SMR38 Colored Dissolved Organic Matter / Dissolved Organic Matter Fluorescence (CDOM/fDOM) Analyzer
Frequently Asked Questions
-
Q1
Basic Product Questions
A- Basic Product Questions
Q1: What is this product?
It is an AQUAS CDOM/fDOM analyzer for real-time monitoring of colored dissolved organic matter or fluorescent dissolved organic matter in water, with data output via RS485 digital communication.
Q2: What parameters can it measure?
It is primarily used to measure colored dissolved organic matter (CDOM/fDOM), serving as a real-time indicator of organic-matter variation in surface water, reservoirs, rivers, drinking-water systems, wastewater, and similar water bodies.
Q3: What is the measurement principle?
It uses a fluorescence absorption principle. Blue excitation light stimulates colored dissolved organic matter molecules in the water to emit green fluorescence, and the fluorescence intensity is proportional to the CDOM/fDOM concentration.
Q4: What applications is it suitable for?
It is suitable for surface-water monitoring and can also be directly installed in reservoirs, rivers, drinking-water systems, wastewater applications, and related environments.
Q5: Is a control box required?
Not necessarily. The analyzer features an integrated miniature housing with a built-in transmitter and sensor. Its rugged IP68 submersible, corrosion-resistant enclosure allows direct field installation.
Q6: Does it support PLC or HMI integration?
Yes. It uses the standard Modbus RTU communication protocol and can be connected to a PLC, HMI, or AQUAS data logger via RS485.
Q7: Is it suitable for outdoor sites without mains power?
Yes. With its ultra-low-power design, it can be powered by batteries or solar power for outdoor or remote monitoring applications.
Q8: Does it have explosion-proof certification?
Yes. It is certified to IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS explosion-proof standards.
Q9: What is the maximum transmission distance?
Using a 1000N high-strength Kevlar cable, the maximum communication distance can reach up to 1.2 km.
Q10: Does it provide self-diagnostic functions?
Yes. It includes built-in self-status monitoring, such as temperature and humidity monitoring, as well as LED light-source abnormality diagnostics.
Q11: Does it provide noise-processing capability?
Yes. It uses advanced algorithms and signal-processing technology to help suppress noise and improve measurement stability.
Q12: Does it support automatic cleaning?
Yes. It can be equipped with an automatic cleaning brush to reduce maintenance caused by deposits on the optical window, sludge, biofouling, or air bubbles.
Q13: What are the features of the optical window material?
The optical window is made of abrasion-resistant, scratch-resistant sapphire glass, making it less prone to wear during cleaning and maintenance. It also includes a nano-coating to reduce sludge deposition and biofouling.
- Basic Product Questions
-
Q2
Installation Questions
B- Installation Questions
Q1: What is the operating voltage?
The operating voltage is 5-12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.
Q2: What installation methods are available?
Immersion, flow-through, or pipe-insertion installation can be used.
Q3: What should be noted during installation?
Avoid pulling or damaging the cable during installation. Ensure the cable length is sufficient and properly secured. Installation in a location with stable water flow is recommended for better measurement performance.
Q4: How deep should the sensing surface be immersed?
During testing, the sensing surface must be fully immersed at least 10 cm below the water surface.
Q5: How long does it take to stabilize after installation?
After installation, allow at least 5 minutes for the analyzer to reach a stable condition. The actual stabilization time may vary depending on sample characteristics and the installation environment.
Q6: Can it be installed close to the bottom of the tank?
It is not recommended. The analyzer should not directly contact or be installed too close to the tank bottom, as obstruction of the sensing area, bottom sludge, or sediment may cause abnormal measurements.
Q7: How should it be secured in the field?
Use a protective pipe and mounting clamps to secure the instrument, preventing impact and installation displacement.
Q8: Are there recommended pressure and flow-velocity ranges for field operation?
Yes. During operation, the pressure is recommended not to exceed 10 kgf/cm², and the sample flow velocity should be maintained between 0.1 and 10 m/s to prevent damage caused by overpressure or water-flow impact.
Q9: What should be considered for outdoor or high-load installations?
For outdoor or high-load installation environments, AQUAS dedicated protective covers and fixtures are recommended to improve overall protection and measurement stability.
Q10: How should the SMR38 be set up for laboratory testing?
Connect the analyzer to a computer, secure it vertically on a stand, insert it into the calibration vessel, and ensure the sensing area is fully immersed in the sample water. Avoid contact with the bottom or sides of the vessel.
Q11: What are the requirements for the laboratory calibration vessel?
The calibration vessel is recommended to be at least 30 cm high and made of black, opaque material to reduce interference from external light.
Q12: Is stirring required during laboratory testing?
It is recommended to place a 2.5 cm magnetic stir bar at the bottom of the vessel and use a magnetic stirrer to maintain stable rotation, preferably around 300 rpm, to prevent sample stratification and sediment from affecting the results.
Q13: Can it be installed in a flow-through holder?
Yes. The manual installation diagrams include FTC38-1a and FTC38-1b flow-through holders, as well as the PIF38-1 simple pipe-insertion holder, which can be selected according to site requirements.
- Installation Questions
-
Q3
Communication and Wiring Questions
C- Communication and Wiring Questions
Q1: What communication interface is used?
It uses an RS485 communication interface.
Q2: What communication protocol is used?
It uses the Modbus RTU communication protocol.
Q3: What are the default communication settings?
Common communication settings are as follows:
Communication interface: RS485
Baud rate: 19200 bps
Data length: 8 bit
Parity: None
Stop bit: 1 bit
Communication mode: RTU
The device supports 9600 and 19200 bps.
Q4: What is the default device address?
The factory default device address is 01. If the device address is changed, the address configured in the host, PLC, HMI, or reading software must also be updated accordingly.
Q5: What is the RS485 wiring configuration?
The SMR male connector wiring is as follows:
PIN 1: GND
PIN 2: RS485-
PIN 3: RS485+
PIN 4: 5-12V DC
PIN 5: Not used
The typical new cable colors are: black for GND, white for RS485-, blue for RS485+, and red for 5-12V DC.
Q6: If AO analog output is available, what is PIN 5?
For the AO version, PIN 5 is the AO analog output wire, typically gray. For the standard non-AO version, PIN 5 is not used.
Q7: Can it be connected directly to a computer?
Yes. It can be connected to a computer using the CAB06 USB-to-RS485 cable, allowing values to be read, calibration to be performed, parameters to be configured, and diagnostics to be conducted through Modbus or utility software.
Q8: Can it be connected to an AQUAS data logger?
Yes. It can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data recording and remote monitoring according to site requirements.
Q9: What should be checked first if the PLC cannot read values?
Check the following in order:
1. Whether the device can be connected to a PC using CAB06 and read through Modbus
2. Whether the PLC and SMR38 communication settings are consistent
3. Whether the Modbus ID is correct and whether there is an ID conflict
4. Whether the PLC Modbus address is correct
5. Whether the power supply is stable at 5-12 VDC
6. Whether the RS485 wiring is correct
7. Whether wiring has been completed according to the appendix wiring diagram
Q10: What happens if RS485 D+ / D- are reversed?
If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
- Communication and Wiring Questions
-
Q4
Modbus Reading Questions
D- Modbus Reading Questions
Q1: Which address should be read for the main CDOM/fDOM parameter value?
The main CDOM/fDOM parameter value can be read from PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with the data type as floating point.
Q2: Which address should be read for the secondary parameter value?
The secondary parameter value can be read from PLC address 2, HEX address 0x0001, Length 2 Words, Function Code 04, with the data type as floating point.
Q3: How can the device model be read?
Read PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with the data type as an ASCII string.
Q4: How can the location, area, or sub-area name be read?
Read PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with the data type as a UTF-16BE string.
Q5: How can latitude and longitude be read?
Read PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type as floating point.
Q6: How can the cleaning brush be activated?
If the device is equipped with a cleaning brush, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.
Q7: How should floating-point data be decoded?
Floating-point data follows the IEEE 754 single-precision format and occupies 4 Bytes. If the decoded value is abnormal, verify whether the byte order or word order needs to be swapped.
Q8: What should be noted about Modbus CRC?
Modbus RTU uses CRC16. During transmission, the CRC is sent Low Byte First and High Byte Last, meaning the low byte is transmitted before the high byte.
Q9: What do Function Codes 03, 04, and 10 represent?
Function Code 03: Read holding registers
Function Code 04: Read input registers
Function Code 10: Write multiple holding registers
Q10: Why do garbled or unreasonable values appear when reading data?
Common causes include inconsistent floating-point byte or word order, incorrect PLC address settings, incorrect Function Code selection, mismatched communication parameters, or incorrect data parsing format. Confirm the Modbus register table and communication format first.
- Modbus Reading Questions
-
Q5
Calibration Questions
E- Calibration Questions
Q1: Is calibration required before first use?
Generally, no. The analyzer is calibrated before shipment. For first use, measurement can begin after proper installation according to the manual. If accuracy confirmation is required before use, standard-solution calibration can be performed.
Q2: What calibration items are available?
Calibration items include standard zero calibration, standard slope calibration, and AO calibration for the optional AO version.
Q3: What should be done before calibration?
Before calibration, thoroughly clean the analyzer. Rinse the surface with tap water to remove debris and residues. If necessary, use a soft brush and neutral detergent for cleaning, and dry all parts of the analyzer body as much as possible.
Q4: What tools are required for calibration?
Prepare the following:
Calibrated reference thermometer
Calibration cup
Magnetic stir bar and magnetic stirrer
Ultrapure water
Q5: What standard solution is required for calibration?
The standard solution listed in the manual is SL38-1 CDOM standard solution, 1500 μg/L, 1000 ml.
Q6: What water should be used for zero calibration?
Zero calibration must use compliant purified water, such as Type II pure water with conductivity below 1.0 μS/cm, or higher-grade pure water.
Q7: How is zero calibration performed?
Place pure water and a magnetic stir bar into the calibration vessel. Immerse the analyzer to the appropriate depth, start the magnetic stirrer at a stable speed, and avoid excessive vortex formation or dense bubbles. Then read the zero calibration value, enter 0 as the zero calibration value to perform calibration, and finally check the real-time CDOM/fDOM value to confirm successful calibration.
Q8: Which address should the zero calibration value be written to?
The zero calibration value can be written to PLC address 13057, HEX address 0x3300, Function Code 10, Length 2 Words, with the data type as floating point.
Q9: Which value should be read to confirm zero calibration?
Read the real-time CDOM/fDOM value at PLC address 1, HEX address 0x0000, Function Code 04, Length 2 Words, to confirm whether zero calibration was successful.
Q10: How is slope calibration performed?
Place the standard solution and a magnetic stir bar into the calibration vessel. Immerse the analyzer to the appropriate depth, start the magnetic stirrer at a stable speed, and avoid excessive vortex formation or dense bubbles. After the real-time CDOM/fDOM value stabilizes, enter the standard-solution concentration as the slope calibration value to perform standard calibration.
Q11: Which address should the slope calibration value be written to?
The slope calibration value can be written to PLC address 13062, HEX address 0x3305, Function Code 10, Length 2 Words, with the data type as floating point.
Q12: What concentration is recommended for slope calibration?
The recommended concentration should be selected according to the target measurement range. The standard solution provided in the manual is SL38-1 CDOM standard solution at 1500 μg/L.
Q13: How can slope calibration be confirmed?
Read the real-time CDOM/fDOM value again to confirm whether the reading has been adjusted according to the standard-solution concentration.
Q14: Why should large vortices and dense bubbles be avoided during calibration?
Large vortices and dense bubbles may interfere with optical readings, causing unstable CDOM/fDOM values and affecting calibration results.
Q15: Why should a black opaque vessel be used during calibration?
Because the analyzer uses optical fluorescence measurement, external light may affect the measured value. A black opaque vessel helps reduce light interference and improve calibration stability.
- Calibration Questions
-
Q6
AO Analog Output Questions
F- AO Analog Output Questions
Q1: Does it support AO 4-20 mA output?
If the AO version is selected, the AO output range can be configured so that 4-20 mA corresponds to the specified CDOM/fDOM measurement range.
Q2: What equipment is required for AO calibration?
AO calibration requires the SMR instrument, suitable power cable, a USB-to-RS485 converter such as CAB06, and a digital multimeter or current calibrator.
Q3: Are reagents required for AO calibration?
No. AO calibration is analog output current calibration and does not require reagents.
Q4: What is the AO1 calibration wiring configuration?
AO1 wiring is as follows:
Brown: AO1+, connect to the meter mA+ terminal
White: AO1-, connect to the meter COM / mA- terminal
Q5: What is the AO2 calibration wiring configuration?
AO2 wiring is as follows:
Blue: AO2+, connect to the meter mA+ terminal
Black: AO2-, connect to the meter COM / mA- terminal
Q6: What mode should the meter be set to during AO calibration?
Set the meter to DC mA mode and ensure the test leads are inserted into the correct mA measurement terminals.
Q7: What does a negative value on the AO meter indicate?
If the meter displays a negative value, the polarity may be reversed. Swap the meter connections and measure again.
Q8: Which addresses should AO Min / Max be written to?
The AO output range can be written to:
Min: PLC address 31241, HEX address 0x7A08
Max: PLC address 31242, HEX address 0x7A09
Q9: Which address should the AO current offset be written to?
The AO current offset can be written to PLC address 31236, HEX address 0x7A03, Function Code 10, with the data type as floating point.
Q10: How can normal output be confirmed during AO calibration?
First set the Min and Max range, then read the real-time measured value and calculate the theoretical current corresponding to that reading within the 4-20 mA range. If the meter reading differs from the theoretical current, write a current offset for correction, then confirm again that the meter current has been calibrated to the theoretical current.
- AO Analog Output Questions
-
Q7
Abnormal Readings and Troubleshooting
G- Abnormal Readings and Troubleshooting
Q1: What should be done if the reading keeps fluctuating?
First clean the sensing area and optical window. Confirm that the installation depth is sufficient, the device is securely fixed, there are no air bubbles, bottom sludge, or contaminants causing interference, and that the power supply and RS485 communication are normal.
Q2: What should be done if the reading remains at 0?
Confirm whether the sample may contain CDOM/fDOM, then check whether the optical window is clean, the sensing area is immersed deeply enough, and zero and slope calibrations are correct. If there is still no response when testing with a standard solution, contact AQUAS.
Q3: What should be done if the reading remains at the upper range?
Possible causes include sample concentration exceeding the measurement range, incorrect calibration, optical window contamination, bubble interference, external light interference, or obstruction of the sensing area. Clean the optical window first and reconfirm standard calibration.
Q4: What should be done if the reading is obviously too high or too low?
Check whether zero and slope calibrations are correct, whether the optical window is dirty, whether the sample is homogeneous, and whether bubbles, bottom sludge, external light, or vessel material are causing interference. Recalibrate if necessary.
Q5: What should be done if the response time becomes slower?
Possible causes include optical window contamination, deposits on the sensing area, unstable sample conditions, or abnormal calibration. Clean the sensing area first and verify the standard-solution reading.
Q6: What should be done if readings differ for the same sample?
Confirm whether the sample is homogeneous, stirring is stable, there is no stratification or sediment, bubbles are not attached, and the calibration procedure is correct.
Q7: What should be done if the reading does not change?
Restore factory settings first and repeat the calibration procedure. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.
Q8: Do air bubbles affect readings?
Yes. Air bubbles may interfere with optical sensing, causing unstable CDOM/fDOM values or calibration errors. Stabilize the water flow, install the analyzer correctly, and avoid dense bubbles attaching to the sensing area.
Q9: Do bottom sludge or sediment affect readings?
Yes. If the analyzer is buried by bottom sludge or the sensing area is blocked by sediment, readings may become abnormal or unstable, or the response may slow down.
Q10: Can external light affect the SMR38?
It may. During laboratory calibration, a black opaque calibration vessel is recommended to reduce interference from external light.
Q11: Can an abnormal LED light source affect readings?
Yes. The analyzer includes LED light-source abnormality diagnostics. If the LED light source is abnormal, readings may drift, become unstable, or show no response. Further inspection should be performed according to the diagnostic result.
Q12: Does a difference between field readings and laboratory values indicate device failure?
Not necessarily. Real-time monitoring instruments are directly exposed to field conditions, while laboratory samples typically undergo sampling, storage, pretreatment, and standardized procedures. Differences in test conditions and timing may cause value differences.
Q13: When should AQUAS technical support be contacted?
If the device still cannot be read or the readings remain unresponsive after checking power supply, wiring, Modbus ID, communication settings, standard-solution testing, calibration, optical-window cleaning, and installation, contact AQUAS technical support and provide site photos, wiring photos, Modbus settings, standard-solution test records, and reading screenshots.
- Abnormal Readings and Troubleshooting
-
Q8
Maintenance and Consumables Questions
H- Maintenance and Consumables Questions
Q1: Is regular maintenance required?
Yes. To keep the analyzer in optimal operating condition, regular maintenance and cleaning are required, especially in environments where grease, sediment, sludge, or biofouling can easily accumulate.
Q2: How often should routine maintenance be performed?
Routine maintenance is recommended at least once per week. The actual frequency can be adjusted according to site water quality, contamination level, and monitoring environment.
Q3: What does routine maintenance include?
Routine maintenance is recommended to include:
Cleaning the analyzer surface and sensing area
Checking battery level, if it is a battery-powered model
Checking consumable status
Checking whether cables and mounting fixtures are secure
Confirming stable data transmission
Q4: How often should periodic maintenance be performed?
Periodic maintenance is recommended at least once per month.
Q5: What does periodic maintenance include?
Periodic maintenance includes trend calibration and standard calibration. Long-term data trends can be used to determine whether the instrument condition needs adjustment.
Q6: What consumables are used?
The main consumables include standard solution and pure water. The standard solution listed in the manual is SL38-1 CDOM standard solution at 1500 μg/L.
Q7: What should be noted during cleaning?
Use a soft cloth, pure water, or neutral detergent to clean the analyzer surface and sensing area. Avoid using hard objects that could scratch the optical window or damage the sensing end.
Q8: Why must the optical window be kept clean?
Because the analyzer uses optical fluorescence measurement, sludge, biofilm, grease, or sediment blocking the optical window may cause reading drift, instability, or slower response.
Q9: Is recalibration required after long-term use?
Determine whether recalibration is required based on field data trends, standard-solution test results, and periodic maintenance conditions.
Q10: How should it be stored when not used for an extended period?
Clean the analyzer thoroughly and store it properly to prevent long-term contamination buildup on the optical window or impact damage.
- Maintenance and Consumables Questions