    1. Product General Questions
      Q1: About product?
      It is a multi-parameter data logger designed primarily for environmental monitoring, water-quality monitoring, and data acquisition from pressure, flow, water-level, and various sensor devices.

      Q2: What devices can it connect to?
      It can connect via RS485 to AQUAS SMR digital water-quality analyzers, MAX multi-parameter water-quality probes, pressure modules, I/O modules, and other sensors that support Modbus RTU.

      Q3: What applications is it suitable for?
      It is suitable for drinking water, pipelines, underground utility networks, hydrology, meteorology, oil and gas, petrochemical facilities, wastewater, rivers, reservoirs, and remote environmental monitoring.

      Q4: Can it be used outdoors?
      Yes. ECO features an integrated sealed design with an IP69K / IP68 waterproof enclosure, making it suitable for humid, submerged, outdoor, or flood-prone environments.

      Q5: Is an external control cabinet required?
      Not necessarily. ECO is an integrated data logger with built-in communication, data logging, and sensor connection functions. It can be installed directly according to site requirements.

      Q6: What communication methods are supported?
      ECO supports communication options such as 4G, GPRS, NB-IoT, Wi-Fi, Ethernet, LoRa, satellite communication, GPS, RS485, and Bluetooth. Actual functions depend on the ordered model.

      Q7: Does it support MQTT?
      Yes. This version of the manual includes MQTT mode settings for the NB-IoT module, allowing data to be uploaded to a specified server via the MQTT protocol.

      Q8: Does it support AQWEB?
      Yes. ECO can be used with the AQWEB cloud management platform for data collection, remote monitoring, alarms, control, chart analysis, and data management.

      Q9: Can it operate independently for long periods?
      Yes. ECO is designed for low power consumption and can be used with an internal battery, external battery, or solar panel for long-term unattended monitoring.

      Q10: Does it have GPS functionality?
      GPS is available as an option or supported depending on the model. GPS can be used for location information, time synchronization, and location-change alarms.

      Q11: Does it have explosion-proof certification?
      IECEx / ATEX Ex ia IIB T5 Ga explosion-proof certification is supported. Actual certification should be confirmed according to the shipped model.

      Q12: Can it upload data to two servers simultaneously?
      Yes. ECO supports simultaneous upload of log and alarm files to two data servers for redundancy or dual data-center architectures.

  • A
    1. Installation and Hardware Questions
      Q1: What field installation methods are available?
      It can be used in wall-mounted, pole-mounted, sewer manhole water-level monitoring, sewer flow-velocity and water-level monitoring, open-water, and pressure-pipeline installation scenarios.

      Q2: What water-level or flow devices can be used with 4G?
      It can be used with the SMR61 radar level meter for sewer manhole water-level monitoring, or with the SMR63 radar velocity and level meter for sewer flow and water-level monitoring.

      Q3: How should the 4G antenna be installed?
      First locate the SMA metal connector on the top of the ECO, check that the ANT-series antenna connector is intact, align the antenna with the SMA connector, and tighten it clockwise. The antenna must remain vertical and secure, without tilting or looseness.

      Q4: What precautions should be taken when installing the antenna?
      Avoid installing the antenna in humid environments to prevent oxidation of the SMA connector. After installation, avoid bending, squeezing, or applying external force to prevent damage to the antenna or wireless module.

      Q5: How do I install a Micro SIM card?
      First make sure the device is powered off and fully disconnected from power. Use a Phillips screwdriver to remove the metal cover marked SIM, insert the activated Micro SIM card into the slot in the correct orientation, and after hearing it click into place, reinstall and tighten the metal cover.

      Q6: Does the SIM card need to be activated first?
      Yes. Before using 4G / GPRS / NB-IoT, the SIM card must be activated, and the data plan should be confirmed to meet the site application requirements.

      Q7: Will 4G communication generate communication fees?
      Yes. Enabling 4G communication will generate data transmission fees. Rates depend on the SIM card and telecom provider plan. Data fees and tariff disputes are the contractual responsibility of the user and the telecom provider.

      Q8: Can non-original batteries be used?
      This is not recommended. The manual prohibits the use of non-original batteries or chargers, as they may cause overheating, fire, or explosion, and may also damage the device or void the warranty.

      Q9: When should the internal battery be replaced?
      If the system detects that the battery voltage is below 3.4 V, it is recommended to replace the battery immediately to maintain stable operation and data integrity.

      Q10: What should be noted when replacing the internal battery?
      Use a Phillips screwdriver to remove the four screws around the LCD and carefully lift the upper cover to avoid damaging the display ribbon cable or internal connectors. After replacement, reinstall the upper cover and restart the device for confirmation.

      Q11: Can solar power be used?
      Yes. ECO can be used with PAN-series solar panels or external batteries, making it suitable for remote monitoring or locations without utility power.

      Q12: What should be noted when inserting connectors?
      Before insertion, confirm that the keyway and pin orientation are correct. Do not insert at an angle, force the connector, or rotate it during insertion, as this may bend pins, cause poor contact, or damage the enclosure.

      Q13: What precautions should be taken when tightening connectors?
      After the connector is fully inserted, tighten the threaded locking ring clockwise to prevent loosening, water ingress, dust ingress, or vibration-related failure. Do not overtighten, as this may damage the threads.

      Q14: How often should connectors be inspected?
      It is recommended to inspect the connector locking condition every 1 to 3 months, especially in outdoor, high-humidity, or high-vibration environments.
  • B
    1. AQCFG Connection and Configuration Questions
      Q1: What is Port 0 used for?
      It is mainly used for system communication. It can connect to a computer and AQCFG software for system configuration, diagnostics, data upload and download, and may also support PLC / DCS / HMI communication depending on the model.

      Q2: What is Port 1 used for?
      Port 1 is the sensor connection port. It can connect to AQUAS SMR digital sensors, MAX multi-parameter water-quality probes, I/O modules, and other devices, and also supplies power to the sensors.

      Q3: What is Port 2 used for?
      Port 2 can be used as an I/O port, supporting digital input, relay output, pulse output, or analog input. Actual functions depend on the I/O module model.

      Q4: What is Port 3 used for?
      Port 3 can connect to a second sensor set or optional pressure modules, barometric pressure modules, and similar devices for pressure, water-level, or compensation applications.

      Q5: How many sensors can be connected?
      Multiple SMR digital sensors and I/O modules can be connected. Some architectures support up to 8 sets of SMR sensors and I/O modules.

      Q6: What cable should be used to connect to a PC?
      It is recommended to use the CAB06 USB to RS485 Cable to connect ECO to a computer, then configure and read data through AQCFG software.

      Q7: What cable should be used to connect to a PLC?
      For PLC connection, the CAB03-A communication cable is recommended.

      Q8: Can an RS485 Hub be used?
      Yes. AQUAS BOX01 and BOX02 can be used as RS485 Hubs for expanded connection of multiple analyzers, improving signal stability and reducing transmission loss.

      Q9: What power supply devices can be used?
      It can be used with power supply devices such as BAT11-2, BAT03-2, PAN02, and ADP02, using batteries, solar panels, or power adapters according to site requirements.

      Q10: What should be noted when removing connectors?
      When removing a connector, first loosen the locking nut counterclockwise, then pull the connector straight out. Do not twist the cable or apply torque to the connector, as this may loosen or damage internal wiring.

  • C
    1. System Time and Battery Configuration Questions
      Q1: What software should be used for configuration?
      Use the AQCFG configuration software provided by AQUAS to perform system configuration, calibration, data reading, communication testing, firmware updates, and diagnostics.

      Q2: How do I install AQCFG?
      After inserting the installation disc or obtaining the installer, run Setup, select the installation path, Start Menu folder, and whether to create a desktop shortcut, then click Install to complete the installation.

      Q3: How do I install the USB Driver?
      After connecting ECO to the computer using CAB06, install CDM20824_Setup from the USB Driver folder. After installation, USB Serial Port will appear in Device Manager.

      Q4: What should be done before connecting to AQCFG?
      First connect ECO to the PC using CAB06, then open AQCFG software and configure COM Port, Modbus ID, Baud Rate, and Parity in the Device Search interface.

      Q5: What are the default connection settings for the ECO Data Logger?
      Common settings are as follows:
      Modbus ID:1
      Baud Rate:115200
      Parity:None
      Data Bits:8
      Stop Bits:1

      Q6: How do I scan for the ECO device?
      Click Scan on the Device Search page. The system will detect connected devices. After the device is found, select Link or double-click the device name to enter Basic Settings.

      Q7: What do the green and red communication statuses in AQCFG mean?
      Green indicates connected; red indicates not connected or disconnected.

      Q8: Why do I need to click System On?
      Before parameter configuration and data logging, ECO must be started by clicking System On. If the system is not started, settings cannot be applied and data logging will not operate.

      Q9: How do I confirm that System On was successful?
      After clicking the green System On icon, if the fields become editable, the system has been initialized successfully and configuration can begin.

      Q10: How can the system time be synchronized?
      You can choose no synchronization, synchronization with PC / Server, or GPS synchronization. If PC / Server is selected, Time Zone must be configured.

      Q11: Is software configuration required after replacing the battery?
      Yes. After replacing the battery, click Reset in the Battery section of AQCFG so the system can recalculate battery capacity.

      Q12: How do I save settings after making changes?
      After making changes, click Apply to write the settings to the device.

  • D
    1. Sensor and Data Logging Questions
      Q1: How do I make ECO scan SMR analyzers?
      Configure the sensor communication parameters in AQCFG. Common SMR analyzer settings are 19200 bps, 8 data bits, 1 stop bit, and Parity None. Then click Scan to automatically search for connected modules and sensors.

      Q2: Why can’t the sensor be detected?
      Possible causes include damaged cables, incorrect COM Port, incorrect Baud Rate, incorrect Modbus ID, or duplicate IDs among multiple sensors.

      Q3: Can the data logging interval of ECO be configured?
      Yes. ECO supports scheduled logging, and the interval can be adjusted according to the application. The manual specifications list a data acquisition interval configurable from 1 second to 60 minutes.

      Q4: Can ECO record historical data?
      Yes. ECO can store a large volume of data records, and Historical Logs, Alarms, and Trend Charts can be viewed through AQCFG.

      Q5: Can ECO data be exported?
      Yes. Go to the Import / Export page in AQCFG and click Export to export records in CSV format for analysis or reporting.

      Q6: Does ECO support SD card storage?
      Yes. The manual specifications list support for storage formats including SD, SDHC, SDXC, MicroSD, MicroSDHC, and MicroSDXC.

      Q7: Can ECO prevent data loss?
      ECO includes a store-and-forward mechanism. If an upload is incomplete, data can be automatically saved and resent, reducing the risk of data loss.

      Q8: What data upload methods does ECO support?
      ECO can upload data via CCS, FTP, CCS / FTP dual mode, MQTT, and other methods, depending on the communication module and configuration.

      Q9: Can ECO upload alarm data?
      Yes. ECO supports alarm communication, can upload alarm logs, and can send SMS or other alarm notifications according to the configuration.

      Q10: Can ECO display real-time data?
      Yes. ECO can display real-time data, historical trends, alarms, and data analysis through AQCFG or AQWEB.
  • F
    1. 4G / FTP / CCS Communication Questions
      Q1: How do I enable 4G communication on ECO?
      On the 4G settings page in AQCFG, select Enable, click Apply to enable the module, and enter information provided by the telecom provider, such as SIM PIN, APN, username, and password.

      Q2: How is 4G signal strength evaluated?
      AQCFG displays signal quality as a percentage and in dBm. The manual notes that if dBm > -93, the signal is too weak and connection may fail.

      Q3: What are the 4G connection statuses?
      Common statuses include:
      Standby: Idle
      Standby → Try Connecting: Attempting to connect
      Standby → Connected → Data Communication: Connected and transmitting data

      Q4: What should be entered for APN?
      The APN should be entered according to the settings provided by the telecom provider of the SIM card. For MQTT mode, the manual example mentions that JTM2M can be used as the APN.

      Q5: What information is required for CCS mode?
      Select CCS or CCS / FTP mode, enter the Server IP and Port. The commonly used default communication port is 502, and a backup server can also be configured.

      Q6: What information is required for FTP mode?
      Enter FTP Server IP, Port, Username, and Password. The default FTP communication port is 21. After setup, click Test Connection to test the connection.

      Q7: Can ECO upload to CCS and FTP simultaneously?
      Yes. To upload to CCS and FTP simultaneously, select CCS / FTP dual mode.

      Q8: What should be checked first when 4G or FTP cannot connect?
      First check whether the SIM card is activated, the APN is correct, the antenna is properly connected, the signal is sufficient, the server IP and Port are correct, and whether the firewall is blocking the communication port.

      Q9: What should I do if 4G communication is unstable?
      Install the device in a location with stronger signal, regularly confirm the SIM card validity period, data usage, and tariff status, and check whether the antenna is loose or damaged.

      Q10: Can ECO use a backup server?
      Yes. In CCS settings, Second Server IP can be enabled to upload data simultaneously to a second server, improving communication redundancy.
  • F
    1. MQTT / NB-IoT Questions
      Q1: What is ECO’s MQTT function used for?
      MQTT mode is mainly used for the NB-IoT module, allowing ECO to upload data to a specified server or domain via the MQTT protocol.

      Q2: What should be configured first for MQTT mode?
      First configure fields such as NB-IoT Band, APN, Username, Password, Protocol, Domain Name, and Server Port.

      Q3: How should the NB-IoT Band be selected?
      Select the band according to the country of use or telecom provider frequency band. The manual default is Band 8 and recommends referring to the NB-IoT Frequency Band Reference Table.

      Q4: What should be entered for the MQTT APN?
      The APN should be entered according to the information provided by the telecom provider. The manual example mentions that JTM2M can be used for MQTT mode.

      Q5: Can a domain name be used for the MQTT Server?
      Yes. The manual example uses a DNS domain, such as 502136.aquas.com.tw, and the system automatically resolves the IP address.

      Q6: What is the default MQTT communication port?
      The default MQTT Server Port is 1883.

      Q7: How do I test after completing MQTT configuration?
      Click Test Connection to confirm whether the device can successfully upload data to the server.

      Q8: Why can’t it connect after MQTT configuration?
      Possible causes include an inactive NB-IoT SIM, incorrect APN, incorrect Band setting, weak signal, Server Port blocked by a firewall, or incorrect DNS / server settings.

      Q9: How can NB-IoT connection failure be troubleshot?
      Temporarily switch the Band setting to Band 3, then switch back to the appropriate Band. Also confirm that the SIM card has NB-IoT service enabled, the APN is correct, and the server communication port is not blocked.

      Q10: Is data usage a concern in MQTT mode?
      Yes. Although NB-IoT data volume is typically small, the SIM card validity period, data plan, and usage should still be checked regularly.

  • F
    1. Alarm and Control Questions
      Q1: Does it support alarm functions?
      Yes. ECO can configure 4-level alarms, such as HH, H, L, and LL, and also supports alarm actions such as SMS, alarm log, and control output.

      Q2: What alarms can be detected?
      Alarms can be generated based on sensor values, I/O status, digital inputs, level switches, manhole cover switches, installation angle changes, GPS location changes, low battery, and other conditions.

      Q3: Can it perform control output?
      Yes. Depending on the I/O module, ECO can support functions such as relay output, pulse output, manual ON/OFF, and IF_THEN_ELSE conditional control.

      Q4: Does it support Peer-to-Peer control?
      Yes. It can support Peer-to-Peer communication between sensors and I/O for alarm and control conditions on remote controllers.

      Q5: Can transient pressure be detected?
      This is available as an option. ECO supports transient pressure detection for events such as water hammer, pipeline rupture, pump abnormalities, or valve abnormalities.

      Q6: What is the transient pressure detection speed?
      Transient pressure detection can reach 0.001 seconds.

      Q7: Can device location changes be detected?
      Yes. ECO supports GPS location offset anti-theft alarms and location-change alarms. Actual availability depends on the model and configuration.

      Q8: Does it provide low-battery alerts?
      Yes. ECO has intelligent battery management that can estimate battery operating days and provide early low-battery warnings to prevent unexpected power loss.


  • F
    1. Firmware Update and Maintenance Questions
      Q1: Can ECO firmware be updated?
      Yes. AQCFG provides the Firmware Information & Update function, allowing .aqs firmware files to be loaded and updates to be performed.

      Q2: How do I check the ECO firmware version?
      The currently installed firmware version and release date can be viewed on the Firmware page in AQCFG.

      Q3: What are the firmware update steps?
      On the Firmware Update page, click Open to load the correct .aqs firmware file, then click Download to start the update. When a confirmation message appears, click YES to continue or NO to cancel.

      Q4: What should be noted during firmware updates?
      During the update, the PC should be connected to only one target device to avoid interference. The update process must not be interrupted, and stable power and communication must be ensured to prevent device failure.

      Q5: Is regular maintenance required?
      Yes. It is recommended to regularly inspect connector locking, antenna condition, SIM card status, battery voltage, enclosure sealing, communication status, and data upload status.

      Q6: Can the enclosure be opened for repair?
      Self-disassembly is not recommended. The manual explicitly prohibits unauthorized disassembly of the enclosure, as it may compromise waterproofing and cause water ingress and device damage.

      Q7: Can the circuit board or wiring be disassembled by the user?
      No. Unauthorized removal of circuit boards or wiring may cause short circuits or component damage.

      Q8: Can it be used in extreme environments?
      Use outside the specified limits is not recommended. The manual prohibits exposing the product to extreme high or low temperatures, high pressure, corrosion, radiation, strong magnetic fields, or severe vibration.

      Q9: Will impact to the enclosure affect the device?
      It may. The manual prohibits squeezing, twisting, dropping, impact, or collision, as these external forces may damage the enclosure, connectors, internal modules, or waterproof performance.

      Q10: Who should perform repairs?
      Internal inspection, modification, repair, or cable replacement should be handled by AQUAS or qualified experts authorized by AQUAS.

  • F
    1. Common Troubleshooting Questions
      Q1: What should I do if the device cannot connect to the computer?
      Confirm that CAB06 is correctly connected, the USB Driver is successfully installed, USB Serial Port appears in Device Manager, the correct COM Port is selected, and the Baud Rate is set to 115200.

      Q2: What should I do if AQCFG cannot detect ECO?
      Confirm that ECO is powered, the COM Port is correct, Modbus ID is 1, Baud Rate is 115200, Parity is None, Data Bits is 8, Stop Bits is 1, and then scan again.

      Q3: What should I do if the SMR analyzer cannot be detected?
      Confirm that the analyzer is connected to Port 1, communication settings are 19200, 8N1, Parity None, and that each analyzer has a unique Modbus ID.

      Q4: What should I do if there is no 4G signal?
      Check whether the SIM card is activated and correctly inserted, whether the APN is correct, the antenna is tightened, the signal is sufficient, and the SIM card data plan is valid.

      Q5: What should I do if data cannot be uploaded to AQWEB?
      Confirm that 4G / Wi-Fi / Ethernet communication is enabled, the server IP or DNS is correct, the Port is correct, Test Connection is successful, and the firewall is not blocking communication.

      Q6: What should I do if MQTT upload fails?
      Confirm the NB-IoT Band, APN, DNS, Server Port 1883, SIM NB-IoT service, signal strength, and firewall settings.

      Q7: What should I do if data logging is not operating?
      Confirm whether System On has been clicked. If the system is not started, settings cannot be applied and data logging will not operate.

      Q8: What should I do if data export fails?
      Confirm that the device is connected, data records exist, the AQCFG Import / Export page is operated correctly, and the computer has permission to write to the selected storage location.

      Q9: What should I do if the device battery drains quickly?
      Check the data logging interval, upload frequency, alarm frequency, communication module usage, number of sensors, and site temperature. Frequent uploads and multiple sensors will increase power consumption.

      Q10: What should I do if connectors have water ingress or communication is unstable?
      Check whether the connector is correctly inserted, the threaded locking ring is tightened, and the O-ring or connector is damaged. Avoid installing or removing antennas or connectors in humid environments.