual screen, the connected USB COM Port will appear automatically after the USB cable is connected.

      Q9: What should I do if AQCFG cannot find ONE?
      Check that CAB06 is connected to Port 0, the USB driver is installed, the COM Port is correct, the baud rate is set to 115200, the ID range includes the host ID, and the device is powered.

      Q10: What should be done first after entering host settings?
      Click the System ON button first. Settings and data logging can only be performed after the system is powered on.

  • F
    1. Host System Settings
      Q1: What can be configured on the system page?
      The system page can be used to configure or view the model, serial number, location, area, sub-area, system time, battery mode, battery voltage, battery level, communication rate, and Modbus ID.

      Q2: Can the model field be modified?
      No. The model is created during production and is used to verify the hardware configuration. It cannot be modified by the user.

      Q3: What is the serial number?
      The serial number is the unique device ID information for each ONE unit, used for device identification and management.

      Q4: What can be entered in the location field?
      The location field can be set by the user and can contain up to 20 letters, making it easier to manage and identify devices in AQWEB.

      Q5: What are the area and sub-area fields used for?
      The area and sub-area fields can be used for platform classification and management, helping distinguish different monitoring areas or stations.

      Q6: How is ONE system time synchronized?
      You can choose no synchronization, PC/Server synchronization, or GPS synchronization. After selecting PC/Server, set the time zone and press Confirm; the system time will synchronize with the server.

      Q7: Can GPS be used for time synchronization?
      Yes. If a GPS module is installed, the host time can be synchronized with GPS.

      Q8: How are settings saved after configuration?
      After all field data is configured, press “Confirm” again to save the data.

      Q9: Can the Modbus ID of ONE be configured?
      Yes. The system page can display and configure the current Modbus ID, making it easier for AQCFG to search and connect.

      Q10: What does System ON do?
      System ON indicates that the host system is activated. If System ON has not been performed, some settings and data logging functions may not operate correctly.

  • G
    1. 4G / GPRS / FTP Communication
      Q1: What communication methods does ONE support?
      Depending on the model, ONE can support GPRS, 4G, NB-IoT, Wi-Fi, Ethernet, LoRa, satellite communication, RS485, and Bluetooth.

      Q2: What data can ONE send back to the server?
      ONE can send initial configuration parameters, real-time values, alarms, logged data, and other information back to the server via GPRS, NB-IoT, or 4G.

      Q3: What should be done before configuring 4G or GPRS?
      Before configuration, check “Enable” and press Confirm. After completing the data settings, press Confirm again to save the data.

      Q4: What does the 4G status “Standby” mean?
      Standby means the module is powered off or in standby mode.

      Q5: What does “Try Connecting” mean?
      Standby → Try Connecting indicates that the device is establishing a connection.

      Q6: What does “Connected → Data Communication” mean?
      It means the connection is successful and data is being transmitted.

      Q7: How is communication signal strength evaluated?
      AQCFG displays both a 0%–100% value and a dBm value. The manual indicates that if the dBm value is greater than -93 dBm, the signal is too weak and the device may not be able to connect to the server.

      Q8: How should the APN be configured?
      The APN, username, and password should be entered according to the mobile operator’s network information. They can also be obtained by selecting the country and carrier from the network service provider menu on the right, or by contacting the carrier.

      Q9: What needs to be configured in CCS mode?
      In CCS mode, enter the server IP and server port. The manual lists the default server port as 502.

      Q10: What is the default AQUAS cloud server setting?
      The quick operation screen in the manual shows the default AQUAS cloud server as 118.163.120.143:502. For actual projects, confirm the setting with the shipment configuration or platform information.

      Q11: Can ONE be configured with a backup server?
      Yes. If ONE data needs to be sent to a second server at the same time, enable the backup server option and enter the IP and Port of the second server.

      Q12: What information is required for FTP mode?
      FTP mode requires the FTP server IP, Port, username, and password. The default FTP Port is 21.

      Q13: What is CCS / FTP mode?
      Select CCS / FTP mode when both the CCS server and FTP server need to receive data uploaded by ONE.

      Q14: How can communication be verified after configuration?
      Click “Connection Test” to confirm whether data can be uploaded to the server successfully.

  • H
    1. GPS and SMS
      Q1: Is GPS a standard feature on ONE?
      GPS is optional. If the GPS module is not installed, coordinates can be entered manually.

      Q2: What can GPS be used for?
      GPS can search for device coordinates, synchronize time, record location, and allow AQWEB or other platforms to display the monitoring station location.

      Q3: What does GPS status “off” mean?
      off means the GPS module is turned off.

      Q4: What does GPS status “Searching” mean?
      Searching means the GPS module is enabled and searching for satellite positioning.

      Q5: What does GPS status “Position” mean?
      Position means positioning has been completed.

      Q6: How is the GPS signal displayed?
      AQCFG displays the GPS module signal strength in a range from 0% to 100%.

      Q7: Can GPS be logged periodically?
      Yes. After “Periodic” is selected, ONE will record GPS coordinate values together with the logging cycle.

      Q8: When is GPS continuous mode suitable?
      Continuous mode keeps the GPS module enabled. The manual recommends using this mode when external power is available.

      Q9: What is GPS search mode?
      Search mode manually turns on the GPS function. After one search is completed, GPS will automatically turn off regardless of whether positioning succeeds.

      Q10: What is the minimum interval for GPS periodic mode?
      The manual states that the minimum interval for GPS periodic mode is 5 minutes.

      Q11: How many SMS phone numbers can ONE store?
      Up to 10 mobile phone numbers can be configured for SMS notifications.

      Q12: When can SMS messages be sent?
      For each phone number, SMS sending can be enabled for alarm occurrence and alarm end events.

      Q13: Why can’t the alarm end SMS option be selected directly?
      The manual notes that “Alarm Occurrence” must be selected before “Alarm End” can be selected.

      Q14: How can the SMS function be tested?
      Press “SMS Test,” and the host will send a test SMS message to the first mobile phone number.

  • I
    1. RS485 Sensors and I/O
      Q1: How does ONE connect to SMR sensors?
      Connect the SMR sensor to ONE Port 1 using CAB12 or an RS485 cable, then scan for the sensor on the AQCFG sensor port page.

      Q2: What is the default communication setting for ONE Port 1?
      The manual shows an example SMR sensor Port 1 setting of 19200 bps, N, 8, 1.

      Q3: How many SMR sensors can ONE scan?
      ONE can connect to up to 4 digital sensors. However, the AQCFG scan list may show Modbus IDs 1–8. The actual usable quantity should be confirmed based on the ONE model and hardware configuration.

      Q4: What restrictions apply to SMR sensor IDs?
      Each SMR sensor ID must be unique and within the host scan range.

      Q5: How is an SMR sensor configured after it is scanned?
      Click the sensor in the AQCFG tree list, or double-click the device in the sensor list to open that sensor’s settings page.

      Q6: Which SMR sensors are shown as examples for ONE?
      The manual screens show example sensors such as SMR04 pH, SMR05 ORP, SMR07 conductivity, and SMR09 dissolved oxygen.

      Q7: Can logging methods be configured for SMR sensors?
      Yes. Periodic logging and alarm logging can be configured on the sensor settings page.

      Q8: Can ONE calibrate SMR sensors?
      ONE can access calibration pages for SMR sensors, such as pH, ORP, conductivity, and dissolved oxygen. Detailed standard solutions and calibration procedures should still follow the corresponding SMR operation manual.

      Q9: Does ONE support I/O modules?
      Yes. The I/O module on ONE Port 2 supports two DI channels for pulse counting or status input.

      Q10: What applications can DI be used for?
      DI can be used for manhole cover switches, rainfall pulses, flow pulses, equipment status, or other digital input signals.

  • J
    1. Operation Settings, Alarms, and Data Logging
      Q1: What can be configured in operation settings?
      Operation settings can configure the system operating mode, scan cycle, logging cycle, alarm logging cycle, communication module, upload schedule, periodic upload, and retry count.

      Q2: Does ONE support continuous operation?
      Yes. After “Continuous” is selected, the system can continuously scan, log, and upload data according to the configured cycles.

      Q3: Does ONE support scheduled operation?
      Yes. A schedule can be configured so the system operates or uploads data only at specified times.

      Q4: What is the scan cycle?
      The scan cycle is the time interval at which ONE scans sensor or I/O values and determines status and alarm conditions.

      Q5: What is the logging cycle?
      The logging cycle is the time interval at which ONE writes sensor or I/O data to the log records.

      Q6: What is the alarm logging cycle?
      The alarm logging cycle is the time interval for accelerated data logging during an alarm or under alarm level conditions.

      Q7: Can data upload be scheduled periodically?
      Yes. Periodic upload can be configured, with daily 00:00 used as the starting time for data transmission according to the settings.

      Q8: Can data upload be scheduled?
      Yes. If schedule mode is selected, the system will send data back to the server only at the selected times.

      Q9: Can data be uploaded immediately when an alarm occurs?
      Yes. If immediate alarm upload is enabled, the host will send data back to the server immediately when an alarm record occurs.

      Q10: What alarm types does ONE support?
      ONE supports system alarms, sensor alarms, and I/O module alarms, such as Wi-Fi communication failure, mobile communication failure, low voltage, and upper/lower limits for sensor values.

      Q11: What alarm levels can be configured?
      Alarm levels such as HH, H, L, and LL can be configured, along with the corresponding values and activation conditions.

      Q12: Can ONE export data?
      Yes. Records and alarm data can be searched and exported from the AQCFG Import/Export page.

      Q13: Can ONE be restored to factory defaults?
      Yes. The AQCFG Import/Export page provides a factory reset function. After the reset, reconnect the device and confirm the settings.

  • K
    1. Firmware Update and Maintenance
      Q1: Can ONE firmware be updated?
      Yes. The AQCFG system firmware page can load the correct version of the .aqs file to perform a firmware update.

      Q2: What should be noted before a firmware update?
      Before updating, make sure only one device is connected to avoid upload failure or unrecoverable device errors.

      Q3: Can the firmware update process be interrupted?
      No. Once the firmware update starts, it must not be interrupted. Wait until the update is complete.

      Q4: What should be done after the firmware update is complete?
      The manual states that after the firmware update is complete, go to the settings page under Import/Export and click “Restore” to restore ONE to factory settings. After reconnecting and powering on the device, go to the records page and delete data to ensure normal host operation.

      Q5: Does ONE require regular maintenance?
      Yes. Regularly check the enclosure, waterproof condition, connectors, battery, communication signal, sensor connections, data logging status, and data upload status.

      Q6: What should be done if the ONE cable is damaged?
      If a sensor or device cable is damaged, contact AQUAS Customer Service. Cutting, folding, damaging, or replacing the power cable by yourself is not recommended.

      Q7: Can ONE be modified by the user?
      No. Unless explicitly stated in the manual, do not disassemble or modify the product or its components by yourself.

      Q8: What should be noted when using ONE outdoors?
      Make sure the top cover is tightened, waterproof connectors are secured, cables are fixed, the antenna is properly positioned, the battery level is sufficient, the communication signal is stable, and platform upload status is normal.

      Q9: What should be noted when installing ONE underground or inside a well?
      Ensure the device is securely fixed, avoid long-term cable tension, and confirm that the antenna or communication module can maintain sufficient signal strength.

  • L
    1. Common Troubleshooting
      Q1: What should I do if ONE cannot connect to the computer?
      Check whether CAB06 is connected to Port 0, the USB driver is installed, Device Manager shows the Serial COM Port, the correct COM Port is selected, and the baud rate is correct.

      Q2: What should I do if AQCFG cannot scan ONE?
      Check that ONE is powered, the COM Port is correct, the baud rate is set to 115200, and the ID range includes the device ID, then scan again.

      Q3: What should I do if ONE cannot upload data?
      Check whether the communication module is enabled, the APN is correct, the SIM card is available, and the server IP and Port are correct, then run a connection test.

      Q4: What should I do if the 4G signal is too weak?
      If the dBm value is greater than -93 dBm, the signal is too weak. Adjust the antenna position or installation location, or confirm mobile network coverage.

      Q5: What should I do if FTP upload fails?
      Confirm that the FTP server IP, Port 21, username, and password are correct, then click Connection Test.

      Q6: What should I do if SMS messages are not received?
      Confirm that SMS is enabled, the phone number is correct, the SIM card supports SMS, and the alarm occurrence and alarm end options are selected correctly, then use SMS Test first.

      Q7: What should I do if GPS cannot obtain a position?
      Confirm whether a GPS module is installed, GPS is enabled, the site has no obstruction, and signal strength is sufficient. If GPS is not installed, enter the coordinates manually.

      Q8: What should I do if sensors cannot be scanned?
      Confirm that the SMR sensor is connected to Port 1, the RS485 wiring is correct, the communication setting is 19200.N81 or matches the sensor setting, and each sensor ID is unique.

      Q9: What should I do if data is not being logged?
      Check System ON, the logging cycle in operation settings, whether the sensor is enabled, and whether periodic or alarm logging is selected.

      Q10: What should I do if alarms are not triggered?
      Check whether the alarm page is enabled, alarm levels and threshold values are configured, and alarm logging is selected for the sensor or I/O.

      Q11: What should I do if the battery level display is inaccurate?
      Confirm that the power mode and battery type are correct. If a primary battery has been replaced, press “Reset” to recalculate the battery level.

      Q12: What should I do if the firmware update fails?
      Confirm that only one device is connected, the firmware file version is correct, and power and communication remain stable during the update. If it still fails, contact AQUAS Technical Support.