SMR09 Optical Dissolved Oxygen and Temperature Analyzer

Frequently Asked Questions

  • A
    1. Product and Basic Use
      Q1: About the product
      The SMR09 is a Dissolved Oxygen (DO) and temperature analyzer designed for real-time monitoring of dissolved oxygen and temperature in water, with data output via RS485 digital communication.

      Q2: What parameters can it measure?
      It primarily measures dissolved oxygen concentration (DO), dissolved oxygen saturation (DO%), and temperature, and includes temperature compensation.

      Q3: What is the measurement principle?
      It uses optical fluorescence quenching measurement technology in accordance with ISO 17289 and ASTM D888-05. Oxygen molecules diffuse into the sensing membrane cap. After the indicator molecules are excited by blue light, they emit red fluorescence. Oxygen molecules quench the fluorescence, and the fluorescence decay time is inversely proportional to the oxygen concentration in water.

      Q4: What applications is it suitable for?
      It is suitable for drinking water, surface water, groundwater, industrial processes, water treatment, wastewater, and related applications.

      Q5: Is a control box required?
      Not necessarily. The SMR09 features an integrated compact housing with a built-in transmitter and sensor. Its rugged IP68 submersible, corrosion-resistant enclosure allows direct installation according to site requirements.

      Q6: Does it support PLC or HMI integration?
      Yes. The SMR09 uses the standard Modbus RTU protocol and can be connected to PLCs, HMIs, or AQUAS data loggers via RS485.

      Q7: Does it have explosion-proof certification?
      Yes. The SMR09 is certified with IECEx CoC / ATEX Cat. 1, Ex ia IIB T4/T5 Ga/TS explosion-proof approval.

      Q8: Does it have an automatic cleaning function?
      Yes. The SMR09 can be equipped with an automatic cleaning brush to reduce measurement interference caused by bubbles, sludge, or deposits.

      Q9: Can the membrane cap be replaced?
      Yes. The SMR09 uses a replaceable membrane cap design, allowing users to replace the cap according to its consumable condition.

      Q10: What is the service life of the membrane cap?
      According to the manual, the sensing membrane cap can last more than 24 months under suitable conditions. Actual service life varies depending on water quality, contamination level, maintenance frequency, and operating environment.

      Q11: Is it suitable for outdoor sites without mains power?
      Yes. The SMR09 features an ultra-low-power design and can be powered by batteries or solar power for outdoor or remote monitoring applications.
  • B
    1. Installation Questions
      Q1: What is the operating voltage?
      The operating voltage is 5–12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.

      Q2: What installation methods are available?
      Immersion, flow-through, or pipe-insertion installation methods can be used.

      Q3: What should be considered during installation?
      Avoid pulling or damaging the cable during installation. Ensure sufficient cable length and secure the cable properly. Installation in a stable-flow location is recommended for better measurement performance.

      Q4: How deep should the sensing surface be immersed?
      During testing, the sensing surface should be fully immersed at least 10 cm below the water surface.

      Q5: How long does it take to stabilize after installation?
      After installation, the analyzer should generally be allowed to stabilize for at least 5 minutes. For the SMR09 dissolved oxygen analyzer, a minimum waiting time of 10 minutes is recommended to achieve a more stable measurement condition. Actual stabilization time may vary depending on sample characteristics and installation conditions.

      Q6: Can it be installed close to the bottom of the tank?
      It is not recommended. The analyzer should not directly touch or be installed too close to the tank bottom, as obstruction of the sensing area or sediment interference may cause abnormal readings. If necessary, install a dedicated protective cover to prevent the sensing area from contacting the bottom.

      Q7: Should the protective cap be removed before use?
      Yes. Before use, remove the protective cap according to the manual to prevent it from affecting the sensing area and measurement results.

      Q8: Does the analyzer need to be fixed vertically during field installation?
      Yes. For field installation, the analyzer should be fixed vertically. Use a protective pipe and mounting clamps to secure the device and prevent impact or installation displacement.

      Q9: What should be considered for outdoor or high-load installation sites?
      For outdoor or high-load environments, AQUAS dedicated protective covers and mounting fixtures are recommended to improve overall protection and measurement stability.

      Q10: How should the SMR09 be set up for laboratory testing?
      Connect the analyzer to a computer, fix it vertically on a stand, place it into the calibration container, remove the electrode protective bottle, and ensure that the sensing area is fully immersed in the sample water. Avoid contact with or close proximity to the bottom of the container.

      Q11: Is stirring required during laboratory testing?
      It is recommended to place a 2.5 cm magnetic stir bar at the bottom of the container and use a magnetic stirrer to maintain stable rotation at approximately 300 rpm. This helps prevent sample stratification and sediment from affecting the results.
  • C
    1. Communication and Wiring Questions
      Q1: What communication interface is used?
      The RS485 communication interface is used.

      Q2: What communication protocol is used?
      The standard Modbus RTU communication protocol is used.

      Q3: What are the default communication settings?
      Common communication settings are as follows:
      Communication interface: RS485
      Baud rate: 19200 bps
      Data length: 8 bit
      Parity: None
      Stop bit: 1 bit
      Communication mode: RTU
      The device supports 9600 and 19200 bps.

      Q4: Can it be connected directly to a computer?
      Yes. It can be connected to a computer using the CAB06 USB-to-RS485 cable, allowing values to be read, calibration to be performed, and parameters to be configured via Modbus tools or AQCFG.

      Q5: What is the default device address?
      The factory default device address is 01. If the device address is changed, the corresponding address in the PLC, HMI, host, or reading software must also be updated.

      Q6: How should the RS485 wiring be connected?
      The SMR male connector wiring is as follows:
      PIN 1: GND
      PIN 2: RS485-
      PIN 3: RS485+
      PIN 4: 5–12 VDC
      PIN 5: Not used
      The typical new cable color code is: black for GND, white for RS485-, blue for RS485+, and red for 5–12 VDC.

      Q7: Can it be connected to an AQUAS data logger?
      Yes. The SMR09 can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data logging and remote monitoring according to site requirements.

      Q8: What should be checked first if the PLC cannot read values?
      Check the following in order:
      1. Whether values can be read via Modbus using a CAB06 connection to a PC
      2. Whether the communication settings of the PLC and SMR09 are consistent
      3. Whether the Modbus ID is correct and whether there is an ID conflict
      4. Whether the PLC Modbus address is correct
      5. Whether the power supply is stable at 5–12 VDC
      6. Whether the RS485 wiring is correct
      7. Whether the wiring has been completed according to the appendix wiring diagram

      Q9: What happens if RS485 D+ / D- are reversed?
      If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
  • D
    1. Modbus Reading Questions
      Q1: Which address should be read for the real-time DO value?
      The DO real-time value can be read from PLC address 51, HEX address 0x0032, Length 2 Words, Function Code 04, with a floating-point data type.

      Q2: Which address should be read for DO% saturation?
      DO% can be read from PLC address 53, HEX address 0x0034, Length 2 Words, Function Code 04, with a floating-point data type.

      Q3: How can the temperature value be read?
      The real-time temperature value can be read from PLC address 49, HEX address 0x0030, Length 2 Words, Function Code 03, with a floating-point data type.

      Q4: How can the device model be read?
      The device model can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with an ASCII string data type.

      Q5: How can the location, area, or sub-area name be read?
      It can be read from PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with a UTF-16BE string data type.

      Q6: How can latitude and longitude be read?
      They can be read from PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with a floating-point data type.

      Q7: How can the cleaning brush be activated?
      If the device is equipped with a cleaning brush, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.

      Q8: How should floating-point data be decoded?
      SMR09 floating-point values follow the IEEE 754 single-precision floating-point format and occupy 4 Bytes. If the decoded value is abnormal, check whether byte order or word order needs to be swapped.

      Q9: What should be noted about Modbus CRC?
      Modbus RTU uses CRC16. During transmission, the Low Byte is sent first and the High Byte last; that is, the low byte precedes the high byte.

      Q10: What do Function Codes 03, 04, and 10 mean?
      Function Code 03: Read Holding Registers
      Function Code 04: Read Input Registers
      Function Code 10: Write Multiple Holding Registers

  • E
    1. Calibration Questions
      Q1: Is calibration required before first use?
      In general, no. The SMR09 is calibrated before shipment. For first-time use, measurement can begin after correct installation according to the manual. If accuracy needs to be verified before use, standard-solution calibration may be performed.

      Q2: When is recalibration recommended?
      Recalibration is recommended when field readings differ significantly from laboratory results, readings drift after long-term use, accuracy must be confirmed after maintenance or cleaning, or field sample conditions differ significantly from standard-solution conditions.

      Q3: What calibration items are included?
      SMR09 calibration includes temperature verification, zero-point calibration, slope calibration, slope calculation, and cap parameter settings after membrane cap replacement.

      Q4: What should be done before calibration?
      Thoroughly clean the analyzer before calibration. It is recommended to rinse the surface with tap water, use a soft brush and neutral detergent if necessary, and dry all parts of the analyzer body as much as possible.

      Q5: What tools are required for calibration?
      Prepare the following:
      Calibrated reference thermometer
      Calibration cup
      Magnetic stir bar and magnetic stirrer
      Ultrapure water

      Q6: What standard solution is required for calibration?
      SL09-1 Dissolved Oxygen Standard Solution 0%, 500 ml, can be used.

      Q7: How is temperature calibration or verification performed?
      Pour pure water or tap water into the calibration cup and immerse both the analyzer and the reference thermometer in the water. Wait at least 2 minutes after immersion, then check whether the real-time temperature value is stable. If the difference between the analyzer and thermometer readings is too large, contact AQUAS.

      Q8: What should be done if the temperature readings differ significantly?
      During calibration, keep the temperature as stable as possible. In general, the analyzer reading should not differ significantly from the reference thermometer. If the deviation is excessive, contact AQUAS.

      Q9: How is zero-point calibration performed?
      Place the 0% dissolved oxygen standard solution into the calibration container and immerse the analyzer to the appropriate depth. Allow it to stand until the reading stabilizes. Before calibration, read the real-time DO and DO% values to confirm the current status. Then enter the DO% offset to perform zero-point correction, and finally read DO% again to confirm whether calibration was successful.

      Q10: Which address should be read for the real-time temperature value?
      The real-time temperature value can be read from PLC address 49, HEX address 0x0030, Function Code 03, Length 2 Words, with a floating-point data type.

      Q11: Which values should be read during zero-point calibration?
      Before zero-point calibration, read the DO real-time value and DO% real-time value:
      DO: PLC address 51, HEX 0x0032
      DO%: PLC address 53, HEX 0x0034

      Q12: Which address should the DO% offset be written to?
      The DO% offset can be written to PLC address 99, HEX address 0x0062, Function Code 10, Length 2 Words, with a floating-point data type.

      Q13: How is slope calibration performed?
      For slope calibration, first place the instrument in air and read DO% before calibration to confirm the status. After obtaining the DO slope value through slope calculation, write the DO slope value, then read DO% to confirm whether it is close to 100%, and read the DO real-time value to verify the current data.

      Q14: Which address should the DO slope value be written to?
      The DO slope value can be written to PLC address 97, HEX address 0x0060, Function Code 10, Length 2 Words, with a floating-point data type.

      Q15: Why is the sensor placed in air during slope calibration?
      During slope calibration, water-saturated air is used as the 100% dissolved oxygen saturation reference to establish or correct the full-scale signal.

      Q16: What is the slope calculation logic for the SMR09?
      Calibration can be regarded as establishing a linear equation, Y = aX + b. The 0% dissolved oxygen standard solution is used to determine offset b so that the output is 0, while the 100% saturation condition in air is used to determine slope a so that the output returns to 100%.

      Q17: Can the cleaning brush be activated during calibration?
      Yes. The cleaning brush can be activated to remove bubbles attached to the optical window or around the membrane cap, helping prevent abnormal readings.

      Q18: What should be noted when cleaning the electrode?
      The cleaning brush can be activated to remove bubbles attached to the optical window and prevent abnormal readings.

  • F
    1. Abnormal Readings and Troubleshooting
      Q1: What should be done if the reading keeps fluctuating?
      First clean the sensing area, then confirm whether the analyzer is installed at the proper water depth, securely fixed, exposed to excessive turbulence, affected by bubble interference, or whether the signal cable is being interfered with by power cables.

      Q2: What should be done if the reading remains at 0?
      Confirm that the sample and standard solution are normal, then check whether the sensing membrane cap or optical window is clean, whether the sensor is immersed to sufficient depth, and whether calibration is correct. If there is still no response when testing with standard solution, contact AQUAS.

      Q3: What should be done if the reading remains at the upper range?
      Possible causes include sample concentration exceeding the measurement range, calibration error, contamination of the membrane cap or optical window, bubble interference, or obstruction of the sensing area. Clean the analyzer first and reconfirm standard-solution calibration.

      Q4: What should be done if the reading is obviously too high or too low?
      Check whether standard-solution calibration is correct, whether the membrane cap is contaminated or aged, whether the sample is homogeneous, and whether bubbles or bottom sludge are causing interference. Recalibrate if necessary.

      Q5: What should be done if the response time becomes slower?
      Possible causes include membrane cap aging, contamination of the sensing area, unstable sample conditions, or deposits on the optical window. Clean the sensing area first and verify readings with standard solution. Replace the membrane cap if necessary.

      Q6: What should be done if readings differ for the same sample?
      Confirm whether the sample is homogeneous, stirring is stable, stratification or sediment is present, bubbles are attached, and the calibration procedure is correct.

      Q7: What should be done if the reading does not change?
      Restore factory settings first and repeat the calibration procedure. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.

      Q8: Does a difference between field readings and laboratory results indicate equipment failure?
      Not necessarily. Real-time monitoring instruments are directly exposed to field conditions, while laboratory samples are typically collected, stored, pretreated, and measured under standardized procedures. Differences in test conditions and timing may cause measurement deviations.

      Q9: Can bubbles affect SMR09 readings?
      Yes. Bubbles may attach to the optical window or sensing membrane cap, causing unstable DO readings or calibration errors. The impact can be reduced by stabilizing the flow, selecting the proper installation position, or activating the cleaning brush.

      Q10: Can bottom sludge or sediment affect SMR09 readings?
      Yes. If the analyzer is buried in bottom sludge or the sensing area is blocked by sediment, readings may become abnormal, unstable, or slower to respond.

      Q11: When should AQUAS technical support be contacted?
      If readings still cannot be obtained or the device remains unresponsive after checking power supply, wiring, Modbus ID, communication settings, standard-solution testing, calibration, cleaning, membrane cap condition, and installation, contact AQUAS technical support and provide site photos, wiring photos, Modbus settings, standard-solution test records, and reading screenshots.
  • F
    1. Membrane Cap Replacement Questions
      Q1: When should the SMR09 membrane cap be replaced?
      Replace the membrane cap when it is aged, damaged, heavily contaminated, shows slower reading response, cannot achieve reasonable calibration results, or is determined during maintenance inspection to have degraded performance.

      Q2: What should be confirmed before replacing the membrane cap?
      Confirm that the serial number on the information strip of the replacement membrane cap matches the serial number on the outer packaging, and check that an O-ring is included.

      Q3: Can membrane cap setting values be shared?
      No. Each membrane cap has its own dedicated setting values. Do not lose the membrane cap information strip or setting data.

      Q4: What parameters must be entered after membrane cap replacement?
      After replacing the membrane cap, enter or confirm the A0, A1, A2, A3, T, and related setting values according to the membrane cap information strip.

      Q5: What is the address for the membrane cap A0 parameter?
      A0 can be read or set at PLC address 107, HEX address 0x006A, Function Code 03, Length 2 Words, with a floating-point data type.

      Q6: What is the address for the membrane cap A1 parameter?
      A1 can be read or set at PLC address 109, HEX address 0x006C, Function Code 03, Length 2 Words, with a floating-point data type.

      Q7: What is the address for the membrane cap A2 parameter?
      A2 can be read or set at PLC address 111, HEX address 0x006E, Function Code 03, Length 2 Words, with a floating-point data type.

      Q8: What is the address for the membrane cap A3 parameter?
      A3 can be read or set at PLC address 113, HEX address 0x0070, Function Code 03, Length 2 Words, with a floating-point data type.

      Q9: What is the address for the membrane cap T parameter?
      T can be read or set at PLC address 125, HEX address 0x007C, Function Code 03, Length 2 Words, with a floating-point data type.

      Q10: Is calibration required after replacing the membrane cap?
      Yes, it is recommended. After replacing the membrane cap, confirm that all parameters are entered correctly and repeat the calibration procedure to ensure accurate measurement results.
  • F
    1. Maintenance Questions
      Q1: Is regular maintenance required?
      Yes. Regular maintenance and cleaning are required to keep the analyzer in optimal operating condition, especially in environments where grease, sediment, or other contaminants may accumulate.

      Q2: How often should it be cleaned?
      Routine maintenance is recommended at least once per week. The actual frequency may be adjusted according to site water quality, contamination level, and monitoring conditions.

      Q3: What does routine maintenance include?
      Routine maintenance includes cleaning the analyzer surface and sensing area, checking battery power, checking consumable status, and confirming that cables and mounting fixtures are secure to prevent looseness or damage that could affect data transmission.

      Q4: How often should periodic maintenance be performed?
      Periodic maintenance is recommended at least once per month and may be adjusted based on long-term monitoring data and site contamination levels.

      Q5: What does periodic maintenance include?
      Periodic maintenance may include trend calibration and standard-solution calibration. It is recommended to analyze long-term data trends and adjust instrument status as needed to ensure stable and accurate measurement results.

      Q6: What should be noted when cleaning the SMR09?
      Use a soft cloth or dedicated neutral cleaner to clean the analyzer surface, especially the sensing area, to prevent grease or contaminants from accumulating and affecting measurement accuracy. When cleaning the electrode or sensing area, rinse with pure water and dry with lens tissue.

      Q7: What consumables are required?
      Main consumables include dissolved oxygen standard solution, pure water, and the replaceable membrane cap. If the device uses specific electrodes or related consumables, they should also be checked and replaced periodically according to maintenance results.

      Q8: When should the membrane cap be replaced?
      When the membrane cap is contaminated, aged, damaged, causes unstable readings or slow response, or normal readings still cannot be obtained after standard-solution calibration, inspect the membrane cap condition and evaluate whether replacement is required.

      Q9: How should the analyzer be stored when not used for an extended period?
      If the analyzer will not be used for an extended period, clean it thoroughly, store it properly according to the manual, and prevent the sensing area from becoming contaminated or damaged.

      Q10: Is recalibration required after replacing consumables?
      Yes, it is recommended. After replacing the membrane cap or related consumables, confirm that all connections and settings are correct and repeat the calibration procedure to ensure measurement accuracy.