SMR27 Chloride (Cl-) Analyzer

Frequently Asked Questions

  • A
    1. Product Overview
      Q1: What is this product?
      SMR27 is an AQUAS chloride ion (Cl⁻) analyzer designed for real-time monitoring of chloride ion concentration in water. Measurement data is output via RS485 digital communication.

      Q2: What parameters can it measure?
      SMR27 primarily measures chloride ion (Cl⁻) concentration and can also read temperature data. It supports temperature compensation, remote configuration, calibration, and fault diagnosis.

      Q3: What is the measurement principle?
      SMR27 uses an ion-selective electrode (ISE) method. The membrane at the electrode tip selectively allows specific ions to pass through, generating an electrochemical potential. The potential difference is then converted into analyte concentration according to the Nernst equation.

      Q4: What applications is it suitable for?
      SMR27 is suitable for drinking water, surface water, groundwater, industrial processes, water treatment, wastewater, and related applications.

      Q5: Is a control box required?
      Not necessarily. SMR27 features an integrated compact housing with a built-in transmitter and sensor. Its rugged IP68 submersible, corrosion-resistant enclosure allows direct installation according to site requirements.

      Q6: Does it support PLC or HMI integration?
      Yes. SMR27 uses the standard Modbus RTU protocol and can connect to PLCs, HMIs, or AQUAS data loggers via RS485.

      Q7: Does it have explosion-proof certification?
      Yes. SMR27 is certified to IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga explosion-proof standards.

      Q8: Does it support automatic cleaning?
      Yes. SMR27 can be equipped with an automatic cleaning brush to reduce maintenance caused by bubbles, contaminants, or deposits.

      Q9: Does the electrode include a protective design?
      Yes. The SMR27 electrode includes an anti-scratch metal membrane protection design to reduce the risk of scratches or damage during field operation.

      Q10: What is the electrode service life?
      The manual states that the electrode service life can exceed 12 months. Actual service life depends on water quality, contamination level, maintenance frequency, and operating environment.

      Q11: Is it suitable for outdoor sites without utility power?
      Yes. SMR27 features an ultra-low-power design and can be powered by batteries, solar power, or similar sources for outdoor or remote monitoring applications.

      Q12: What is the maximum transmission distance?
      According to the manual, SMR27 uses a 1000 N high-strength Kevlar cable and supports a maximum transmission distance of up to 1.2 km.
  • B
    1. Installation
      Q1: What is the operating voltage?
      SMR27 operates at 5–12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.

      Q2: What installation methods are available?
      SMR27 supports immersion, flow-through, and pipe-insertion installation.

      Q3: What should be considered during installation?
      Avoid pulling or damaging the connection cable during installation. Ensure sufficient cable length and secure it properly. Install the analyzer where the water flow is stable to obtain better measurement results.

      Q4: How deep should the sensing surface be immersed?
      During testing, the sensing surface should be fully immersed at least 10 cm below the water surface.

      Q5: How long does it take to stabilize after installation?
      After installation, wait at least 5 minutes for the analyzer to stabilize. Actual stabilization time may vary depending on sample characteristics and installation conditions.

      Q6: Can it be installed close to the bottom of the tank?
      This is not recommended. The analyzer should not directly contact or be installed too close to the tank bottom, as obstruction of the sensing area, sludge, or sediment may cause abnormal measurements.

      Q7: Should the analyzer be fixed vertically in the field?
      Yes. For field installation, the analyzer should be fixed vertically. Use a protective pipe and mounting clamps to secure the unit and prevent impact or misalignment.

      Q8: What should be considered for outdoor or high-load installation sites?
      For outdoor or high-load environments, use AQUAS dedicated protective covers and fixtures to improve overall protection and measurement stability.

      Q9: How should SMR27 be set up for laboratory testing?
      Connect the analyzer to a computer, secure it vertically on a stand, insert it into the calibration vessel, remove the electrode protection bottle, and fully immerse the electrode sensing area in the sample water. Avoid contact with the bottom or sides of the vessel.

      Q10: Is stirring required during laboratory testing?
      It is recommended to place a 2.5 cm magnetic stir bar at the bottom of the vessel and use a magnetic stirrer to maintain stable rotation at approximately 300 rpm. This helps prevent sample stratification and sediment from affecting the results.

      Q11: Can it be installed in a flow-through holder?
      Yes. The manual lists the FTC27 series flow-through holder and PIF27 pipe-insertion holder, which can be selected according to flow cell, pipeline, or fixed insertion requirements.
  • C
    1. Communication and Wiring
      Q1: What communication interface is used?
      The analyzer uses an RS485 communication interface.

      Q2: What communication protocol is used?
      It uses the Modbus RTU communication protocol.

      Q3: What are the default communication settings?
      The common communication settings are as follows:
      Communication interface: RS485
      Baud rate: 19200 bps
      Data length: 8 bit
      Parity: None
      Stop bit: 1 bit
      Communication mode: RTU
      The device supports 9600 and 19200 bps.

      Q4: What is the default device address?
      The factory default device address is 01. If the device address is changed, the address setting in the PLC, HMI, host, or reading software must be updated accordingly.

      Q5: What is the RS485 wiring configuration?
      The SMR male connector wiring is as follows:
      PIN 1: GND
      PIN 2: RS485-
      PIN 3: RS485+
      PIN 4: 5–12 VDC
      PIN 5: Not used
      The typical new cable color assignment is: Black = GND, White = RS485-, Blue = RS485+, Red = 5–12 VDC.

      Q6: If AO analog output is available, what is PIN 5?
      For the AO version, PIN 5 is the AO analog output wire and is gray. For standard non-AO versions, PIN 5 is not used.

      Q7: Can it be connected directly to a computer?
      Yes. Use the CAB06 USB-to-RS485 cable to connect the analyzer to a computer, then read values, perform calibration, and configure parameters using a Modbus tool or AQCFG.

      Q8: Can it be connected to an AQUAS data logger?
      Yes. SMR27 can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data logging and remote monitoring according to site requirements.

      Q9: What should be checked first if the PLC cannot read values?
      Check the following in order:
      1. Whether the analyzer can be connected to a PC using CAB06 and read via Modbus
      2. Whether the communication settings of the PLC and SMR27 match
      3. Whether the Modbus ID is correct and whether there is an ID conflict
      4. Whether the PLC Modbus address is correct
      5. Whether the power supply is stable at 5–12 VDC
      6. Whether the RS485 wiring is correct
      7. Whether the wiring has been completed according to the appendix wiring diagram

      Q10: What happens if RS485 D+ / D- are reversed?
      If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
  • D
    1. Modbus Reading
      Q1: Which address should be read for the Cl⁻ real-time value?
      The Cl⁻ main parameter value can be read from PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with a floating-point data type.

      Q2: How can the real-time temperature value be read?
      The real-time temperature value can be read from PLC address 2, HEX address 0x0001, Function Code 04, Length 2 Words, with a floating-point data type.

      Q3: How can the Cl⁻ mV real-time value be read?
      The Cl⁻ mV real-time value can be read from PLC address 49, HEX address 0x0030, Function Code 04, Length 2 Words, with a floating-point data type.

      Q4: How can the device model be read?
      The device model can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with an ASCII string data type.

      Q5: How can the location, area, or zone name be read?
      The location, area, or zone name can be read from PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with a UTF-16BE string data type.

      Q6: How can latitude and longitude be read?
      Latitude and longitude can be read from PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with a floating-point data type.

      Q7: How can the cleaning brush be activated?
      If the device is equipped with a cleaning brush, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.

      Q8: How should floating-point data be decoded?
      Floating-point values follow the IEEE 754 single-precision format and occupy 4 bytes. If the decoded value is abnormal, check whether byte order or word order needs to be swapped.

      Q9: What should be noted about Modbus CRC?
      Modbus RTU uses CRC16. During transmission, the low byte is sent first and the high byte last.

      Q10: What do Function Codes 03, 04, and 10 mean?
      Function Code 03: Read Holding Registers
      Function Code 04: Read Input Registers
      Function Code 10: Write Multiple Holding Registers

  • E
    1. Calibration
      Q1: Is calibration required before first use?
      In general, no. SMR27 is calibrated before shipment. For initial use, measurements can begin after correct installation according to the manual. If accuracy needs to be verified before use, standard solution calibration can be performed.

      Q2: When is recalibration required?
      Recalibration is recommended when field readings differ significantly from laboratory results, when readings drift after long-term use, when accuracy needs to be verified after maintenance or cleaning, after electrode replacement, or when field sample conditions differ significantly from standard solution conditions.

      Q3: What calibration items are available?
      SMR27 calibration includes temperature calibration, standard solution calibration, field calibration, background value subtraction, and AO calibration for the optional AO version.

      Q4: What should be done before calibration?
      Before calibration, thoroughly clean the analyzer. Remove the electrode protection cap, rinse the surface with tap water, and, if necessary, clean with a soft brush and neutral detergent. Dry the analyzer body as much as possible and shake excess water from the electrode.

      Q5: What tools are required for calibration?
      Prepare the following:
      Calibrated reference thermometer
      Calibration cup
      Magnetic stir bar and magnetic stirrer
      Ultrapure water

      Q6: What standard solutions are required for calibration?
      The following can be used:
      SL27-1 chloride ion (Cl⁻) standard solution, 50 mg/L, 500 ml
      SL27-2 chloride ion (Cl⁻) standard solution, 500 mg/L, 500 ml
      SL27-3 four-in-one calibration solution, low concentration, 500 ml
      SL27-4 four-in-one calibration solution, high concentration, 500 ml

      Q7: How is temperature calibration performed?
      Pour ultrapure water or tap water into the calibration cup, then place the analyzer and reference thermometer into the water. Wait at least 2 minutes and confirm that the real-time temperature value is stable. Write the temperature correction value as “real-time temperature value − thermometer reading.”

      Q8: Which address should the temperature correction value be written to?
      The temperature correction value can be written to PLC address 28679, HEX address 0x7006, Function Code 10, Length 2 Words, with a floating-point data type.

      Q9: How many calibration points are required for standard solution calibration?
      SMR27 chloride ion standard solution calibration uses a two-point calibration with 50 mg/L and 500 mg/L standard solutions.

      Q10: How is the first-point standard solution calibration performed?
      Place the 50 mg/L standard solution and magnetic stir bar into the calibration vessel. Immerse the analyzer to the appropriate depth and start the magnetic stirrer at a constant speed, avoiding excessive vortexing and dense bubbles. Wait approximately 15 minutes for the Cl⁻ mV real-time value to stabilize, then enter 1 in the Cl⁻ mV first-point calibration value to perform calibration.

      Q11: Which address should the Cl⁻ mV first-point calibration value be written to?
      The Cl⁻ mV first-point calibration value can be written to PLC address 561, HEX address 0x0230, Function Code 10, Length 2 Words, with a floating-point data type.

      Q12: How is the second-point standard solution calibration performed?
      After completing the first-point calibration, remove the analyzer, rinse the electrode with ultrapure water, and dry it. Then repeat the same calibration procedure using the 500 mg/L standard solution. Enter 1 in the Cl⁻ mV second-point calibration value to perform the second-point calibration.

      Q13: Which address should the Cl⁻ mV second-point calibration value be written to?
      The Cl⁻ mV second-point calibration value can be written to PLC address 562, HEX address 0x0231, Function Code 10, Length 2 Words, with a floating-point data type.

      Q14: Which value should be read to confirm stability during standard solution calibration?
      During standard solution calibration, read the Cl⁻ mV real-time value from PLC address 49, HEX address 0x0030, Function Code 04, Length 2 Words.

      Q15: How long should the analyzer stabilize during standard solution calibration?
      After stirring begins, wait approximately 15 minutes until the Cl⁻ mV real-time value becomes stable before performing calibration.

      Q16: How deep should the electrode be immersed during calibration?
      When immersing the electrode in the standard solution, ensure that the electrode tip is at least 3 cm below the liquid surface and does not touch the bottom of the calibration cup.

      Q17: What should be noted when cleaning the Cl⁻ electrode?
      Rinse the electrode with ultrapure water and dry it with lens tissue. Avoid residual standard solution, contaminants, or water droplets that may affect the next calibration step.

      Q18: Why should large vortices and dense bubbles be avoided during calibration?
      Large vortices and dense bubbles may interfere with electrode sensing and reading stability, causing unstable Cl⁻ mV or concentration values and affecting calibration results.
  • F
    1. Field Calibration and Background Value Subtraction
      Q1: When is field calibration required?
      Field calibration may be performed when real-time field readings differ from laboratory analysis results, provided that cleaning, installation, and standard solution calibration have been confirmed to be normal.

      Q2: How is field calibration performed?
      Install the analyzer correctly and place it into a mixed solution with a known concentration. Immerse the analyzer to the appropriate depth. After the Cl⁻ real-time value stabilizes, enter the field calibration value, then read the Cl⁻ real-time value again to confirm whether calibration was successful.

      Q3: Which address should the field calibration value be written to?
      The field calibration value can be written to PLC address 13058, HEX address 0x3301, Function Code 10, Length 2 Words, with a floating-point data type.

      Q4: Which address should be read for the Cl⁻ real-time value during field calibration?
      The Cl⁻ real-time value can be read from PLC address 1, HEX address 0x0000, Function Code 04, Length 2 Words, with a floating-point data type.

      Q5: Can the cleaning brush be activated during field calibration?
      Yes. The cleaning brush can be activated to remove bubbles attached near the sensing area or optical window to prevent abnormal readings.

      Q6: What is background value subtraction?
      Background value subtraction is a correction function for background interference in field samples. It can be used to subtract the known background contribution from the Cl⁻ reading.

      Q7: When is background value subtraction not recommended?
      If the concentration trend of the field sample has not been verified through laboratory testing, background value subtraction is not recommended.

      Q8: Which address should the background correction value be written to?
      The background correction value can be written to PLC address 28678, HEX address 0x7005, Function Code 10, Length 2 Words, with a floating-point data type.

      Q9: How can successful background value subtraction be confirmed?
      Read the Cl⁻ real-time value again to confirm whether the reading has been corrected according to the background value, and compare it with a known concentration or laboratory analysis result.
  • G
    1. AO Analog Output
      Q1: Does it support AO 4–20 mA output?
      With the optional AO version, the AO output range can be configured so that 4–20 mA corresponds to the specified Cl⁻ measurement range.

      Q2: What equipment is required for AO calibration?
      AO calibration requires the SMR instrument, suitable power wiring, a USB-to-RS485 converter, preferably CAB06, and a digital multimeter or current calibrator.

      Q3: Are reagents required for AO calibration?
      No. AO calibration is an analog output current calibration and does not require reagents.

      Q4: What is the AO1 calibration wiring?
      AO1 wiring is as follows:
      Brown: AO1+, connect to the meter mA+ terminal
      White: AO1−, connect to the meter COM / mA− terminal

      Q5: What is the AO2 calibration wiring?
      AO2 wiring is as follows:
      Blue: AO2+, connect to the meter mA+ terminal
      Black: AO2−, connect to the meter COM / mA− terminal

      Q6: What mode should the meter be set to during AO calibration?
      Set the meter to DC mA mode and ensure the test leads are inserted into the correct mA measurement terminals.

      Q7: What does a negative AO meter reading indicate?
      If the meter displays a negative value, the polarity may be reversed. Swap the meter leads and measure again.

      Q8: Which addresses should AO Min / Max be written to?
      The AO output range can be written to:
      Min: PLC address 31241, HEX address 0x7A08
      Max: PLC address 31242, HEX address 0x7A09

      Q9: Which address should the AO current offset be written to?
      The AO current offset can be written to PLC address 31236, HEX address 0x7A03, Function Code 10, with a floating-point data type.

      Q10: How can normal AO output be confirmed during calibration?
      First set the Min and Max range, then read the real-time value of the measured parameter and calculate the theoretical current within the 4–20 mA range. If the meter reading differs from the theoretical current, write a current offset correction, then confirm that the meter current has been calibrated to the theoretical value.
  • H
    1. Abnormal Readings and Troubleshooting
      Q1: What should be done if the reading keeps fluctuating?
      First clean the sensing area. Confirm that the installation depth is sufficient, the device is firmly fixed, and there are no bubbles, sludge, or contaminants interfering with the measurement. Also check whether the power supply and RS485 communication are normal.

      Q2: What should be done if the reading remains at 0?
      Confirm that the sample actually contains chloride ions, then check whether the electrode is clean, whether it is immersed to a sufficient depth, and whether standard solution calibration and field calibration are correct. If there is still no response when testing with a standard solution, contact AQUAS.

      Q3: What should be done if the reading remains at the upper range?
      Possible causes include Cl⁻ concentration exceeding the measurement range, incorrect calibration, electrode contamination, bubble interference, obstruction of the sensing area, or electrode aging. Clean the analyzer first and reconfirm standard solution calibration.

      Q4: What should be done if the reading is obviously too high or too low?
      Check whether standard solution calibration is correct, whether the electrode is dirty or damaged, whether the sample is homogeneous, and whether bubbles or sludge are interfering. Recalibrate or perform field calibration if necessary.

      Q5: What should be done if the response time becomes slower?
      This may be caused by electrode aging, electrode contamination, deposits on the sensing area, unstable sample conditions, or degraded electrode performance. Clean the electrode first and verify the standard solution reading. Replace the electrode if necessary.

      Q6: What should be done if readings from the same sample are inconsistent?
      Confirm that the sample is homogeneous, stirring is stable, there is no stratification or sediment, no bubbles are attached, and the calibration procedure is correct.

      Q7: What should be done if the reading does not change?
      Restore factory settings and perform the calibration procedure again. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.

      Q8: Do bubbles affect readings?
      Yes. Bubbles may interfere with sensor readings, causing unstable Cl⁻ values or calibration errors. Maintain stable flow, install the analyzer correctly, and avoid dense bubbles attaching to the sensing area.

      Q9: Do sludge or sediment affect readings?
      Yes. If the analyzer is buried in sludge or the sensing area is blocked by sediment, readings may become abnormal, unstable, or slower to respond.

      Q10: Do cracks or wear on the electrode affect readings?
      Yes. Cracks, damage, or wear on the electrode may cause unstable readings, slower response, or calibration failure. Replace the electrode immediately.

      Q11: Does a difference between field readings and laboratory results indicate device failure?
      Not necessarily. Real-time monitoring instruments are directly exposed to field conditions, while laboratory samples usually undergo sampling, storage, pretreatment, and standardized procedures. Differences in test conditions and timing may lead to different results. Field calibration can improve consistency.

      Q12: When should AQUAS technical support be contacted?
      If readings still cannot be obtained or there is no response after checking power supply, wiring, Modbus ID, communication settings, standard solution testing, calibration, cleaning, electrode condition, and installation, contact AQUAS technical support and provide site photos, wiring photos, Modbus settings, standard solution test records, and reading screenshots.
  • I
    1. Maintenance and Consumables
      Q1: Is regular maintenance required?
      Yes. Regular maintenance and cleaning are required to keep the analyzer in optimal operating condition, especially in environments where grease, sediment, or other contaminants can easily accumulate.

      Q2: How often should routine maintenance be performed?
      Routine maintenance is recommended at least once per week. The actual frequency may be adjusted according to water quality, contamination level, and monitoring environment.

      Q3: What tasks are included in routine maintenance?
      Routine maintenance should include:
      Cleaning the analyzer surface and sensing area
      Checking battery level for battery-powered models
      Checking consumable condition
      Checking whether cables and mounting fixtures are secure
      Confirming stable data transmission

      Q4: How often should periodic maintenance be performed?
      Periodic maintenance is recommended at least once per month.

      Q5: What is included in periodic maintenance?
      Periodic maintenance includes trend calibration and standard solution calibration. Long-term data trends can be used to determine whether instrument adjustment is required.

      Q6: What consumables are required?
      Main consumables include Cl⁻ standard solutions, four-in-one calibration solutions, ultrapure water, 3M KCl storage solution, and Cl⁻ electrodes and reference electrodes that may need replacement depending on use conditions.

      Q7: When should the electrode be replaced?
      Electrode replacement is recommended when electrode response becomes slow, readings are unstable, electrode slope decreases significantly, or physical damage such as cracks or breakage is observed.

      Q8: How should the Cl⁻ electrode and reference electrode be stored during long-term non-use?
      When the electrodes will not be used for an extended period, clean the sensor thoroughly and store the Cl⁻ electrode and reference electrode immersed in 3M KCl storage solution.

      Q9: Is calibration required after electrode replacement?
      Yes, it is recommended. After replacing the electrode, confirm that all connections and installation are correct, then perform calibration again to ensure accurate measurement results.

      Q10: What should be noted during cleaning?
      Use a soft cloth, ultrapure water, or neutral detergent to clean the analyzer surface and sensing area. Avoid using hard objects that may scratch the electrode or damage the sensing tip.