SMR11 Salinity and Temperature Analyzer
Frequently Asked Questions
-
Q1
Basic Product Questions
A- Basic Product Questions
Q1: About the Product?
This is the AQUAS salinity and temperature analyzer, designed for real-time monitoring of salinity and temperature in water. Measurement data is output via RS485 digital communication.
Q2: What parameters can it measure?
It mainly measures salinity and temperature, and includes temperature compensation.
Q3: What is the measurement principle?
It uses a two-electrode measurement principle. The analyzer measures the resistance of the water sample between two opposing electrodes, calculates conductivity from the reciprocal of resistance and the cell constant, and then converts the result into salinity.
Q4: What applications is it suitable for?
It is suitable for drinking water, surface water, groundwater, industrial processes, water treatment, wastewater, and related applications.
Q5: Is a control box required?
Not necessarily. The analyzer features an integrated compact enclosure with a built-in transmitter and sensor. Its rugged IP68 submersible, corrosion-resistant housing allows direct installation according to site requirements.
Q6: Does it support PLC or HMI integration?
Yes. It uses the standard Modbus RTU communication protocol and can be connected to a PLC, HMI, or AQUAS data logger via RS485.
Q7: Does it have explosion-proof certification?
Yes. It is certified for IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS explosion protection.
Q8: Is it suitable for outdoor sites without mains power?
Yes. Its ultra-low-power design supports battery or solar power supply for outdoor or remote monitoring applications.
Q9: Does it have self-diagnostic functions?
Yes. It includes internal status monitoring such as temperature and humidity, as well as abnormal salinity electrode diagnostics.
Q10: Does it comply with EPA automatic monitoring methods?
It provides temperature compensation and measurement functions and is indicated as compliant with the EPA-published automatic monitoring facility method NIEA W218.51C.
- Basic Product Questions
-
Q2
Installation Questions
B- Installation Questions
Q1: What is the operating voltage?
The operating voltage is 5-12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.
Q2: What installation methods are available?
Immersion, flow-through, or pipe-insertion installation methods may be used.
Q3: What should be noted during installation?
Avoid pulling or damaging the connection cable during installation. Ensure sufficient cable length and secure it properly. Installation in a location with stable flow is recommended for better measurement performance.
Q4: How deep should the sensing surface be submerged?
During testing, the sensing surface should be fully submerged at least 10 cm below the water surface.
Q5: How long does it take to stabilize after installation?
After installation, wait at least 5 minutes for the analyzer to reach a stable state. The actual stabilization time may vary depending on sample characteristics and installation conditions.
Q6: Can it be installed close to the tank bottom?
This is not recommended. The analyzer should not directly contact or be installed too close to the tank bottom, as shielding of the sensing area, sludge, or sediment may cause abnormal measurements.
Q7: Should the analyzer be fixed vertically during field installation?
Yes. During field installation, the analyzer should be fixed vertically. Use a protective pipe and mounting clamps to secure the device and prevent impact or misalignment.
Q8: What should be considered for outdoor or high-load installations?
For outdoor or high-load sites, using AQUAS dedicated protective covers and fixtures is recommended to improve overall protection and measurement stability.
Q9: How should the SMR11 be set up for laboratory testing?
Connect the analyzer to a computer, fix it vertically on a stand, insert it into the calibration container, remove the electrode protection bottle, and fully immerse the electrode sensing area in the sample water. Avoid contact with, or close proximity to, the container bottom.
Q10: Is stirring required during laboratory testing?
It is recommended to place a 2.5 cm magnetic stir bar at the bottom of the container and use a magnetic stirrer to maintain stable rotation at approximately 300 rpm, preventing sample stratification and sediment from affecting the results.
Q11: Can it be installed in a flow-through mounting holder?
Yes. The manual lists FTC11 series flow-through holders and PIF11 pipe-insertion holders, which can be selected according to site requirements for flow cells, piping, or fixed insertion installation.
- Installation Questions
-
Q3
Communication and Wiring Questions
C- Communication and Wiring Questions
Q1: What communication interface is used?
It uses an RS485 communication interface.
Q2: What communication protocol is used?
The SMR11 uses the Modbus RTU communication protocol.
Q3: What are the default communication settings?
Common communication settings are as follows:
Communication interface: RS485
Baud rate: 19200 bps
Data length: 8 bit
Parity: None
Stop bit: 1 bit
Communication mode: RTU
The device supports 9600 and 19200 bps.
Q4: What is the default device address?
The factory default device address is 01. If the device address is changed, the corresponding device address in the PLC, HMI, host, or reading software must also be updated.
Q5: What is the RS485 wiring configuration?
The SMR male connector wiring is as follows:
PIN 1:GND
PIN 2:RS485-
PIN 3:RS485+
PIN 4:5~12V DC
PIN 5: Not used
Typical new cable colors are: black for GND, white for RS485-, blue for RS485+, and red for 5-12 VDC.
Q6: If AO analog output is included, what is PIN 5?
For the AO version, PIN 5 is the AO analog output wire, typically gray. For standard non-AO versions, PIN 5 is not used.
Q7: Can it be connected directly to a computer?
Yes. It can be connected to a computer using the CAB06 USB to RS485 Cable. Values can be read, calibration can be performed, and parameters can be configured using a Modbus tool or AQCFG.
Q8: Can it be connected to an AQUAS data logger?
Yes. It can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data logging and remote monitoring according to site requirements.
Q9: What should be checked first if the PLC cannot read SMR11 values?
Check the following in sequence:
1. Whether the device can be connected to a PC using CAB06 and read via Modbus
2. Whether the PLC and SMR11 communication settings match
3. Whether the Modbus ID is correct and whether there is any ID conflict
4. Whether the PLC Modbus address is correct
5. Whether the power supply is stable at 5-12 VDC
6. Whether the RS485 wiring is correct
7. Whether the wiring has been completed according to the appendix wiring diagram
Q10: What happens if RS485 D+ / D- are reversed?
If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
- Communication and Wiring Questions
-
Q4
Modbus Reading Questions
D- Modbus Reading Questions
Q1: Which address should be read for salinity values?
The main salinity parameter value can be read from PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with a floating point data type.
Q2: How can the real-time temperature value be read?
The real-time temperature value can be read from PLC address 2, HEX address 0x0001, Function Code 04, Length 2 Words, with a floating point data type.
Q3: How can the device model be read?
It can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with an ASCII string data type.
Q4: How can the location, area, or sub-area name be read?
It can be read from PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with a UTF-16BE string data type.
Q5: How can longitude and latitude be read?
They can be read from PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with a floating point data type.
Q6: How can the cleaning brush be activated?
If the device is equipped with a cleaning brush, write to PLC address 5, HEX address 0x0004, Function Code 10. A data value of 0 indicates standby, and 1 indicates activation.
Q7: How should floating point data be decoded?
SMR11 floating point data follows the IEEE 754 single-precision format and occupies 4 Bytes. If the decoded value is abnormal, verify whether the high/low bytes or word order need to be swapped.
Q8: What should be noted about Modbus CRC?
Modbus RTU uses CRC16. During transmission, it follows Low Byte First and High Byte Last, meaning the low byte is transmitted before the high byte.
Q9: What do Function Codes 03, 04, and 10 mean?
Function Code 03: Read holding registers
Function Code 04: Read input registers
Function Code 10: Write multiple holding registers
- Modbus Reading Questions
-
Q5
Calibration Questions
E- Calibration Questions
Q1: Does the SMR11 need calibration before first use?
In general, no. The device is calibrated before shipment. For first use, measurement can begin after proper installation according to the manual. If accuracy needs to be verified before use, standard-solution calibration may be performed.
Q2: When is recalibration required?
Recalibration is recommended when field readings differ significantly from laboratory measurements, when readings drift after long-term use, when accuracy needs to be verified after maintenance or cleaning, after electrode replacement, or when field sample conditions differ significantly from standard-solution conditions.
Q3: What calibration items does the SMR11 support?
Calibration items include temperature calibration, standard-solution zero calibration, standard-solution slope calibration, field calibration, and AO calibration for the optional AO version.
Q4: What should be done before calibration?
Before calibration, thoroughly clean the analyzer. It is recommended to remove the electrode protective cap, rinse the surface with tap water, use a soft brush and neutral detergent if necessary, dry the analyzer body as much as possible, and shake excess water off the electrode.
Q5: What tools are required for calibration?
Prepare the following:
Calibrated reference thermometer
WTW conductivity meter
Calibration cup
Magnetic stir bar and magnetic stirrer
Ultrapure water
Q6: What standard solutions are required for calibration?
Prepare SL11-1 salinity standard solution, 50 PSU, 500 ml.
Q7: How is temperature calibration performed?
Pour pure water or tap water into the calibration cup, then place the analyzer and the reference thermometer into the cup. Wait at least 2 minutes and confirm that the temperature is stable. Write the temperature correction value as "real-time temperature value - thermometer reading."
Q8: Which address should the temperature correction value be written to?
The temperature correction value can be written to PLC address 714, HEX address 0x02C9, Function Code 10, Length 2 Words, with a floating point data type.
Q9: What water should be used for zero calibration?
Zero calibration must use compliant pure water, such as Type II grade pure water with conductivity below 1.0 µS/cm, or a higher-grade pure water.
Q10: How is zero calibration performed?
Place pure water and a magnetic stir bar into the calibration container. Immerse the analyzer to the appropriate depth and start the magnetic stirrer at a constant speed, avoiding large vortices and dense bubbles. Then read the zero calibration value, enter 0 as the zero calibration value, enter 1 to execute zero calibration, and finally check the real-time salinity value to confirm completion.
Q11: Which address should the zero calibration value be written to?
The zero calibration value can be written to PLC address 519, HEX address 0x0206, Function Code 10, with a floating point data type.
Q12: Which address should be written to execute zero calibration?
Write 1 to PLC address 518, HEX address 0x0205, Function Code 10, to execute zero calibration.
Q13: How is slope calibration performed?
Place the salinity standard solution and a magnetic stir bar into the calibration container. Immerse the analyzer to the appropriate depth and start the magnetic stirrer at a constant speed. After the real-time salinity value stabilizes, enter the standard-solution concentration as the slope calibration value, then enter 1 to execute slope calibration.
Q14: Which address should the slope calibration value be written to?
The slope calibration value can be written to PLC address 521, HEX address 0x0208, Function Code 10, with a floating point data type.
Q15: Which address should be written to execute slope calibration?
Write 1 to PLC address 520, HEX address 0x0207, Function Code 10, to execute slope calibration.
Q16: What concentration is recommended for SMR11 slope calibration?
Select the standard solution concentration according to the target measurement range. The manual lists a 50 PSU salinity standard solution.
Q17: What are Zero lock and Span lock?
Zero lock and Span lock are locked values in the calibration logic, used for the zero reference and slope reference. Zero lock can be read from PLC address 733, HEX address 0x02DC; Span lock can be read from PLC address 734, HEX address 0x02DD.
Q18: What should be noted before field calibration?
After the analyzer is installed on site, wait for the reading to stabilize. The target calibration value should preferably refer to laboratory test results. After successful calibration, the slope value will be updated.
Q19: What parameter is updated by field calibration?
After successful field calibration, the slope value is updated to correct deviations caused by differences between field sample conditions and standard-solution conditions.
Q20: Why should large vortices and dense bubbles be avoided during calibration?
Large vortices and dense bubbles may interfere with electrode readings, causing unstable salinity values and affecting calibration results.
- Calibration Questions
-
Q7
AO Analog Output Questions
F- AO Analog Output Questions
Q1: Does the SMR11 support AO 4-20 mA output?
If the optional AO version is selected, the AO output range can be configured so that 4-20 mA corresponds to the specified salinity measurement range.
Q2: What equipment is required for AO calibration?
AO calibration requires the SMR instrument, suitable power wiring, a USB-to-RS485 converter (CAB06 recommended), and a digital multimeter or current calibrator.
Q3: Are reagents required for AO calibration?
No. AO calibration is analog output current calibration and does not require reagents.
Q4: What is the wiring for AO1 calibration?
AO1 wiring is as follows:
Brown: AO1+, connect to multimeter mA+
White: AO1-, connect to multimeter COM / mA-
Q5: What is the wiring for AO2 calibration?
AO2 wiring is as follows:
Blue: AO2+, connect to multimeter mA+
Black: AO2-, connect to multimeter COM / mA-
Q6: What mode should the multimeter be set to during AO calibration?
Set the multimeter to DC mA mode and ensure the test leads are inserted into the correct mA measurement terminals.
Q7: What does a negative reading on the AO multimeter indicate?
If the multimeter shows a negative value, the polarity may be reversed. Swap the multimeter leads and measure again.
Q8: Which addresses should AO Min / Max be written to?
The AO output range can be written to:
Min: PLC address 31241, HEX address 0x7A08
Max: PLC address 31242, HEX address 0x7A09
Q9: Which address should the AO current offset be written to?
The AO current offset can be written to PLC address 31236, HEX address 0x7A03, Function Code 10, with a floating point data type.
Q10: How can normal output be confirmed during AO calibration?
First set the Min and Max range, then read the real-time value of the measured parameter and calculate the theoretical current within the 4-20 mA range. If there is a deviation between the multimeter reading and the theoretical current, write the current offset correction, then confirm again that the multimeter current has been calibrated to the theoretical value.
- AO Analog Output Questions
-
Q8
Abnormal Readings and Troubleshooting
F- Abnormal Readings and Troubleshooting
Q1: What should be done if the reading keeps fluctuating?
First clean the sensing area, confirm that the installation depth is sufficient, the device is securely fixed, and there are no bubbles, sludge, or contaminants interfering on site. Also check whether the power supply and RS485 communication are normal.
Q2: What should be done if the reading remains at 0?
Confirm that the sample actually has salinity or conductivity, then check whether the electrode is clean, whether it is submerged deep enough, and whether zero and slope calibrations are correct. If there is still no response when testing with a standard solution, contact AQUAS.
Q3: What should be done if the reading remains at the upper limit?
Possible causes include sample salinity exceeding the measurement range, calibration error, electrode contamination, bubble interference, or blockage of the sensing area. Clean the analyzer first and reconfirm the calibration.
Q4: What should be done if the reading is significantly high or low?
Check whether zero and slope calibrations are correct, whether the electrode is dirty or damaged, whether the sample is homogeneous, and whether bubbles or sludge are interfering. Recalibrate if necessary.
Q5: What should be done if the response time becomes slower?
Possible causes include electrode contamination, deposits on the sensing area, unstable sample conditions, or degraded electrode performance. Clean the electrode first and verify readings with a standard solution. Activate or replace the electrode if necessary.
Q6: What should be done if readings are inconsistent for the same sample?
Confirm that the sample is homogeneous, stirring is stable, there is no stratification or sediment, no bubbles are attached, and the calibration procedure is correct.
Q7: What should be done if the reading does not change?
Restore factory settings and repeat the calibration procedure first. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.
Q8: Can bubbles affect the reading?
Yes. Bubbles may interfere with sensor readings, causing unstable salinity values or calibration errors. Maintain stable flow, install the analyzer correctly, and avoid dense bubbles adhering to the sensing area.
Q9: Can sludge or sediment affect the reading?
Yes. If the analyzer is buried in sludge or the sensing area is blocked by sediment, readings may become abnormal, unstable, or slower to respond.
Q10: Can the electrode be activated?
Yes. The manual recommends the following EC electrode activation procedure: first rinse the electrode surface with pure water, then soak it in 0.15M NaOH for at least 30 minutes, rinse with clean water, and then soak it in 0.15M HCl for at least 30 minutes. The sequence must be alkaline first, then acid, and must not be changed. After activation, calibrate according to the standard procedure and observe whether the values are normal.
Q11: What should be done if readings are still incorrect after electrode activation?
If readings remain incorrect after activation and calibration, electrode replacement is recommended. If the new electrode is still abnormal, contact AQUAS.
Q12: Does a difference between field readings and laboratory values mean the device is faulty?
Not necessarily. Real-time monitoring instruments are directly exposed to field conditions, while laboratory samples typically undergo sampling, storage, pretreatment, and standardized procedures. Differences in test conditions and timing may cause value discrepancies. Field calibration can improve consistency.
Q13: When should AQUAS technical support be contacted?
If readings still cannot be obtained or the analyzer remains unresponsive after checking power supply, wiring, Modbus ID, communication settings, standard-solution testing, calibration, cleaning, electrode activation, and installation, contact AQUAS technical support and provide site photos, wiring photos, Modbus settings, standard-solution test records, and reading screenshots.
- Abnormal Readings and Troubleshooting
-
Q9
Maintenance and Consumables Questions
F- Maintenance and Consumables Questions
Q1: Is regular maintenance required?
Yes. To ensure optimal analyzer operation, regular cleaning and calibration are required, especially in environments where grease, sediment, or other contaminants can easily adhere.
Q2: How often should it be cleaned?
Cleaning approximately once per week is recommended. The frequency may be adjusted based on site water quality conditions.
Q3: How often should calibration solution be used for calibration?
Calibration with calibration solution is recommended approximately every two weeks. The frequency may be adjusted based on site water quality conditions.
Q4: How should it be stored when not used for an extended period?
If it will not be used for an extended period, clean the analyzer thoroughly and store it dry.
Q5: What maintenance cleaning procedure is recommended?
It is recommended to clean the entire unit first with pure water and a soft brush, then rinse with pure water and confirm that the electrode and body are free of dirt. If obvious fouling is present, soak it in 5% dilute sulfuric acid for 15 minutes, then place it in pure water for zero-point verification.
Q6: What reading condition should be met during zero-point verification?
During zero-point verification in pure water, the measured parameter should be less than 5% of full scale. If the reading remains greater than 15% of full scale after acid soaking, return for maintenance is recommended.
Q7: What standard-solution test error is considered acceptable?
When testing with calibration solution, a measured-value error of less than 5% may be considered acceptable and maintenance complete. If the error is between 5% and 20%, single-point slope calibration may be performed. If the error exceeds 20%, confirm whether bubbles are present in the sensing area and whether the standard solution has deteriorated or become contaminated.
Q8: What consumables are required?
The main consumables include salinity standard solution and pure water.
Q9: When should the electrode be replaced?
Electrode replacement is recommended when the response becomes slower, readings are unstable, electrode performance has clearly degraded, or the electrode has physical damage such as cracks, wear, or breakage.
Q10: Can an electrode that has been used for more than one year continue to be used?
If the electrode has been used for more than one year, it may have aged. Even if it can still produce readings, it is recommended to determine whether replacement is needed based on standard-solution calibration and reading stability.
- Maintenance and Consumables Questions