CTD Level and Conductivity Analyzer

Frequently Asked Questions

  • A
    1. Product Basics
      Q1: About CTD?
      CTD is the AQUAS conductivity, depth, and temperature multiparameter water quality analyzer. It is used for real-time monitoring of water level, conductivity, temperature, and related calculated water quality parameters.

      Q2: What parameters can CTD measure?
      CTD measures water level, conductivity, and temperature. Conductivity can also be used to calculate specific conductance, salinity, total dissolved solids (TDS), resistivity, density, and other parameters.

      Q3: What applications is CTD suitable for?
      CTD is suitable for drinking water, surface water, groundwater, and mine water quality monitoring. It is also suitable for hydrological monitoring applications such as groundwater aquifers, boreholes, wells, reservoirs, lakes, coastal estuaries, and wetlands.

      Q4: Is CTD suitable for groundwater or borehole applications?
      Yes. With a diameter of less than 1 inch, CTD is suitable for space-constrained installations such as groundwater wells, boreholes, and mining sites.

      Q5: Does CTD require sampling or pretreatment?
      No. CTD supports in-situ, online, real-time monitoring without the need for sampling or pretreatment equipment.

      Q6: Can CTD be submerged for long-term use?
      Yes. CTD uses a rugged submersible housing. The manual specifies an IP68 protection rating and continuous submersion capability down to 250 m.

      Q7: Does CTD have a built-in battery?
      Yes. CTD is powered by a built-in battery. The manual states that it can operate for more than 10 years at maximum, while actual battery life depends on logging frequency, upload frequency, operating environment, and battery condition.

      Q8: Does CTD support data storage?
      Yes. CTD supports SD card data storage and backup for long-term monitoring data retention.

      Q9: Does CTD support PLC or HMI integration?
      Yes. CTD provides the standard Modbus RTU communication protocol and can connect to PLCs, HMIs, or data loggers.

      Q10: Can CTD connect to ECO?
      Yes. The manual appendix provides a connection diagram for CTD and the ECO host, using an RS485 cable.

      Q11: Can CTD connect to a computer?
      Yes. CTD can connect to a computer via the CAB06 USB-to-RS485 cable, and AQCFG software can be used for setup, readings, and calibration.

      Q12: Does CTD have self-diagnostic functions?
      Yes. CTD includes built-in status monitoring and abnormality diagnostics covering temperature, humidity, water level sensor status, and related conditions.

  • B
    1. Measurement Principles
      Q1: What is the conductivity measurement principle of CTD?
      CTD uses graphite electrodes for conductivity measurement and calculates conductivity by measuring the electrical resistance of the solution.

      Q2: Does CTD conductivity measurement comply with EPA methods?
      The manual states that the CTD conductivity measurement method complies with EPA automatic monitoring facility method NIEA W204.51C.

      Q3: What is the temperature measurement principle of CTD?
      CTD uses an RTD Pt1000 platinum resistance sensor to measure temperature. The sensor resistance changes with temperature.

      Q4: Does CTD temperature measurement comply with EPA methods?
      The manual states that CTD temperature measurement complies with EPA automatic monitoring facility method NIEA W218.51C.

      Q5: What is the water level measurement principle of CTD?
      CTD uses piezoresistive sensing technology. Liquid pressure is transmitted through a diaphragm to a silicon piezoresistive sensing element, then converted into water level data through signal processing and temperature compensation.

      Q6: What components are included in the CTD water level sensing structure?
      The CTD water level measurement structure includes a silicon piezoresistive sensing element, titanium alloy diaphragm, signal processing circuit, and oil-filled isolation structure.

      Q7: Does CTD provide temperature compensation?
      Yes. CTD provides temperature and density compensation to improve the stability and accuracy of water level, conductivity, and related measurements.

      Q8: Are specific conductance, TDS, salinity, resistivity, and density directly measured?
      These parameters are mainly calculated from conductivity, including specific conductance, TDS, salinity, resistivity, and density.

  • C
    1. Specifications and Models
      Q1: What housing materials are available for CTD?
      CTD housing materials include POM, SS316L, and titanium alloy. The actual material depends on the ordered model.

      Q2: What are the dimensions of CTD?
      The manual dimension drawing shows CTD1 at approximately Ø18 x 210 mm and CTD2 at approximately Ø18 x 310 mm.

      Q3: What is the weight of CTD?
      The manual specifications list approximately 1.8 kg for the SS316L version and approximately 1.5 kg for the titanium version.

      Q4: What are the features of the CTD cable?
      CTD uses a Kevlar-reinforced PUR cable with 1500 N tensile strength, improving durability for field installation and long-term use.

      Q5: What connector types does CTD use?
      CTD uses an M16 plug fixed cable or an M12 connector. The 5-pin connector type depends on the model.

      Q6: What is the operating voltage of CTD?
      The CTD external power supply is 5-12 VDC, 1 A.

      Q7: What is the operating temperature range of CTD?
      The manual specifications list the operating temperature as -40 to 85°C.

      Q8: What is the protection rating of CTD?
      CTD has an IP68 protection rating and is suitable for humid, flooded, or submerged environments.

      Q9: What water level measurement ranges are available for CTD?
      Depending on the model, water level or depth measurement ranges may include 0-10 m, 0-50 m, 0-200 m, and 0-350 m. Calibration report examples also list 0-10 m, 0-50 m, and 0-100 m.

      Q10: What is the conductivity measurement range of CTD?
      The manual technical specifications list a conductivity measurement range of 0-100 mS/cm. Calibration report examples list conductivity up to 0-500 mS/cm; the actual range should be confirmed based on the delivered model and calibration data.

      Q11: What is the temperature measurement range of CTD?
      The CTD temperature measurement range is -20 to 65°C.

      Q12: What is the conductivity resolution of CTD?
      The manual specifications list the conductivity resolution as 0.1 μS/cm.

      Q13: What is the temperature resolution of CTD?
      The CTD temperature resolution is 0.01°C.

      Q14: What is the water level resolution of CTD?
      The CTD water level resolution is ±0.005% FS.

  • D
    1. Installation
      Q1: How is CTD installed?
      CTD is installed by immersion and can be installed in wells, tanks, pipelines, rivers, groundwater sites, or other hydrological monitoring locations according to site requirements.

      Q2: What should be checked before installing CTD?
      Before installation, confirm that the cable length is sufficient and properly secure and protect the cable to prevent damage caused by pulling.

      Q3: Where is CTD recommended to be installed?
      It is recommended to install CTD at a location with stable water flow for better measurement results.

      Q4: Does the CTD signal cable need shielding?
      Yes. To ensure signal stability and reduce electromagnetic interference, use cables with metallic braided shielding or shielding foil and properly manage the shielding protection.

      Q5: How should the CTD shield be grounded?
      The manual recommends grounding the shield at one end only to avoid ground loop interference.

      Q6: Can the CTD signal cable be bundled with power cables?
      No. Signal cables should not be bundled in parallel with power cables, and sharing the same conduit or cable tray is not recommended.

      Q7: What distance is recommended between signal and power cables?
      It is recommended to keep signal and power cables at least 30 cm apart. Use separate cable trays where conditions allow.

      Q8: What should be considered if there are inverters or high-power equipment on site?
      Check for electromagnetic interference sources. If necessary, adjust equipment location, add metal conduits, or install additional shielding devices.

      Q9: Can CTD be installed on a flat surface or suspended?
      Yes. The manual installation diagrams include flat-surface installation, suspended installation, pole-mounted installation, and tank/pipeline application layouts.

      Q10: Does CTD need to be secured after installation?
      Yes. CTD should be secured according to the recommended method to prevent sensor displacement or cable damage caused by water flow, vibration, or pulling.

  • E
    1. Communication and Wiring
      Q1: What communication interface does CTD use?
      CTD uses an RS485 digital communication interface.

      Q2: What communication protocol does CTD use?
      CTD uses the Modbus RTU communication protocol.

      Q3: What are the default CTD communication settings?
      The factory default is 19200 bps, no parity, 8 data bits, and 1 stop bit: 19200, N, 8, 1.

      Q4: Which baud rates does CTD support?
      The manual states that CTD supports 9600 and 19200 bps.

      Q5: What is the default Modbus ID of CTD?
      The manual COM Port setup example lists Modbus ID = 1, and the Modbus table also shows the default MODBUS ID as 0x01.

      Q6: What should be considered when multiple CTDs or sensors connect to the same host?
      Each analyzer must be assigned a different device ID to avoid communication conflicts.

      Q7: Can CTD be set to 9600 or 19200?
      Yes. The AQCFG basic page can set the baud rate to 19200 or 9600.

      Q8: What communication formats can CTD be set to?
      Communication data formats such as N81, O81, and E81 can be configured.

      Q9: What is the wiring for the CTD SMR male connector?
      The SMR male connector wiring is as follows:
      PIN 1: Brown, GND PIN 2: White, RS485- PIN 3: Blue, RS485+ PIN 4: Black, 5-12 VDC PIN 5: Gray, not used

      Q10: What is the wiring when CTD uses 4-20 mA SAO02?
      SAO02 wiring is as follows:
      PIN 1: Brown, GND PIN 2: Red, RS485 D- PIN 3: Orange, RS485 D+ PIN 4: Yellow, power supply 5-12 VDC PIN 5: Green, not used PIN 6: Blue, output 4-20 mA CH1 PIN 7: Purple, output 4-20 mA CH2 PIN 8: Gray, GND

      Q11: Can CTD connect via RS232?
      Yes. The manual appendix provides an RS232 connection diagram. An RS232/485 converter and DC 5-12 V power supply are required.

      Q12: Can CTD connect to a PLC or HMI?
      Yes. The manual appendix provides a diagram for connection to a controller or HMI, including D+, D-, DC 12V+, GND, and shield grounding.

  • F
    1. AQCFG Software Operation
      Q1: Which software is used to configure CTD?
      CTD uses AQCFG software for device search, reading display, calibration, parameter configuration, firmware update, and fault diagnosis.

      Q2: How is AQCFG installed?
      After inserting the disc or obtaining the installer, open the corresponding folder, click the Setup icon, configure the installation path, access folder, and desktop shortcut, then click Install to complete installation.

      Q3: How does CTD connect to a computer?
      Use the CAB06 USB 485 cable to connect CTD to the computer, then open AQCFG software.

      Q4: How do I configure the COM Port?
      In the device search interface, click Setup to configure the COM Port, Modbus ID, baud rate, and parity.

      Q5: What are the steps for scanning CTD in AQCFG?
      After setting the COM Port and communication parameters, return to the device search page and click Scan. Once CTD is found, click Link or double-click the device to enter the basic settings window.

      Q6: What do the green and red AQCFG communication statuses mean?
      Green indicates connected; red indicates disconnected.

      Q7: Can AQCFG set decimal places?
      Yes. AQCFG Data Format can set the number of decimal places for displayed values.

      Q8: Can AQCFG switch languages?
      Yes. AQCFG language settings can change the interface language. The software must be restarted after the change.

      Q9: What information is displayed on the CTD basic page?
      The basic page displays analyzer, model, serial number, location, region, subregion, GPS positioning information, measurement range, water level, conductivity, temperature, specific conductance, TDS, salinity, resistivity, and density.

      Q10: How do I access calibration functions?
      On the AQCFG main page, click Advanced to display calibration and firmware update functions, then select water level calibration or conductivity calibration.

      Q11: Can CTD be restored to factory defaults?
      Yes. The Advanced page provides a factory reset function that restores factory settings and resets all parameters to their default values.

  • G
    1. Modbus Readings
      Q1: Which address should be read for CTD water level values?
      Water level values can be read from 3xxxx read-only register 0x0000. The data type is Float, length is 2 words, and Function Code is 04.

      Q2: Which address should be read for CTD conductivity values?
      Conductivity values can be read from 0x0001. The data type is Float, length is 2 words, and Function Code is 04.

      Q3: Which address should be read for CTD temperature values?
      Temperature values can be read from 0x0003. The data type is Float, length is 4 words, and Function Code is 04.

      Q4: How do I read the CTD model?
      Read 0x1001, length 20 words, data type Ansi-C String, Function Code 04.

      Q5: How do I read the CTD internal temperature?
      Internal temperature can be read from 0x04F1, corresponding to decimal 1266, length 2 words.

      Q6: How do I read the CTD internal humidity?
      Internal humidity can be read from 0x04F3, corresponding to decimal 1268, length 2 words.

      Q7: How do I read raw conductivity before temperature compensation?
      Read 0x00C, corresponding to decimal 13, length 2 words.

      Q8: How do I read depth pressure?
      Depth pressure can be read from 0x0022, corresponding to decimal 35, length 2 words.

      Q9: How do I read and write the location name?
      The location name can use 4xxxx read/write register 0x0000, length 20 words, data type Ansi-C String.

      Q10: How do I read and write the Modbus ID?
      Modbus ID can use 4xxxx read/write register 0x0001, length 1 word, default value 0x01.

      Q11: How do I set the CTD communication protocol?
      Protocol can be configured using 0x0002, length 1 word. The default value is 12, representing 19200 N81.

      Q12: What is the difference between 3xxxx and 4xxxx registers?
      3xxxx registers are read-only Input Registers and support Function Code 04. 4xxxx registers are Holding Registers and support Function Codes 03, 06, and 16 for reading and writing configuration parameters.

      Q13: What should I do if Modbus readings are garbled or unreasonable?
      Confirm that the Function Code, starting address, length, floating-point decoding method, communication parameters, and Modbus ID are correct, and verify them against the manual Modbus table.

  • H
    1. Water Level Calibration
      Q1: When does CTD require water level calibration?
      When water level readings are abnormal, zero offset occurs, or there is a clear difference from a standard measuring device, water level zero or slope calibration can be performed.

      Q2: What tools are required for water level calibration?
      A tape measure or other water level measuring tool is required, as well as a container at least 1.5 m deep, depending on calibration requirements.

      Q3: Are chemicals required for water level calibration?
      No. No chemicals are required for water level calibration.

      Q4: How is water level zero calibration performed?
      Open the water level calibration function on the Advanced settings page, place CTD in air, click Start on the zero page, enter the target liquid density (pure water can be treated as 1), then click OK to complete.

      Q5: Why is liquid density required during water level zero calibration?
      Because CTD water level calculation is affected by liquid density. Entering the correct density improves water level calculation accuracy.

      Q6: What target liquid density should be entered for pure water?
      The manual states that pure water can be treated as 1.

      Q7: How is water level slope calibration performed?
      Open the slope calibration page, place CTD in the target liquid at a depth of at least 1 m, measure the actual immersion height with a tape measure, click Start to perform slope calibration, enter the measured water level value, then click OK.

      Q8: How deep should the water be for water level slope calibration?
      The manual procedure states that CTD should be placed in the target liquid at least 1 m deep. Tool preparation recommends a container at least 1.5 m deep, depending on requirements.

      Q9: Is water level calibrated before shipment?
      Yes. The manual notes that all devices are calibrated with a standard pressure fixture before shipment.

      Q10: What should be done first when water level readings are abnormal?
      It is recommended to perform basic cleaning and zero calibration first. If readings still differ significantly after zero calibration, use a standard measuring device to measure the actual water level depth and perform slope calibration.

      Q11: Can CTD perform temperature correction?
      Yes. The water level calibration page provides temperature correction, allowing the correct temperature value to be entered for correction.

  • I
    1. Conductivity Calibration
      Q1: When does CTD require conductivity calibration?
      Conductivity calibration can be performed when conductivity readings drift, readings are unstable, accuracy needs to be confirmed after maintenance, or periodic maintenance is required.

      Q2: What tools are required for conductivity calibration?
      An AQUAS dedicated calibration cup, wash bottle, and lens tissue are required.

      Q3: What standard solution is required for conductivity calibration?
      A 1.413 mS/cm standard solution is required.

      Q4: What should be done before conductivity calibration?
      Before calibration, thoroughly clean the device. Rinse the sensor surface with tap water, use a soft brush and neutral detergent if necessary, then rinse with pure water and wipe dry.

      Q5: How is conductivity zero calibration performed?
      Open the conductivity calibration function on the Advanced page, place the electrode in air, click Start on the zero page, enter 0 in the pop-up window, and the slope button will appear after completion.

      Q6: How is conductivity slope calibration performed?
      Open the slope calibration page, rinse the EC electrode with pure water and wipe it dry with lens tissue. Pour 1.413 mS/cm standard solution into the calibration cup, insert the electrode into the standard solution, ensure it is fully immersed without touching the bottom of the cup, click Start, enter 1.413 mS/cm, and click OK to complete.

      Q7: Can the electrode touch the bottom of the cup during conductivity calibration?
      No. The electrode must be fully immersed in the standard solution but must not touch the bottom of the calibration cup.

      Q8: Do bubbles affect conductivity calibration results?
      Yes. The manual reminds users to ensure that there are no bubbles around the electrode in the standard solution.

      Q9: What temperature conditions should be considered during conductivity calibration?
      It is recommended to maintain the ambient temperature at 25°C ± 0.5°C. If ambient temperature cannot be controlled, ensure that the temperature difference between zero and slope calibration is less than 1°C.

      Q10: How do I verify conductivity calibration after completion?
      Return to the basic page to view the real-time conductivity value, confirm that the standard solution reading is close to 1.413 mS/cm, and check whether the reading is stable.

  • J
    1. Firmware Update
      Q1: Can CTD firmware be updated?
      Yes. The AQCFG Advanced page provides firmware version display and firmware update functions.

      Q2: How do I check the CTD firmware version?
      Open the AQCFG Advanced page to view the currently installed firmware version and date.

      Q3: What are the CTD firmware update steps?
      Select Open to load the correct .aqs firmware file, then select Download to start the update. When the AQUAS INC DISCLAIMS window appears, click YES to start installation or No to exit.

      Q4: What should be considered during firmware update?
      During the update, the computer should be connected only to the target device to avoid update failure or unrecoverable system errors caused by multiple connected devices.

      Q5: Can firmware update be interrupted?
      No. The update process must not be interrupted. Ensure stable power and communication to prevent device damage caused by power loss.

  • K
    1. Maintenance
      Q1: Does CTD require regular maintenance?
      Yes. Regular maintenance and cleaning are required to keep CTD in optimal operating condition, especially in environments where grease, sediment, or contaminants easily accumulate.

      Q2: How often should routine CTD maintenance be performed?
      The manual recommends routine maintenance at least once per week. The actual frequency can be adjusted based on the application and monitoring environment.

      Q3: What items are included in routine CTD maintenance?
      Routine maintenance includes cleaning the analyzer surface and sensing area, checking battery level, checking consumables, and confirming that cables and mounting devices are secure.

      Q4: How often should periodic CTD maintenance be performed?
      The manual recommends periodic maintenance at least once per month.

      Q5: What items are included in periodic CTD maintenance?
      Periodic maintenance includes trend calibration when necessary and using standard solutions to confirm instrument performance and accuracy.

      Q6: What can be used to clean CTD?
      Use a soft cloth or dedicated neutral cleaning agent to clean the analyzer surface, especially the sensing area, to prevent contaminants from affecting measurement accuracy.

      Q7: What CTD consumables should be checked?
      The manual notes that consumables requiring periodic replacement, such as electrodes, should be checked, and spare parts should be prepared for timely replacement.

      Q8: When should the electrode be replaced?
      Consider replacing the electrode when response becomes slow, readings are unstable, electrode slope decreases significantly, or cracks, breakage, or physical damage are found.

      Q9: Is calibration required after electrode replacement?
      Yes. The manual recommends performing calibration after electrode replacement to ensure optimal measurement accuracy.

      Q10: What items are recommended in the CTD maintenance checklist?
      The sample maintenance checklist includes housing condition, cable and connector appearance, analyzer accuracy, calibration status, and data collection testing.

  • L
    1. Troubleshooting
      Q1: What should be checked first when a PLC cannot read CTD values?
      First connect CTD to a PC using CAB06 and try reading the device with AQCFG to confirm that the CTD unit, power supply, and communication are normal.

      Q2: What are possible causes when a PLC cannot communicate with CTD?
      Common causes include incorrect communication settings, ID conflicts, incorrect PLC Modbus address settings, unstable power supply, or wiring errors.

      Q3: How should CTD power supply be verified?
      The CTD operating voltage range is 5-12 VDC. Although a computer 5 V supply can drive the analyzer in many cases, an external power supply is recommended for stable operation when multiple options are installed.

      Q4: What problems can wiring errors cause?
      Wiring errors may cause connection failure, no readings, or unstable communication. Check GND, RS485-, RS485+, power, and shield according to the wiring diagram.

      Q5: What should I do if CTD values fluctuate?
      First clean the sensing area and confirm that installation depth is appropriate, mounting is secure, and there is no sediment burial, bubble interference, or electrode damage.

      Q6: What should I do if CTD readings remain at 0?
      Confirm that the sensor is properly immersed, conductivity or water level actually changes, wiring and communication are normal, and check whether cleaning or recalibration is required.

      Q7: What should I do if CTD readings remain at the maximum range?
      Possible causes include sensor contamination, calibration error, mismatched measurement range, bubbles, sediment, or electrode damage. Clean first and confirm calibration status.

      Q8: What should I do if CTD readings are too high or too low?
      Check whether the sensing area is dirty, installation depth is correct, bubbles or sediment are present, and whether water level and conductivity calibration are correct.

      Q9: What should I do if CTD response time becomes slow?
      Possible causes include sensor surface contamination, electrode aging, sediment buildup, or poor installation environment. Clean the sensing area first and check electrode condition.

      Q10: What should I do if measurements differ for the same sample?
      Confirm that sample conditions are stable, the sensor is clean, there are no bubbles, the installation position is stable, and calibration is correct.

      Q11: What should I do if conductivity readings are abnormal or unchanged?
      You may first restore factory defaults, then perform the calibration procedure again. If temperature data fluctuates abnormally, the manual recommends replacing the electrode.

      Q12: What should I do if abnormal conditions persist after electrode replacement?
      If abnormalities persist after electrode replacement and recalibration, contact the AQUAS technical support team.

      Q13: When should AQUAS be contacted?
      If readings cannot be obtained or remain abnormal after checking power supply, wiring, Modbus ID, communication settings, AQCFG testing, cleaning, zero calibration, and slope calibration, contact AQUAS technical support.

  • A
    1. Product Basics
      Q1: What is ONE?
      ONE is AQUAS’s compact and cost-effective data logger, designed for applications such as water level, water quality, flow, rainfall, and manhole cover monitoring.

      Q2: What types of sites is ONE suitable for?
      ONE is suitable for groundwater, sewer systems, mining sites, underground aquifers, flooding or stormwater locations, municipal water supply, construction and mining applications, and meteorological monitoring environments.

      Q3: How many digital sensors can ONE connect to?
      ONE has a built-in RS485 serial port and can connect to up to 4 digital sensors.

      Q4: Does ONE require an external power supply?
      Not necessarily. ONE supports battery operation, and its ultra-low-power design allows long-term operation without external power. Depending on the model, it can also be configured with external power, an internal battery, an external battery, or solar power.

      Q5: How long can ONE operate?
      According to the manual, ONE can operate continuously for more than 5 years on battery power. With a primary lithium battery, some configurations can operate for up to 10 years. Actual operating time depends on the logging interval, upload frequency, communication method, number of sensors, I/O modules, and site conditions.

      Q6: Is ONE waterproof?
      Yes. ONE uses an integrated fully sealed design with a rugged IP68 submersible enclosure, allowing direct installation in groundwater monitoring wells, wet areas, flooded environments, or outdoor sites.

      Q7: Does ONE require a control cabinet?
      In most cases, no. ONE has a fully sealed IP68 enclosure and can be installed directly in outdoor or underground monitoring sites, reducing the need for control cabinets, surge protectors, and signal converters.

      Q8: Can ONE be used with AQWEB?
      Yes. ONE can upload data to the AQWEB cloud management platform, enabling remote display, control, alarm overview, trend charts, statistical data analysis, and report queries.

      Q9: Does ONE support alarm functions?
      Yes. ONE can set alarm conditions based on water level, water quality, flow, rainfall, manhole cover switches, or other sensor states. When an abnormal condition occurs, it can log the alarm and send notifications.

      Q10: Does ONE support SMS alarms?
      Yes. ONE can store up to 10 mobile phone numbers and can send SMS messages when an alarm occurs or when the alarm ends.

      Q11: Does ONE support GPS?
      GPS is an optional feature. Coordinates can be entered manually, or the GPS module can be used to search for coordinates and store the location record.

      Q12: Can ONE log data offline?
      Yes. ONE includes data logging functionality. If communication is unavailable, it can store data locally and automatically transmit the buffered data after communication is restored.

  • B
    1. Product Features and Applications
      Q1: What are the main advantages of ONE?
      ONE offers ultra-low power consumption, a rugged IP68 waterproof enclosure, multi-parameter monitoring, multiple communication options, cloud data upload, alarm notification, battery-powered operation, and quick installation.

      Q2: Is ONE suitable for groundwater monitoring?
      Yes. ONE can be installed directly in groundwater monitoring wells and used with water level, water quality, or other SMR sensors for long-term data logging and upload.

      Q3: Is ONE suitable for sewer or manhole monitoring?
      Yes. ONE can be used for sewer monitoring, manhole cover switch monitoring, water level, flow, and water quality monitoring, and can send alarms when abnormal conditions occur.

      Q4: Can ONE be used for flooding or stormwater monitoring?
      Yes. With its IP68 waterproof enclosure and low-power communication capability, ONE is suitable for flooded, stormwater, outdoor, or humid environments.

      Q5: Does ONE support multi-parameter monitoring?
      Yes. Through RS485 sensors, ONE can monitor different parameters such as water level, water quality, flow, rainfall, and manhole cover switch status.

      Q6: Can ONE be integrated with PLC or SCADA systems?
      Yes. ONE supports communication protocols such as Modbus TCP, FTP, MQTT, RESTful, and HART, and can integrate with monitoring systems through OPC Server or database interfaces.

      Q7: Does ONE support database integration?
      Yes. The manual states that data can be linked with databases such as SQL and MySQL, and can also be integrated through AQWEB or AQOPC.

      Q8: Does ONE support mobile or web operation?
      Yes. ONE can support data viewing and remote operation through cloud platforms, mobile phones, or mobile devices. Actual functions depend on the communication module and platform configuration.

  • C
    1. Hardware, Connections, and Installation
      Q1: What are the main ports on ONE?
      ONE mainly includes Port 0 for system communication, Port 1 for RS485 sensors, Port 2 for I/O, and connectors related to antennas or external power.

      Q2: What is Port 0 used for?
      Port 0 is the system communication port. It can connect to a PC via CAB06 for AQCFG software configuration, operating mode setup, alarm settings, control parameters, and data upload/download.

      Q3: What is Port 1 used for?
      Port 1 is the RS485 sensor port and can connect to SMR digital sensors or Modbus instruments.

      Q4: What is Port 2 used for?
      Port 2 is the I/O port and supports two DI channels for status input or counter input.

      Q5: What sensors can ONE connect to?
      ONE can connect to SMR sensors for water level, pH, ORP, conductivity, dissolved oxygen, turbidity, salinity, suspended solids, ammonia nitrogen, nitrate nitrogen, oil in water, chemical oxygen demand, and other parameters.

      Q6: What are the dimensions of ONE?
      The dimensional drawing in the manual shows that ONE is approximately 85 mm × 171 mm × 52 mm.

      Q7: What is the weight of ONE?
      The technical specifications in the manual list ONE at approximately 600 g.

      Q8: What is the enclosure material of ONE?
      The ONE enclosure is made of ABS.

      Q9: What installation methods does ONE support?
      ONE can be wall-mounted, pole-mounted, ground-mounted, or installed underground depending on site conditions.

      Q10: Can ONE be installed with a solar panel?
      Yes. ONE can be used with a solar panel and rechargeable lithium battery, making it suitable for sites without mains power.

      Q11: What should be noted when replacing or installing the SIM card?
      Remove the screw protection cap on the top of the unit, loosen the four screws, and open the top cover. Disconnect the internal battery cable first to prevent short circuits before installing the SIM card. After installation, make sure no cables are pinched and secure the top cover again.

      Q12: Can ONE be disassembled by the user?
      The manual provides disassembly steps for SIM card replacement. However, general internal inspection, modification, and repair should still be performed by AQUAS or authorized qualified specialists to prevent waterproofing failure or equipment damage.

  • D
    1. Power Supply and Battery
      Q1: What power supply options does ONE support?
      ONE can support external power, an internal primary battery, an internal rechargeable battery, an external battery box, a solar panel, and external battery configurations.

      Q2: What internal battery types are available for ONE?
      The manual lists BAT05-1 as a 3.6 VDC 13 Ah rechargeable lithium battery, and BAT10 as a 3.6 VDC 13 Ah primary lithium battery.

      Q3: What external battery options are available for ONE?
      The manual lists BAT11-2 as a 3.6 VDC 240 Ah rechargeable lithium battery box.

      Q4: How long can ONE operate with a primary battery?
      The manual specifications state that the BAT10 primary lithium battery can provide a service life of up to 10 years. Actual life should be evaluated based on logging and upload frequency.

      Q5: How long can ONE operate with a rechargeable battery?
      The manual lists the 3.6 VDC 13 Ah rechargeable lithium battery as having a life of approximately 300 charge/discharge cycles. When used with a solar panel and external battery, outdoor operating time can be extended.

      Q6: What should be done after replacing a primary battery?
      After replacing a primary battery, press “Reset” on the AQCFG system page so the system can correctly calculate the battery level.

      Q7: Why is it important to select the correct power mode?
      The power mode affects battery voltage, capacity, and remaining runtime estimation. Incorrect settings may cause inaccurate battery level display.

      Q8: Is there a low-battery alarm?
      Yes. ONE supports low-battery alarms. Conditions can be configured on the alarm page for logging or notification.

      Q9: What are the risks of using an incorrect battery?
      The manual warns that using an incorrect battery type or a non-original charger may create an explosion risk. Used batteries must be handled according to the manufacturer’s instructions.

  • E
    1. AQCFG Software Installation and Connection
      Q1: Which software is used to configure ONE?
      ONE uses AQUAS AQCFG software for device search, connection, system settings, sensor settings, communication settings, alarm settings, data export, and firmware updates.

      Q2: How is AQCFG installed?
      Insert the CD or USB drive, select the AQCFG folder, and click the Setup icon to start installation. Follow the prompts by clicking Next, then click Install to complete the installation.

      Q3: How does ONE connect to a computer?
      Use the CAB06 communication cable to connect ONE Port 0 to the PC, then open AQCFG to search for and configure the device.

      Q4: What must be installed before using CAB06?
      The USB driver must be installed first. Open the USB Driver folder and run CDM20824_Setup. After installation, a new Serial COM Port will appear in Device Manager.

      Q5: What should be configured before searching for ONE in AQCFG?
      Click “Settings” and select the COM Port, baud rate, and Modbus ID scan range, then close the settings page when complete.

      Q6: What is the default AQCFG baud rate for ONE?
      The manual’s quick operation and connection instructions indicate a default baud rate of 115200 bps.

      Q7: How do you enter the system after AQCFG finds ONE?
      Click “Link/Connect,” or double-click the ONE device to enter the system settings page.

      Q8: Does AQCFG automatically display the USB COM Port?
      Yes. According to the manual screen, the connected USB COM Port will appear automatically after the USB cable is connected.

      Q9: What should I do if AQCFG cannot find ONE?
      Check that CAB06 is connected to Port 0, the USB driver is installed, the COM Port is correct, the baud rate is set to 115200, the ID range includes the host ID, and the device is powered.

      Q10: What should be done first after entering host settings?
      Click the System ON button first. Settings and data logging can only be performed after the system is powered on.

  • F
    1. Host System Settings
      Q1: What can be configured on the system page?
      The system page can be used to configure or view the model, serial number, location, area, sub-area, system time, battery mode, battery voltage, battery level, communication rate, and Modbus ID.

      Q2: Can the model field be modified?
      No. The model is created during production and is used to verify the hardware configuration. It cannot be modified by the user.

      Q3: What is the serial number?
      The serial number is the unique device ID information for each ONE unit, used for device identification and management.

      Q4: What can be entered in the location field?
      The location field can be set by the user and can contain up to 20 letters, making it easier to manage and identify devices in AQWEB.

      Q5: What are the area and sub-area fields used for?
      The area and sub-area fields can be used for platform classification and management, helping distinguish different monitoring areas or stations.

      Q6: How is ONE system time synchronized?
      You can choose no synchronization, PC/Server synchronization, or GPS synchronization. After selecting PC/Server, set the time zone and press Confirm; the system time will synchronize with the server.

      Q7: Can GPS be used for time synchronization?
      Yes. If a GPS module is installed, the host time can be synchronized with GPS.

      Q8: How are settings saved after configuration?
      After all field data is configured, press “Confirm” again to save the data.

      Q9: Can the Modbus ID of ONE be configured?
      Yes. The system page can display and configure the current Modbus ID, making it easier for AQCFG to search and connect.

      Q10: What does System ON do?
      System ON indicates that the host system is activated. If System ON has not been performed, some settings and data logging functions may not operate correctly.

  • G
    1. 4G / GPRS / FTP Communication
      Q1: What communication methods does ONE support?
      Depending on the model, ONE can support GPRS, 4G, NB-IoT, Wi-Fi, Ethernet, LoRa, satellite communication, RS485, and Bluetooth.

      Q2: What data can ONE send back to the server?
      ONE can send initial configuration parameters, real-time values, alarms, logged data, and other information back to the server via GPRS, NB-IoT, or 4G.

      Q3: What should be done before configuring 4G or GPRS?
      Before configuration, check “Enable” and press Confirm. After completing the data settings, press Confirm again to save the data.

      Q4: What does the 4G status “Standby” mean?
      Standby means the module is powered off or in standby mode.

      Q5: What does “Try Connecting” mean?
      Standby → Try Connecting indicates that the device is establishing a connection.

      Q6: What does “Connected → Data Communication” mean?
      It means the connection is successful and data is being transmitted.

      Q7: How is communication signal strength evaluated?
      AQCFG displays both a 0%–100% value and a dBm value. The manual indicates that if the dBm value is greater than -93 dBm, the signal is too weak and the device may not be able to connect to the server.

      Q8: How should the APN be configured?
      The APN, username, and password should be entered according to the mobile operator’s network information. They can also be obtained by selecting the country and carrier from the network service provider menu on the right, or by contacting the carrier.

      Q9: What needs to be configured in CCS mode?
      In CCS mode, enter the server IP and server port. The manual lists the default server port as 502.

      Q10: What is the default AQUAS cloud server setting?
      The quick operation screen in the manual shows the default AQUAS cloud server as 118.163.120.143:502. For actual projects, confirm the setting with the shipment configuration or platform information.

      Q11: Can ONE be configured with a backup server?
      Yes. If ONE data needs to be sent to a second server at the same time, enable the backup server option and enter the IP and Port of the second server.

      Q12: What information is required for FTP mode?
      FTP mode requires the FTP server IP, Port, username, and password. The default FTP Port is 21.

      Q13: What is CCS / FTP mode?
      Select CCS / FTP mode when both the CCS server and FTP server need to receive data uploaded by ONE.

      Q14: How can communication be verified after configuration?
      Click “Connection Test” to confirm whether data can be uploaded to the server successfully.

  • H
    1. GPS and SMS
      Q1: Is GPS a standard feature on ONE?
      GPS is optional. If the GPS module is not installed, coordinates can be entered manually.

      Q2: What can GPS be used for?
      GPS can search for device coordinates, synchronize time, record location, and allow AQWEB or other platforms to display the monitoring station location.

      Q3: What does GPS status “off” mean?
      off means the GPS module is turned off.

      Q4: What does GPS status “Searching” mean?
      Searching means the GPS module is enabled and searching for satellite positioning.

      Q5: What does GPS status “Position” mean?
      Position means positioning has been completed.

      Q6: How is the GPS signal displayed?
      AQCFG displays the GPS module signal strength in a range from 0% to 100%.

      Q7: Can GPS be logged periodically?
      Yes. After “Periodic” is selected, ONE will record GPS coordinate values together with the logging cycle.

      Q8: When is GPS continuous mode suitable?
      Continuous mode keeps the GPS module enabled. The manual recommends using this mode when external power is available.

      Q9: What is GPS search mode?
      Search mode manually turns on the GPS function. After one search is completed, GPS will automatically turn off regardless of whether positioning succeeds.

      Q10: What is the minimum interval for GPS periodic mode?
      The manual states that the minimum interval for GPS periodic mode is 5 minutes.

      Q11: How many SMS phone numbers can ONE store?
      Up to 10 mobile phone numbers can be configured for SMS notifications.

      Q12: When can SMS messages be sent?
      For each phone number, SMS sending can be enabled for alarm occurrence and alarm end events.

      Q13: Why can’t the alarm end SMS option be selected directly?
      The manual notes that “Alarm Occurrence” must be selected before “Alarm End” can be selected.

      Q14: How can the SMS function be tested?
      Press “SMS Test,” and the host will send a test SMS message to the first mobile phone number.

  • I
    1. RS485 Sensors and I/O
      Q1: How does ONE connect to SMR sensors?
      Connect the SMR sensor to ONE Port 1 using CAB12 or an RS485 cable, then scan for the sensor on the AQCFG sensor port page.

      Q2: What is the default communication setting for ONE Port 1?
      The manual shows an example SMR sensor Port 1 setting of 19200 bps, N, 8, 1.

      Q3: How many SMR sensors can ONE scan?
      ONE can connect to up to 4 digital sensors. However, the AQCFG scan list may show Modbus IDs 1–8. The actual usable quantity should be confirmed based on the ONE model and hardware configuration.

      Q4: What restrictions apply to SMR sensor IDs?
      Each SMR sensor ID must be unique and within the host scan range.

      Q5: How is an SMR sensor configured after it is scanned?
      Click the sensor in the AQCFG tree list, or double-click the device in the sensor list to open that sensor’s settings page.

      Q6: Which SMR sensors are shown as examples for ONE?
      The manual screens show example sensors such as SMR04 pH, SMR05 ORP, SMR07 conductivity, and SMR09 dissolved oxygen.

      Q7: Can logging methods be configured for SMR sensors?
      Yes. Periodic logging and alarm logging can be configured on the sensor settings page.

      Q8: Can ONE calibrate SMR sensors?
      ONE can access calibration pages for SMR sensors, such as pH, ORP, conductivity, and dissolved oxygen. Detailed standard solutions and calibration procedures should still follow the corresponding SMR operation manual.

      Q9: Does ONE support I/O modules?
      Yes. The I/O module on ONE Port 2 supports two DI channels for pulse counting or status input.

      Q10: What applications can DI be used for?
      DI can be used for manhole cover switches, rainfall pulses, flow pulses, equipment status, or other digital input signals.

  • J
    1. Operation Settings, Alarms, and Data Logging
      Q1: What can be configured in operation settings?
      Operation settings can configure the system operating mode, scan cycle, logging cycle, alarm logging cycle, communication module, upload schedule, periodic upload, and retry count.

      Q2: Does ONE support continuous operation?
      Yes. After “Continuous” is selected, the system can continuously scan, log, and upload data according to the configured cycles.

      Q3: Does ONE support scheduled operation?
      Yes. A schedule can be configured so the system operates or uploads data only at specified times.

      Q4: What is the scan cycle?
      The scan cycle is the time interval at which ONE scans sensor or I/O values and determines status and alarm conditions.

      Q5: What is the logging cycle?
      The logging cycle is the time interval at which ONE writes sensor or I/O data to the log records.

      Q6: What is the alarm logging cycle?
      The alarm logging cycle is the time interval for accelerated data logging during an alarm or under alarm level conditions.

      Q7: Can data upload be scheduled periodically?
      Yes. Periodic upload can be configured, with daily 00:00 used as the starting time for data transmission according to the settings.

      Q8: Can data upload be scheduled?
      Yes. If schedule mode is selected, the system will send data back to the server only at the selected times.

      Q9: Can data be uploaded immediately when an alarm occurs?
      Yes. If immediate alarm upload is enabled, the host will send data back to the server immediately when an alarm record occurs.

      Q10: What alarm types does ONE support?
      ONE supports system alarms, sensor alarms, and I/O module alarms, such as Wi-Fi communication failure, mobile communication failure, low voltage, and upper/lower limits for sensor values.

      Q11: What alarm levels can be configured?
      Alarm levels such as HH, H, L, and LL can be configured, along with the corresponding values and activation conditions.

      Q12: Can ONE export data?
      Yes. Records and alarm data can be searched and exported from the AQCFG Import/Export page.

      Q13: Can ONE be restored to factory defaults?
      Yes. The AQCFG Import/Export page provides a factory reset function. After the reset, reconnect the device and confirm the settings.

  • K
    1. Firmware Update and Maintenance
      Q1: Can ONE firmware be updated?
      Yes. The AQCFG system firmware page can load the correct version of the .aqs file to perform a firmware update.

      Q2: What should be noted before a firmware update?
      Before updating, make sure only one device is connected to avoid upload failure or unrecoverable device errors.

      Q3: Can the firmware update process be interrupted?
      No. Once the firmware update starts, it must not be interrupted. Wait until the update is complete.

      Q4: What should be done after the firmware update is complete?
      The manual states that after the firmware update is complete, go to the settings page under Import/Export and click “Restore” to restore ONE to factory settings. After reconnecting and powering on the device, go to the records page and delete data to ensure normal host operation.

      Q5: Does ONE require regular maintenance?
      Yes. Regularly check the enclosure, waterproof condition, connectors, battery, communication signal, sensor connections, data logging status, and data upload status.

      Q6: What should be done if the ONE cable is damaged?
      If a sensor or device cable is damaged, contact AQUAS Customer Service. Cutting, folding, damaging, or replacing the power cable by yourself is not recommended.

      Q7: Can ONE be modified by the user?
      No. Unless explicitly stated in the manual, do not disassemble or modify the product or its components by yourself.

      Q8: What should be noted when using ONE outdoors?
      Make sure the top cover is tightened, waterproof connectors are secured, cables are fixed, the antenna is properly positioned, the battery level is sufficient, the communication signal is stable, and platform upload status is normal.

      Q9: What should be noted when installing ONE underground or inside a well?
      Ensure the device is securely fixed, avoid long-term cable tension, and confirm that the antenna or communication module can maintain sufficient signal strength.

  • L
    1. Common Troubleshooting
      Q1: What should I do if ONE cannot connect to the computer?
      Check whether CAB06 is connected to Port 0, the USB driver is installed, Device Manager shows the Serial COM Port, the correct COM Port is selected, and the baud rate is correct.

      Q2: What should I do if AQCFG cannot scan ONE?
      Check that ONE is powered, the COM Port is correct, the baud rate is set to 115200, and the ID range includes the device ID, then scan again.

      Q3: What should I do if ONE cannot upload data?
      Check whether the communication module is enabled, the APN is correct, the SIM card is available, and the server IP and Port are correct, then run a connection test.

      Q4: What should I do if the 4G signal is too weak?
      If the dBm value is greater than -93 dBm, the signal is too weak. Adjust the antenna position or installation location, or confirm mobile network coverage.

      Q5: What should I do if FTP upload fails?
      Confirm that the FTP server IP, Port 21, username, and password are correct, then click Connection Test.

      Q6: What should I do if SMS messages are not received?
      Confirm that SMS is enabled, the phone number is correct, the SIM card supports SMS, and the alarm occurrence and alarm end options are selected correctly, then use SMS Test first.

      Q7: What should I do if GPS cannot obtain a position?
      Confirm whether a GPS module is installed, GPS is enabled, the site has no obstruction, and signal strength is sufficient. If GPS is not installed, enter the coordinates manually.

      Q8: What should I do if sensors cannot be scanned?
      Confirm that the SMR sensor is connected to Port 1, the RS485 wiring is correct, the communication setting is 19200.N81 or matches the sensor setting, and each sensor ID is unique.

      Q9: What should I do if data is not being logged?
      Check System ON, the logging cycle in operation settings, whether the sensor is enabled, and whether periodic or alarm logging is selected.

      Q10: What should I do if alarms are not triggered?
      Check whether the alarm page is enabled, alarm levels and threshold values are configured, and alarm logging is selected for the sensor or I/O.

      Q11: What should I do if the battery level display is inaccurate?
      Confirm that the power mode and battery type are correct. If a primary battery has been replaced, press “Reset” to recalculate the battery level.

      Q12: What should I do if the firmware update fails?
      Confirm that only one device is connected, the firmware file version is correct, and power and communication remain stable during the update. If it still fails, contact AQUAS Technical Support.