SMR45 TOC Analyzer
Frequently Asked Questions
-
Q1
Basic Product Questions
A- Basic Product Questions
Q1: What is this product?
It is an AQUAS Total Organic Carbon (TOC) analyzer designed for real-time monitoring of TOC concentration in water, with data output via RS485 digital communication.
Q2: What parameters can it measure?
It primarily measures Total Organic Carbon (TOC) and also provides suspended solids measurement with automatic interference compensation to improve the stability of field water-quality monitoring.
Q3: What is the measurement principle?
It uses 254 nm ultraviolet light to measure the absorbance of organic compounds in water that contain conjugated double bonds or polycyclic aromatic hydrocarbons, thereby estimating Total Organic Carbon (TOC). At the same time, a 530 nm optical beam detects suspended solids concentration for automatic compensation.
Q4: Are chemical reagents required?
No. It uses optical measurement and enables real-time monitoring without adding chemical reagents.
Q5: What applications is it suitable for?
It is suitable for drinking water, surface water, industrial processes, water treatment, wastewater, and related applications.
Q6: Is a control box required?
Not necessarily. The analyzer features an integrated miniature housing with a built-in transmitter and sensor. Its rugged IP68 submersible, corrosion-resistant enclosure allows direct field installation.
Q7: Does it support PLC or HMI integration?
Yes. It uses the standard Modbus RTU communication protocol and can be connected to a PLC, HMI, or AQUAS data logger via RS485.
Q8: Is it suitable for outdoor sites without mains power?
Yes. With its ultra-low-power design, it can be powered by batteries or solar power for outdoor or remote monitoring applications.
Q9: Does it have explosion-proof certification?
Yes. It is certified to IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS explosion-proof standards.
Q10: What is the maximum transmission distance?
Using a 1000N high-strength Kevlar cable, the maximum communication distance can reach up to 1.2 km.
Q11: Does it provide self-diagnostic functions?
Yes. It includes built-in self-status monitoring, such as temperature and humidity monitoring, as well as LED light-source abnormality diagnostics.
Q12: Does it provide automatic compensation?
Yes. Built-in reference-light measurement enables automatic compensation for light-source aging and optical-window fouling, and the analyzer also provides automatic compensation for turbidity and suspended solids interference.
Q13: Does it support automatic cleaning?
Yes. It can be equipped with an automatic cleaning brush, as well as water or air automatic cleaning devices, to reduce maintenance caused by optical-window fouling, bubbles, or deposits.
Q14: What are the features of the optical window material?
The optical window is made of abrasion-resistant, scratch-resistant sapphire glass, making it less prone to wear during cleaning and maintenance. It also includes a nano-coating to reduce sludge deposition and biofouling.
- Basic Product Questions
-
Q2
Installation Questions
B- Installation Questions
Q1: What is the operating voltage?
The operating voltage is 5-12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.
Q2: What installation methods are available?
Immersion, flow-through, or pipe-insertion installation can be used.
Q3: What should be noted during installation?
Avoid pulling or damaging the cable during installation. Ensure the cable length is sufficient and properly secured. Installation in a location with stable water flow is recommended for better measurement performance.
Q4: How deep should the sensing surface be immersed?
During testing, the sensing surface must be fully immersed at least 10 cm below the water surface.
Q5: How long does it take to stabilize after installation?
After installation, allow at least 5 minutes for the analyzer to reach a stable condition. The actual stabilization time may vary depending on sample characteristics and the installation environment.
Q6: Can it be installed close to the bottom of the tank?
It is not recommended. The analyzer should not directly contact or be installed too close to the tank bottom, as obstruction of the sensing area or sediment interference may cause abnormal measurements.
Q7: How should the sensing area be protected during field installation?
If there is a risk of bottom contact, impact, sediment, or water-flow shock on site, install a dedicated protective cover or use mounting fixtures to prevent the sensing area from being obstructed.
Q8: Can it be installed in a flow-through holder?
Yes. The manual installation diagrams include the FTC45-1 flow-through holder and PIF45-1 simple pipe-insertion holder, which can be selected according to on-site pipeline or flow-cell requirements.
Q9: Can measurement begin immediately after installation?
Yes. The analyzer is calibrated before shipment. For first use, measurement can begin after proper installation according to the manual. If accuracy confirmation is required before use, standard-solution calibration can be performed.
Q10: Can it be connected to an AQUAS data logger?
Yes. It can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data recording and remote monitoring according to site requirements.
```
- Installation Questions
-
Q3
Communication and Wiring Questions
C- Communication and Wiring Questions
Q1: What communication interface is used?
It uses an RS485 communication interface.
Q2: What communication protocol is used?
It uses the Modbus RTU communication protocol.
Q3: What are the default communication settings?
Common communication settings are as follows:
Communication interface: RS485
Baud rate: 19200 bps
Data length: 8 bit
Parity: None
Stop bit: 1 bit
Communication mode: RTU
The device supports 9600 and 19200 bps.
Q4: What is the default device address?
The factory default device address is 01. If the device address is changed, the address configured in the host, PLC, HMI, or reading software must also be updated accordingly.
Q5: What is the RS485 wiring configuration?
The SMR male connector wiring is as follows:
PIN 1: GND
PIN 2: RS485-
PIN 3: RS485+
PIN 4: 5-12V DC
PIN 5: Not used
The typical new cable colors are: black for GND, white for RS485-, blue for RS485+, and red for 5-12V DC.
Q6: If AO analog output is available, what is PIN 5?
For the AO version, PIN 5 is the AO analog output wire, typically gray. For the standard non-AO version, PIN 5 is not used.
Q7: Can it be connected directly to a computer?
Yes. It can be connected to a computer using the CAB06 USB-to-RS485 cable, allowing values to be read, calibration to be performed, parameters to be configured, and diagnostics to be conducted through Modbus or utility software.
Q8: What should be checked first if the PLC cannot read values?
Check the following in order:
1. Whether the device can be connected to a PC using CAB06 and read through Modbus
2. Whether the PLC and SMR45 communication settings are consistent
3. Whether the Modbus ID is correct and whether there is an ID conflict
4. Whether the PLC Modbus address is correct
5. Whether the power supply is stable at 5-12 VDC
6. Whether the RS485 wiring is correct
7. Whether wiring has been completed according to the appendix wiring diagram
Q9: What happens if RS485 D+ / D- are reversed?
If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
Q10: Can multiple devices be connected in series?
Yes. Using RS485 digital communication, multiple devices can be connected in a loop to reduce wiring and engineering costs. When multiple devices are connected, ensure that each device has a unique Modbus ID with no conflicts.
- Communication and Wiring Questions
-
Q4
Modbus Reading Questions
D- Modbus Reading Questions
Q1: Which address should be read for the main TOC parameter value?
The main TOC parameter value can be read from PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with the data type as floating point.
Q2: Which address should be read for the secondary parameter value?
The secondary parameter value can be read from PLC address 2, HEX address 0x0001, Length 2 Words, Function Code 04, with the data type as floating point.
Q3: Which address should be read for the real-time temperature value?
The real-time temperature value can be read from PLC address 3, HEX address 0x0002, Length 2 Words, Function Code 04, with the data type as floating point.
Q4: How can the device model be read?
Read PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with the data type as an ASCII string.
Q5: How can the location, area, or sub-area name be read?
Read PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with the data type as a UTF-16BE string.
Q6: How can latitude and longitude be read?
Read PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type as floating point.
Q7: How can the cleaning brush be activated?
If the device is equipped with a cleaning brush, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.
Q8: How should floating-point data be decoded?
Floating-point data follows the IEEE 754 single-precision format and occupies 4 Bytes. If the decoded value is abnormal, verify whether the byte order or word order needs to be swapped.
Q9: What should be noted about Modbus CRC?
Modbus RTU uses CRC16. During transmission, the CRC is sent Low Byte First and High Byte Last, meaning the low byte is transmitted before the high byte.
Q10: What do Function Codes 03, 04, and 10 represent?
Function Code 03: Read holding registers
Function Code 04: Read input registers
Function Code 10: Write multiple holding registers
Q11: Why do garbled or unreasonable values appear when reading data?
Common causes include inconsistent floating-point byte or word order, incorrect PLC address settings, incorrect Function Code selection, mismatched communication parameters, or incorrect data parsing format. Confirm the Modbus register table and communication format first.
- Modbus Reading Questions
-
Q5
Calibration Questions
E- Calibration Questions
Q1: Is calibration required before first use?
Generally, no. The analyzer is calibrated before shipment. For first use, measurement can begin after proper installation according to the manual. If accuracy confirmation is required before use, standard-solution calibration can be performed.
Q2: What calibration items are available?
Calibration items include temperature calibration, standard zero calibration, standard slope calibration, field calibration, background subtraction, and AO calibration for the optional AO version.
Q3: What should be done before calibration?
Before calibration, clean the entire analyzer and all calibration-related vessels with detergent and a soft brush to prevent calibration abnormalities caused by contamination. Also confirm that the optical window is fully immersed below the liquid surface.
Q4: What tools are required for calibration?
Prepare the following:
Calibrated reference thermometer
Calibration cup
Magnetic stir bar and magnetic stirrer
Ultrapure water
Q5: What standard solutions are required for calibration?
The standard solutions listed in the manual include:
SL45-1 Total Organic Carbon standard solution, 10 mg/L, 500 ml
SL45-2 Total Organic Carbon standard solution, 100 mg/L, 500 ml
Q6: How is temperature calibration performed?
Pour pure water or tap water into the calibration cup, then immerse the analyzer and reference thermometer in the water. Wait at least 2 minutes and confirm that the real-time temperature value is stable. Write the value of “real-time temperature value - thermometer reading” as the temperature correction value.
Q7: Which address should the temperature correction value be written to?
The temperature correction value can be written to PLC address 12289, HEX address 0x3000, Function Code 10, Length 2 Words, with the data type as floating point.
Q8: Which address should be read to confirm temperature calibration?
Read the real-time temperature value at PLC address 3, HEX address 0x0002, Function Code 04, Length 2 Words, to confirm whether the final temperature reading is correct.
Q9: What water should be used for zero calibration?
Zero calibration must use compliant purified water, such as Type II pure water with conductivity below 1.0 μS/cm, or higher-grade pure water.
Q10: How is zero calibration performed?
Place pure water and a magnetic stir bar into the calibration vessel. Immerse the analyzer to the appropriate depth, start the magnetic stirrer at a stable speed, and avoid excessive vortex formation or dense bubbles. After the temperature-compensated count value stabilizes, enter 1 to perform zero calibration.
Q11: Which address should the zero calibration value be written to?
The zero calibration value can be written to PLC address 20491, HEX address 0x500A, Function Code 10, Length 2 Words, with the data type as floating point.
Q12: What should be checked after zero calibration?
After zero calibration, read the absorbance value, abs, at PLC address 50, HEX address 0x0031, Function Code 04, Length 2 Words. After successful calibration, the abs value should be 0 ± 0.5.
Q13: How is standard-solution slope calibration performed?
Place the standard solution and a magnetic stir bar into the calibration vessel. Immerse the analyzer to the appropriate depth, start the magnetic stirrer at a stable speed, and avoid excessive vortex formation or dense bubbles. After the standard reading stabilizes, enter the standard-solution concentration as the standard calibration value to perform calibration.
Q14: Which address should be read for the standard-solution value?
The standard-solution value can be read from PLC address 36, HEX address 0x0023, Function Code 04, Length 2 Words, with the data type as floating point.
Q15: Which address should the standard calibration value be written to?
The standard calibration value can be written to PLC address 13062, HEX address 0x3305, Function Code 10, Length 2 Words, with the data type as floating point.
Q16: What concentration is recommended for slope calibration?
The recommended concentration should be selected according to the target measurement value. The manual provides 10 mg/L and 100 mg/L TOC standard solutions, which can be selected according to the monitoring range.
Q17: Can the cleaning brush be activated during calibration?
Yes. The manual recommends activating the cleaning brush to remove bubbles attached to the optical window and prevent abnormal readings.
Q18: Why should large vortices and dense bubbles be avoided during calibration?
Large vortices and dense bubbles may interfere with optical measurement, causing unstable count values, absorbance, or TOC readings and affecting calibration results.
- Calibration Questions
-
Q6
Field Calibration and Background Subtraction Questions
F- Field Calibration and Background Subtraction Questions
Q1: When is field calibration required?
Field calibration can be performed when real-time field readings differ from laboratory analysis values and the instrument has been confirmed to be properly cleaned, installed, and calibrated with standard solutions.
Q2: What should be noted before field calibration?
Because there is a time difference between laboratory testing and field reading calibration, first observe the concentration trend and select a relatively stable concentration period for field reading calibration.
Q3: How is field calibration performed?
Install the analyzer correctly in the field water body and immerse it to the appropriate depth. After the real-time TOC value stabilizes, enter the field calibration value, then read the real-time TOC value again to confirm whether calibration was successful.
Q4: Which address should the field calibration value be written to?
The field calibration value can be written to PLC address 13058, HEX address 0x3301, Function Code 10, Length 2 Words, with the data type as floating point.
Q5: Which address should be read for the real-time TOC value during field calibration?
The real-time TOC value can be read from PLC address 1, HEX address 0x0000, Function Code 04, Length 2 Words, with the data type as floating point.
Q6: What is background subtraction?
Background subtraction is a correction function for background interference in field samples. It can be used to subtract the effect of a known background value on TOC readings.
Q7: When is background subtraction not recommended?
If the field sample concentration trend has not been characterized through laboratory testing, background subtraction is not recommended.
Q8: Which address should the background correction value be written to?
The background correction value can be written to PLC address 28678, HEX address 0x7005, Function Code 10, Length 2 Words, with the data type as floating point.
Q9: How can successful background subtraction be confirmed?
Read the real-time TOC value again to confirm whether the reading has been corrected according to the background value, and compare it with a known concentration or laboratory analysis result.
Q10: Can the cleaning brush be activated during field calibration or background subtraction?
Yes. The cleaning brush can be activated to remove bubbles attached to the optical window and prevent abnormal readings.
- Field Calibration and Background Subtraction Questions
-
Q7
AO Analog Output Questions
G- AO Analog Output Questions
Q1: Does it support AO 4-20 mA output?
If the AO version is selected, the AO output range can be configured so that 4-20 mA corresponds to the specified TOC measurement range.
Q2: What equipment is required for AO calibration?
AO calibration requires the SMR instrument, suitable power cable, a USB-to-RS485 converter such as CAB06, and a digital multimeter or current calibrator.
Q3: Are reagents required for AO calibration?
No. AO calibration is analog output current calibration and does not require reagents.
Q4: What is the AO1 calibration wiring configuration?
AO1 wiring is as follows:
Brown: AO1+, connect to the meter mA+ terminal
White: AO1-, connect to the meter COM / mA- terminal
Q5: What is the AO2 calibration wiring configuration?
AO2 wiring is as follows:
Blue: AO2+, connect to the meter mA+ terminal
Black: AO2-, connect to the meter COM / mA- terminal
Q6: What mode should the meter be set to during AO calibration?
Set the meter to DC mA mode and ensure the test leads are inserted into the correct mA measurement terminals.
Q7: What does a negative value on the AO meter indicate?
If the meter displays a negative value, the polarity may be reversed. Swap the meter connections and measure again.
Q8: Which addresses should AO Min / Max be written to?
The AO output range can be written to:
Min: PLC address 31241, HEX address 0x7A08
Max: PLC address 31242, HEX address 0x7A09
Q9: Which address should the AO current offset be written to?
The AO current offset can be written to PLC address 31236, HEX address 0x7A03, Function Code 10, with the data type as floating point.
Q10: How can normal output be confirmed during AO calibration?
First set the Min and Max range, then read the real-time measured value and calculate the theoretical current corresponding to that reading within the 4-20 mA range. If the meter reading differs from the theoretical current, write a current offset for correction, then confirm again that the meter current has been calibrated to the theoretical current.
- AO Analog Output Questions
-
Q8
Abnormal Readings and Troubleshooting
H- Abnormal Readings and Troubleshooting
Q1: What should be done if the reading keeps fluctuating?
First confirm whether the analyzer is immersed to the appropriate depth and whether the optical window is fully below the liquid surface. Check whether the window is blocked by bubbles, fouling, sediment, or the cleaning brush.
Q2: What should be done if the reading remains at 0?
Confirm whether the sample may contain TOC, then check whether the optical window is clean, the light source is normal, and zero and standard calibrations are correct. If there is still no response after replacing the standard solution, contact AQUAS.
Q3: What should be done if the reading remains at the upper range?
Possible causes include TOC concentration exceeding the measurement range, excessive suspended solids concentration, optical-window fouling, bubble interference, or incorrect calibration. Clean the optical window first and confirm standard calibration.
Q4: What should be done if the reading is obviously too high or too low?
Check whether zero calibration, slope calibration, field calibration, and background subtraction settings are correct, and confirm that the optical window has no deposits, sediment, or bubbles.
Q5: What should be done if the response time becomes slower?
Possible causes include optical-window fouling, deposits on the sensing area, unstable sample conditions, suspended solids interference, or abnormal calibration. Clean the optical window first and verify the standard-solution reading.
Q6: What should be done if readings differ for the same sample?
Confirm whether the sample is homogeneous, whether sediment or suspended solids are changing rapidly, whether bubbles are attached, whether stirring is stable, and whether the calibration procedure is correct.
Q7: What should be done if the reading does not change?
Restore factory settings first and repeat the calibration procedure. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.
Q8: Do air bubbles affect readings?
Yes. Air bubbles may attach to the optical window and cause abnormal optical readings. The cleaning brush can be activated to remove bubbles, and dense bubbles should be avoided during calibration or measurement.
Q9: Will readings be affected if the cleaning brush does not return to its home position?
Yes. If the cleaning brush blocks the optical window’s receiving or emitting light path, it may cause abnormal readings. Confirm that the cleaning brush has returned to the correct position.
Q10: How can the light source be checked?
Insert white paper into the optical path and visually check whether the lower optical window emits LED light at a fixed frequency. If the light source is abnormal, contact AQUAS.
Q11: Can excessive suspended solids affect the SMR45?
Yes. Although the analyzer provides suspended solids measurement and automatic interference compensation, excessively high suspended solids concentration may still cause abnormal testing or unstable readings.
Q12: Does a difference between field readings and laboratory values indicate device failure?
Not necessarily. Real-time monitoring instruments are directly exposed to field conditions, while laboratory samples typically undergo sampling, storage, pretreatment, and standardized procedures. Differences in test conditions and timing may cause value differences.
Q13: When should AQUAS technical support be contacted?
If the device still cannot be read or the readings remain unresponsive after checking power supply, wiring, Modbus ID, communication settings, standard-solution testing, zero calibration, slope calibration, optical-window cleaning, and light-source check, contact AQUAS technical support and provide site photos, wiring photos, Modbus settings, standard-solution test records, and reading screenshots.
- Abnormal Readings and Troubleshooting
-
Q9
Maintenance and Consumables Questions
I- Maintenance and Consumables Questions
Q1: Is regular maintenance required?
Yes. To keep the analyzer in optimal operating condition, regular maintenance and cleaning are required, especially in environments where grease, sediment, sludge, or suspended solids can easily accumulate.
Q2: How often should cleaning be performed?
The manual recommends cleaning the analyzer approximately once per week. The actual frequency can be adjusted according to site water quality and contamination level.
Q3: How often should calibration solution be used for calibration?
The manual recommends calibrating the analyzer with calibration solution approximately once every two weeks. The actual frequency can be adjusted according to site water quality and data trends.
Q4: What does routine maintenance include?
Routine maintenance is recommended to include:
Cleaning the entire instrument with detergent and a soft brush
Rinsing with tap water
Confirming that the optical window and instrument body are free of dirt
Confirming that no bubbles are attached to the optical window
Confirming that the standard solution has not deteriorated or become contaminated
Confirming that cables and mounting fixtures are not loose or damaged
Q5: Can acid be used for cleaning?
The maintenance procedure in the manual states that if visible contamination is present, 5% diluted sulfuric acid can be used for a 15-minute soak, followed by zero-condition verification in pure water.
Q6: What should the reading be during zero verification?
When the analyzer is placed in pure water for zero verification, the measured parameter value should be less than 5% of full scale.
Q7: What should be done if the reading remains high after acid soaking?
If the reading remains greater than 15% of full scale after acid soaking, the manual recommends return for service and maintenance. Calibration can still be continued at this stage, but reliability will decrease.
Q8: What error is considered normal when testing with calibration solution?
According to the maintenance procedure in the manual, when testing with calibration solution, a measured-parameter error of less than 5% can be considered acceptable after maintenance.
Q9: What should be done if the calibration-solution test error is 5%-20%?
If the error is between 5% and 20%, single-point slope calibration can be performed.
Q10: What should be done if the calibration-solution test error is greater than 20%?
If the error is greater than 20%, check whether bubbles are attached to the optical window and whether the standard solution has deteriorated or become contaminated. If necessary, remove the bubbles or replace the standard solution and test again.
Q11: What consumables are used?
The main consumables include TOC standard solutions and pure water. The standard solutions listed in the manual are SL45-1 10 mg/L and SL45-2 100 mg/L Total Organic Carbon standard solutions.
Q12: How should it be stored when not used for an extended period?
If the analyzer will not be used for an extended period, clean it thoroughly and store it dry to prevent long-term contamination buildup on the optical window or impact damage.
Q13: What should be noted during cleaning?
Use a soft brush, detergent, pure water, or neutral cleaning method to clean the analyzer surface and optical window. Avoid hard objects that may scratch the optical window or damage the sensing end.
- Maintenance and Consumables Questions