SMR43 UV254 Analyzer
Frequently Asked Questions
-
Q1
Basic Product Questions
A- Basic Product Questions
Q1: What is this product?
The SMR43 is an AQUAS UV254 ultraviolet absorbance and suspended solids analyzer for real-time monitoring of UV254, suspended solids, and related water-quality indicators in water, with data output via RS485 digital communication.
Q2: What parameters can it measure?
It can directly output UV254 ultraviolet absorbance and suspended solids measurement data, and can also output monitoring values such as COD, BOD, and TOC through correlation settings.
Q3: What is the measurement principle?
It measures the absorbance of organic compounds containing conjugated double bonds or polycyclic aromatic hydrocarbons at 254 nm ultraviolet light, and detects suspended solids concentration through a 530 nm optical beam system for automatic compensation.
Q4: Are chemical reagents required?
No. It uses optical measurement and does not require chemical reagents, making it suitable for real-time continuous monitoring.
Q5: What applications is it suitable for?
It is suitable for drinking water, surface water, industrial processes, water treatment, wastewater, and related applications.
Q6: Is a control box required?
Not necessarily. The analyzer features an integrated miniature housing with a built-in transmitter and sensor. Its rugged IP68 submersible, corrosion-resistant enclosure allows direct installation according to site requirements.
Q7: Does it support PLC or HMI integration?
Yes. It uses the standard Modbus RTU communication protocol and can be connected to a PLC, HMI, or AQUAS data logger via RS485.
Q8: Does it have explosion-proof certification?
Yes. It is certified to IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS explosion-proof standards.
Q9: Does it support automatic cleaning?
Yes. It is equipped with an automatic cleaning brush to reduce maintenance caused by optical-window contamination, bubble attachment, or sludge deposition.
Q10: What are the features of the optical window material?
The optical window is made of abrasion-resistant, scratch-resistant sapphire glass and includes a nano-coating to reduce sludge deposition and biofouling, improving maintenance convenience.
Q11: Can it compensate for light-source aging or optical-window fouling?
Yes. Built-in reference-light measurement enables automatic compensation for light-source aging, optical-window fouling, and related conditions.
- Basic Product Questions
-
Q2
Installation Questions
B- Installation Questions
Q1: What is the operating voltage?
The operating voltage is 5-12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.
Q2: What installation methods are available?
Immersion, flow-through, or pipe-insertion installation can be used.
Q3: What should be noted during installation?
Avoid pulling or damaging the cable during installation. Ensure the cable length is sufficient and properly secured. Installation in a location with stable water flow is recommended for better measurement performance.
Q4: How deep should the sensing surface be immersed?
During testing, the sensing surface must be fully immersed at least 10 cm below the water surface.
Q5: How long does it take to stabilize after installation?
After installation, allow at least 5 minutes for the analyzer to reach a stable condition. The actual stabilization time may vary depending on sample characteristics and the installation environment.
Q6: Can it be installed close to the bottom of the tank?
It is not recommended. The analyzer should not directly contact or be installed too close to the tank bottom, as obstruction of the sensing area, sediment interference, or optical-window contamination may cause abnormal measurements.
Q7: Can it be installed in a flow-through holder?
Yes. The manual lists the FTC43-1 flow-through holder and PIF43-1 simple pipe-insertion holder, which can be selected according to on-site pipeline or flow-cell requirements.
Q8: What should be considered for outdoor or high-load installations?
For outdoor or high-load installation environments, AQUAS dedicated protective covers, holders, or fixtures are recommended to improve protection and measurement stability.
Q9: Is calibration required before first use?
Generally, no. The analyzer is calibrated before shipment. For first use, measurement can begin after proper installation according to the manual.
- Installation Questions
-
Q3
Communication and Wiring Questions
C- Communication and Wiring Questions
Q1: What communication interface is used?
It uses an RS485 communication interface.
Q2: What communication protocol is used?
It uses the Modbus RTU communication protocol.
Q3: What are the default communication settings?
Common communication settings are as follows:
Communication interface: RS485
Baud rate: 19200 bps
Data length: 8 bit
Parity: None
Stop bit: 1 bit
Communication mode: RTU
The device supports 9600 and 19200 bps.
Q4: What is the default device address?
The factory default device address is 01. If the device address is changed, the address configured in the PLC, HMI, host, or reading software must also be updated accordingly.
Q5: What is the RS485 wiring configuration?
The SMR male connector wiring is as follows:
PIN 1: GND
PIN 2: RS485-
PIN 3: RS485+
PIN 4: 5-12V DC
PIN 5: Not used
The typical new cable colors are: black for GND, white for RS485-, blue for RS485+, and red for 5-12V DC.
Q6: Can it be connected directly to a computer?
Yes. It can be connected to a computer using the CAB06 USB-to-RS485 cable, allowing values to be read, calibration to be performed, and parameters to be configured through Modbus tools or AQCFG.
Q7: Can it be connected to an AQUAS data logger?
Yes. It can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data recording and remote monitoring according to site requirements.
Q8: What should be checked first if the PLC cannot read values?
Check the following in order:
1. Whether the device can be connected to a PC using CAB06 and read through Modbus
2. Whether the PLC and SMR43 communication settings are consistent
3. Whether the Modbus ID is correct and whether there is an ID conflict
4. Whether the PLC Modbus address is correct
5. Whether the power supply is stable at 5-12 VDC
6. Whether the RS485 wiring is correct
7. Whether wiring has been completed according to the appendix wiring diagram
Q9: What happens if RS485 D+ / D- are reversed?
If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
- Communication and Wiring Questions
-
Q4
Modbus Reading Questions
D- Modbus Reading Questions
Q1: Which address should be read for the main parameter value?
The main parameter value can be read from PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with the data type as floating point.
Q2: Which address should be read for the secondary parameter value?
The secondary parameter value can be read from PLC address 2, HEX address 0x0001, Length 2 Words, Function Code 04, with the data type as floating point.
Q3: How can the real-time temperature value be read?
The real-time temperature value can be read from PLC address 3, HEX address 0x0002, Function Code 04, Length 2 Words, with the data type as floating point.
Q4: How can the device model be read?
Read PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with the data type as an ASCII string.
Q5: How can the location, area, or sub-area name be read?
Read PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with the data type as a UTF-16BE string.
Q6: How can latitude and longitude be read?
Read PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type as floating point.
Q7: How can the cleaning brush be activated?
If the device is equipped with a cleaning brush, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.
Q8: How should floating-point data be decoded?
SMR43 floating-point data follows the IEEE 754 single-precision format and occupies 4 Bytes. If the decoded value is abnormal, verify whether the byte order or word order needs to be swapped.
Q9: What should be noted about Modbus CRC?
Modbus RTU uses CRC16. During transmission, the CRC is sent Low Byte First and High Byte Last, meaning the low byte is transmitted before the high byte.
Q10: What do Function Codes 03, 04, and 10 represent?
Function Code 03: Read holding registers
Function Code 04: Read input registers
Function Code 10: Write multiple holding registers
- Modbus Reading Questions
-
Q5
Calibration Questions
E- Calibration Questions
Q1: When is recalibration required?
Recalibration is recommended when field readings differ significantly from laboratory values, readings drift after long-term use, accuracy needs to be confirmed after maintenance and cleaning, or field sample conditions differ greatly from standard-solution conditions.
Q2: What calibration methods are available?
SMR43 calibration includes temperature calibration, standard zero calibration, standard slope calibration, field calibration, and background subtraction.
Q3: What should be done before calibration?
Before calibration, thoroughly clean the entire analyzer and all calibration-related vessels with detergent and a soft brush. Confirm that the optical window is fully immersed below the liquid surface. If necessary, activate the cleaning brush to remove attached bubbles and prevent abnormal readings.
Q4: What tools are required for calibration?
Prepare the following:
Calibrated reference thermometer
Calibration cup
Magnetic stir bar and magnetic stirrer
Ultrapure water
Q5: What standard solution is required for calibration?
Prepare SL43-1 KHP standard solution, concentration 1,000 mg/L, volume 500 ml.
Q6: How is temperature calibration performed?
Pour pure water or tap water into the calibration cup, then place the analyzer and reference thermometer into the water. Wait at least 2 minutes and confirm that the real-time temperature value is stable. Write the value of “real-time temperature value - thermometer reading” as the temperature correction value.
Q7: Which address should the temperature correction value be written to?
The temperature correction value can be written to PLC address 12289, HEX address 0x3000, Function Code 10, Length 2 Words, with the data type as floating point.
Q8: What water should be used for zero calibration?
Zero calibration must use compliant purified water, such as Type II pure water with conductivity below 1.0 µS/cm, or higher-grade pure water.
Q9: How is zero calibration performed?
Place pure water and a magnetic stir bar into the calibration vessel. Immerse the analyzer to the appropriate depth, start the magnetic stirrer at a stable speed, and avoid excessive vortex formation or dense bubbles. After the temperature-compensated count value stabilizes, enter 1 to perform zero calibration, then confirm that the absorbance value is 0 ± 0.5.
Q10: Which address should the zero calibration value be written to?
The zero calibration value can be written to PLC address 20491, HEX address 0x500A, Function Code 10, Length 2 Words, with the data type as floating point.
Q11: Which count value should be read during zero calibration?
Read the temperature-compensated count value at PLC address 292, HEX address 0x0123, Function Code 04, Length 2 Words, with the data type as floating point.
Q12: How can successful zero calibration be confirmed?
Read the absorbance value, abs, at PLC address 50, HEX address 0x0031, Function Code 04. After successful calibration, the abs value should display 0 ± 0.5.
Q13: How is slope calibration performed?
Place the SL43-1 KHP standard solution and a magnetic stir bar into the calibration vessel. Immerse the analyzer to the appropriate depth, start the magnetic stirrer at a stable speed, and after the standard reading stabilizes, enter the standard-solution concentration as the standard calibration value to perform calibration.
Q14: Which address should be read for the standard-solution value?
The standard-solution value can be read from PLC address 36, HEX address 0x0023, Function Code 04, Length 2 Words, with the data type as floating point.
Q15: Which address should the standard calibration value be written to?
The standard calibration value can be written to PLC address 13062, HEX address 0x3305, Function Code 10, Length 2 Words, with the data type as floating point.
Q16: Why should large vortices and dense bubbles be avoided during calibration?
Large vortices and dense bubbles may interfere with the optical window and optical-path measurement, causing unstable UV254 or SAC readings and affecting calibration results.
Q18: Can the cleaning brush be activated during calibration?
Yes. The cleaning brush can be activated to remove bubbles attached to the optical window and prevent abnormal readings.
- Calibration Questions
-
Q7
Field Calibration and Background Subtraction Questions
F- Field Calibration and Background Subtraction Questions
Q1: When is field calibration required?
Field calibration can be performed when real-time field readings differ from laboratory analysis values and the instrument has been confirmed to be properly cleaned, installed, and calibrated with standard solutions.
Q2: What should be noted before field calibration?
Because there is a time difference between laboratory testing and field reading calibration, first observe the concentration trend and select a relatively stable concentration period for field reading calibration.
Q3: How is field calibration performed?
Install the analyzer correctly in the field water body and immerse it to the appropriate depth. After the real-time SAC value stabilizes, enter the field calibration value to perform calibration, then read the real-time SAC value again to confirm success.
Q4: Which address should the field calibration value be written to?
The field calibration value can be written to PLC address 13058, HEX address 0x3301, Function Code 10, Length 2 Words, with the data type as floating point.
Q5: Which address should be read for the real-time SAC value during field calibration?
The real-time SAC value can be read from PLC address 1, HEX address 0x0000, Function Code 04, Length 2 Words, with the data type as floating point.
Q6: What is background subtraction?
Background subtraction is a correction function for background interference in field samples. It can be used to subtract the effect of a known background value on SAC or related readings.
Q7: When is background subtraction not recommended?
If the field sample concentration trend has not been characterized through laboratory testing, background subtraction is not recommended.
Q8: Which address should the background correction value be written to?
The background correction value can be written to PLC address 28678, HEX address 0x7005, Function Code 10, Length 2 Words, with the data type as floating point.
- Field Calibration and Background Subtraction Questions
-
Q8
Abnormal Readings and Troubleshooting
F- Abnormal Readings and Troubleshooting
Q1: What should be done if the reading keeps fluctuating?
First confirm whether the analyzer is immersed to the appropriate depth and whether the optical window is fully below the liquid surface. Check whether the window is blocked by bubbles, deposits, sediment, or the cleaning brush.
Q2: What should be done if the reading remains at 0?
Confirm whether the sample actually has a UV254 or SAC response, then check whether the optical window is clean, whether it is immersed deeply enough, and whether calibration is correct. If there is still no response when testing with a standard solution, contact AQUAS.
Q3: What should be done if the reading remains at the upper range?
Possible causes include sample concentration exceeding the measuring range, excessive suspended solids concentration, optical-window contamination, incorrect calibration, or bubble interference. Clean the analyzer first and reconfirm calibration.
Q4: What should be done if the reading is obviously too high or too low?
Check whether zero and slope calibrations are correct, whether the optical window is dirty, whether the sample is homogeneous, whether the suspended solids concentration is too high, and whether bubbles or sediment are causing interference.
Q5: What should be done if the response time becomes slower?
Possible causes include optical-window contamination, unstable sample conditions, suspended solids interference, or deposits affecting the sensing area. Clean the optical window first and verify the standard-solution reading.
Q6: What should be done if readings differ for the same sample?
Confirm whether the sample is homogeneous, field concentration is stable, there is no stratification, sediment, or bubble attachment, and the calibration procedure is correct.
Q7: What should be done if the reading does not change?
Restore factory settings first and repeat the calibration procedure. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.
Q8: How can the SMR43 light source be checked?
Insert white paper into the optical path and visually check whether the lower optical window emits LED light at a fixed frequency. If there is no response, the light source may be abnormal; contact AQUAS.
Q9: Can excessive suspended solids affect measurements?
Yes. Excessive suspended solids concentration may cause abnormal testing or reading deviation. Confirm whether both the target analyte concentration and suspended solids concentration are within the analyzer’s measuring range.
Q10: Will readings be affected if the cleaning brush does not return to its home position?
Yes. If the cleaning brush does not return to the correct position, it may block the optical window’s receiving or emitting light path, causing abnormal readings.
Q11: Does a difference between field readings and laboratory values indicate device failure?
Not necessarily. Real-time monitoring instruments are directly exposed to field conditions, while laboratory samples typically undergo sampling, storage, pretreatment, and standardized procedures. Differences in test conditions and timing may cause value differences.
Q12: When should AQUAS technical support be contacted?
If the device still cannot be read or the readings remain unresponsive after checking power supply, wiring, Modbus ID, communication settings, standard-solution testing, calibration, cleaning, light-source check, and installation, contact AQUAS technical support and provide site photos, wiring photos, Modbus settings, calibration records, and reading screenshots.
- Abnormal Readings and Troubleshooting
-
Q9
Maintenance and Consumables Questions
F- Maintenance and Consumables Questions
Q1: Is regular maintenance required?
Yes. To keep the analyzer in optimal operating condition, regular cleaning and calibration are required, especially in environments where grease, sediment, or other contaminants can easily adhere.
Q2: How often should cleaning be performed?
Cleaning is recommended approximately once per week, with the frequency adjusted according to site water-quality conditions.
Q3: How often should calibration solution be used for calibration?
Calibration with calibration solution is recommended approximately once every two weeks, with the frequency adjusted according to site water-quality conditions.
Q4: How should it be stored when not used for an extended period?
If the analyzer will not be used for an extended period, clean it thoroughly and store it dry.
Q5: What maintenance cleaning procedure is recommended?
It is recommended to clean the entire instrument with detergent and a soft brush, then rinse with tap water and confirm that the optical window and instrument body are free of dirt. If visible contamination remains, soak in 5% diluted sulfuric acid for 15 minutes, then place it in pure water to perform zero verification.
Q6: What condition should the reading meet during zero verification?
When performing zero verification in pure water, the measured parameter value should be less than 5% of full scale. If the reading remains greater than 15% of full scale after acid soaking, factory maintenance is recommended.
Q7: What standard-solution test error is considered normal?
When testing with calibration solution, a measured-parameter error of less than 5% can be considered acceptable after maintenance. If the error is between 5% and 20%, single-point slope calibration can be performed. If the error is greater than 20%, check whether there are bubbles on the optical window or whether the standard solution has deteriorated or become contaminated.
Q8: What consumables are used?
The main consumables include SL43-1 KHP standard solution and pure water.
Q9: What should be noted during cleaning?
Use a soft cloth, soft brush, or neutral detergent for cleaning, and avoid hard objects that may scratch the optical surface. If acid soaking is required, use 5% diluted sulfuric acid according to the maintenance procedure and control the soaking time.
- Maintenance and Consumables Questions