SMR08 TDS and Temperature Analyzer

Frequently Asked Questions

  • A
    1. Product Overview
      Q1: What is this product?
      It is the AQUAS Total Dissolved Solids (TDS) and Temperature Analyzer, designed for real-time monitoring of total dissolved solids concentration and temperature in water, with data output via RS485 digital communication.

      Q2: What parameters can it measure?
      It primarily measures Total Dissolved Solids (TDS) and temperature, and includes temperature compensation.

      Q3: What is the measurement principle?
      It uses a two-electrode measurement principle. The resistance of the water sample is measured between two opposing electrodes; conductivity is then calculated from the reciprocal of resistance and the cell constant, and finally converted into total dissolved solids concentration.

      Q4: What applications is it suitable for?
      It is suitable for drinking water, surface water, groundwater, industrial processes, water treatment, wastewater, and related applications.

      Q5: Is a control box required?
      Not necessarily. The unit features an integrated compact housing with a built-in transmitter and sensor. Its rugged IP68 submersible, corrosion-resistant enclosure allows direct installation according to site requirements.

      Q6: Does it support PLC or HMI integration?
      Yes. It uses the standard Modbus RTU protocol and can connect to a PLC, HMI, or AQUAS data logger via RS485.

      Q7: Does it have explosion-proof certification?
      Yes. The SMR08 is certified to IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS.

      Q8: Is it suitable for outdoor sites without mains power?
      Yes. Its ultra-low-power design allows operation with batteries, solar power, or similar power sources for outdoor or remote monitoring sites.

      Q9: Does it include self-diagnostic functions?
      Yes. It includes internal status monitoring such as temperature and humidity, as well as diagnostic functions for abnormal TDS electrode conditions.

      Q10: Does it comply with EPA automatic monitoring methods?
      It provides temperature compensation and measurement functions, and is indicated as complying with Taiwan EPA announced automatic monitoring facility method NIEA W218.51C.
  • B
    1. Installation
      Q1: What is the operating voltage?
      The operating voltage is 5-12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by voltage abnormalities.

      Q2: What installation methods are available?
      Immersion, flow-through, and pipe-insertion installation methods are available.

      Q3: What should be noted during installation?
      Avoid pulling or damaging the cable during installation. Ensure sufficient cable length and secure it properly. Installation in a stable-flow location is recommended for better measurement performance.

      Q4: How deep should the sensing surface be immersed?
      During testing, the sensing surface must be fully immersed at least 10 cm below the water surface.

      Q5: How long does it take to stabilize after installation?
      After installation, allow at least 5 minutes for the analyzer to reach a stable condition. The actual stabilization time may vary depending on sample characteristics and the installation environment.

      Q6: Can it be installed close to the bottom of the tank?
      This is not recommended. The analyzer should not directly touch or be installed too close to the bottom, as obstruction of the sensing area, sludge, or sediment interference may cause abnormal readings.

      Q7: Should it be fixed vertically during field installation?
      Yes. For field installation, the analyzer should be fixed vertically. Use a protective pipe and mounting clamps to secure the device and prevent impact or installation displacement.

      Q8: What should be considered for outdoor or high-load installations?
      For outdoor or high-load applications, AQUAS dedicated protective covers and fixtures are recommended to improve overall protection and measurement stability.

      Q9: How should the SMR08 be set up for laboratory testing?
      Connect the analyzer to a computer, fix it vertically on a stand, insert it into the calibration vessel, remove the electrode protection bottle, and ensure that the electrode sensing area is fully immersed in the sample water. Avoid contact with the bottom or proximity to the vessel wall.

      Q10: Is stirring required during laboratory testing?
      It is recommended to place a 2.5 cm magnetic stir bar at the bottom of the vessel and use a magnetic stirrer to maintain stable rotation, preferably at approximately 300 rpm, to prevent sample stratification and sediment from affecting the results.

      Q11: Can it be installed in a flow-through holder?
      Yes. The manual lists FTC08 series flow-through holders and PIF08 pipe-insertion holders, which can be selected according to flow cell, pipeline, or fixed insertion requirements.
  • C
    1. Communication and Wiring
      Q1: What communication interface is used?
      The SMR08 uses an RS485 communication interface.

      Q2: What communication protocol is used?
      The SMR08 uses the Modbus RTU communication protocol.

      Q3: What are the default communication settings?
      The commonly used communication settings are as follows:
      Communication interface: RS485
      Baud rate: 19200 bps
      Data length: 8 bit
      Parity: None
      Stop bit: 1 bit
      Communication mode: RTU
      The device supports 9600 and 19200 bps.

      Q4: What is the default device address?
      The factory default device address is 01. If the device address is changed, the corresponding address in the PLC, HMI, host, or reading software must also be updated.

      Q5: What is the RS485 wiring configuration?
      The SMR male connector wiring is as follows:
      PIN 1:GND
      PIN 2:RS485-
      PIN 3:RS485+
      PIN 4:5-12V DC
      PIN 5: Not used
      The standard new wire colors are typically: black for GND, white for RS485-, blue for RS485+, and red for 5-12V DC.

      Q6: For versions with AO analog output, what is PIN 5?
      For AO versions, PIN 5 is the AO analog output wire, with a gray wire color. On standard non-AO versions, PIN 5 is not used.

      Q7: Can it be connected directly to a computer?
      Yes. It can be connected to a computer using the CAB06 USB to RS485 cable, allowing values to be read, calibration to be performed, and parameters to be configured through Modbus tools or AQCFG.

      Q8: Can it be connected to an AQUAS data logger?
      Yes. The SMR08 can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data logging and remote monitoring according to site requirements.

      Q9: What should be checked first if the PLC cannot read SMR08 values?
      Check the following items in order:
      1. Whether the device can be connected to a PC using CAB06 and read via Modbus
      2. Whether the PLC and SMR08 communication settings match
      3. Whether the Modbus ID is correct and whether there is any ID conflict
      4. Whether the PLC Modbus address is correct
      5. Whether the power supply is stable at 5-12 VDC
      6. Whether the RS485 wiring is correct
      7. Whether the wiring has been completed according to the appendix wiring diagram

      Q10: What happens if RS485 D+ / D- are reversed?
      If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
  • D
    1. Modbus Readings
      Q1: Which address should be read for the TDS value?
      The TDS main parameter value can be read from PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with a floating point data type.

      Q2: How can the real-time temperature value be read?
      The real-time temperature value can be read from PLC address 2, HEX address 0x0001, Function Code 04, Length 2 Words, with a floating point data type.

      Q3: How can the device model be read?
      It can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with an ASCII string data type.

      Q4: How can the location, area, or sub-area name be read?
      It can be read from PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with a UTF-16BE string data type.

      Q5: How can latitude and longitude be read?
      They can be read from PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with a floating point data type.

      Q6: How can the cleaning brush be activated?
      If the device is equipped with a cleaning brush, it can be controlled by writing to PLC address 5, HEX address 0x0004, Function Code 10. Data value 0 indicates standby, and 1 indicates activation.

      Q7: How should floating point data be decoded?
      SMR08 floating point values follow the IEEE 754 single-precision floating point format and occupy 4 Bytes. If the decoded value is abnormal, check whether the high/low byte or word order needs to be swapped.

      Q8: What should be noted about the Modbus CRC?
      Modbus RTU uses CRC16. During transmission, Low Byte First and High Byte Last are used, meaning the low byte is transmitted first and the high byte last.

      Q9: What do Function Codes 03, 04, and 10 represent?
      Function Code 03: Read Holding Registers
      Function Code 04: Read Input Registers
      Function Code 10: Write Multiple Holding Registers

  • E
    1. Calibration
      Q1: Is calibration required before first use?
      In general, no. The SMR08 is calibrated before shipment. For first use, measurements can begin after proper installation according to the manual. If accuracy needs to be verified before use, standard solution calibration can be performed.

      Q2: When is recalibration required?
      Recalibration is recommended when field readings differ significantly from laboratory measurements, when readings drift after long-term use, when accuracy must be verified after maintenance or cleaning, after electrode replacement, or when field sample conditions differ substantially from standard solution conditions.

      Q3: What calibration items are available?
      SMR08 calibration items include temperature calibration, standard solution zero calibration, standard solution span/slope calibration, field calibration, and AO calibration for optional AO versions.

      Q4: What should be done before calibration?
      The analyzer should be thoroughly cleaned before calibration. It is recommended to remove the electrode protective cap, rinse the surface with tap water, clean with a soft brush and neutral detergent if necessary, wipe the body as dry as possible, and shake excess water from the electrode.

      Q5: What tools are required for calibration?
      Prepare the following:
      Calibrated reference thermometer
      WTW conductivity meter
      Calibration cup
      Magnetic stir bar and magnetic stirrer
      Ultrapure water

      Q6: What standard solutions are required for calibration?
      The following may be used:
      SL08-1 Total Dissolved Solids Standard Solution, 50 mg/L, 500 ml
      SL08-2 Total Dissolved Solids Standard Solution, 300 mg/L, 500 ml
      SL08-3 Total Dissolved Solids Standard Solution, 1,000 mg/L, 500 ml

      Q7: How is temperature calibration performed?
      Pour purified water or tap water into the calibration cup, then place the analyzer and reference thermometer into the cup. Wait at least 2 minutes and confirm that the temperature is stable. Write the result of “real-time temperature value - thermometer reading” as the temperature correction value.

      Q88: Which address should the temperature correction value be written to?
      The temperature correction value can be written to PLC address 714, HEX address 0x02C9, Function Code 10, Length 2 Words, with a floating point data type.

      Q9: What type of water should be used for zero calibration?
      Zero calibration must be performed using compliant purified water, such as Type II purified water with conductivity below 1.0 µS/cm, or a higher-grade purified water.

      Q10: How is zero calibration performed?
      Place purified water and a magnetic stir bar into the calibration vessel. Immerse the analyzer to the proper depth and start the magnetic stirrer at a constant speed, avoiding large vortices and dense bubbles. Then read the zero calibration value, enter 0 as the zero calibration value, enter 1 to execute zero calibration, and finally check the real-time TDS value to confirm completion.

      Q11: Which address should the zero calibration value be written to?
      The zero calibration value can be written to PLC address 519, HEX address 0x0206, Function Code 10, with a floating point data type.

      Q12: Which address should be written to execute zero calibration?
      Write 1 to PLC address 518, HEX address 0x0205, Function Code 10, to execute zero calibration.

      Q13: How is slope calibration performed?
      Place the TDS standard solution and a magnetic stir bar into the calibration vessel. Immerse the analyzer to the proper depth and start the magnetic stirrer at a constant speed. After the real-time TDS value stabilizes, enter the standard solution concentration as the slope calibration value, then enter 1 to execute slope calibration.

      Q14: Which address should the slope calibration value be written to?
      The slope calibration value can be written to PLC address 521, HEX address 0x0208, Function Code 10, with a floating point data type.

      Q15: Which address should be written to execute slope calibration?
      Write 1 to PLC address 520, HEX address 0x0207, Function Code 10, to execute slope calibration.

      Q16: What concentration is recommended for slope calibration?
      Select the standard solution concentration according to the target measurement range. The standard solutions listed in the manual include 50 mg/L, 300 mg/L, and 1,000 mg/L.

      Q17: What are Zero lock and Span lock?
      Zero lock and Span lock are locked values in the calibration logic, used as the zero reference and slope/span reference. Zero lock can be read from PLC address 733, HEX address 0x02DC; Span lock can be read from PLC address 734, HEX address 0x02DD.

      Q18: What should be noted before field calibration?
      After the analyzer is installed in the field, wait until the reading stabilizes. The target calibration value should be based on laboratory test results. After successful calibration, the slope value will be updated.

      Q19: What parameter is updated by field calibration?
      After successful field calibration, the slope value is updated to correct deviations caused by differences between field sample conditions and standard solution conditions.

      Q20: Why should large vortices and dense bubbles be avoided during calibration?
      Large vortices and dense bubbles may interfere with electrode readings, causing unstable TDS values and affecting calibration results.
  • F
    1. AO Analog Output
      Q1: Does it support AO 4-20 mA output?
      If the AO option is selected, the AO output range can be configured so that 4-20 mA corresponds to the specified TDS measurement range.

      Q2: What equipment is required for AO calibration?
      AO calibration requires the SMR instrument, suitable power wiring, a USB-to-RS485 adapter (CAB06 recommended), and a digital multimeter or current calibrator.

      Q3: Are reagents required for AO calibration?
      No. AO calibration is an analog output current calibration and does not require reagents.

      Q4: What is the AO1 calibration wiring configuration?
      AO1 wiring is as follows:
      Brown: AO1+, connect to multimeter mA+
      White: AO1-, connect to multimeter COM / mA-

      Q5: What is the AO2 calibration wiring configuration?
      AO2 wiring is as follows:
      Blue: AO2+, connect to multimeter mA+
      Black: AO2-, connect to multimeter COM / mA-

      Q6: What mode should the multimeter be set to during AO calibration?
      Set the multimeter to DC mA mode and ensure that the probes are inserted into the correct mA measurement terminals.

      Q7: What does a negative AO current reading on the multimeter indicate?
      A negative reading indicates that the polarity may be reversed. Swap the multimeter leads and measure again.

      Q8: Which addresses should AO Min / Max be written to?
      The AO output range can be written to:
      Min: PLC address 31241, HEX address 0x7A08
      Max: PLC address 31242, HEX address 0x7A09

      Q9: Which address should the AO current offset be written to?
      The AO current offset can be written to PLC address 31236, HEX address 0x7A03, Function Code 10, with a floating point data type.

      Q10: How can normal AO output be verified during calibration?
      First set the Min and Max ranges, then read the real-time value of the measured parameter and calculate the theoretical current within the 4-20 mA range. If there is an error between the multimeter reading and the theoretical current, write a current offset correction. Finally, confirm again that the multimeter current has been calibrated to the theoretical current.
  • F
    1. Abnormal Readings and Troubleshooting
      Q1: What should be done if the reading keeps fluctuating?
      First clean the sensing area, confirm that the installation depth is sufficient, the device is securely fixed, and there are no bubbles, sludge, or contaminants interfering at the site. Also check that the power supply and RS485 communication are normal.

      Q2: What should be done if the reading remains at 0?
      Confirm whether the sample actually contains dissolved solids or has conductivity. Then check whether the electrode is clean, sufficiently immersed, and whether zero and slope calibration have been performed correctly. If there is still no response when testing with a standard solution, contact AQUAS.

      Q3: What should be done if the reading remains at the upper range limit?
      Possible causes include sample TDS exceeding the measurement range, calibration errors, electrode contamination, bubble interference, or obstruction of the sensing area. Clean the analyzer first and reconfirm the calibration.

      Q4: What should be done if the reading is obviously too high or too low?
      Check whether zero and slope calibration are correct, whether the electrode is dirty or damaged, whether the sample is homogeneous, and whether bubbles or sludge are causing interference. Recalibrate if necessary.

      Q5: What should be done if the response time becomes slow?
      Possible causes include electrode contamination, deposits on the sensing area, unstable sample conditions, or degraded electrode performance. Clean the electrode first and verify the standard solution reading. Activate or replace the electrode if necessary.

      Q6: What should be done if readings from the same sample are inconsistent?
      Confirm whether the sample is homogeneous, stirring is stable, stratification or sediment is present, bubbles are attached, and the calibration procedure is correct.

      Q7: What should be done if the reading does not change?
      Restore factory settings first and repeat the calibration procedure. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.

      Q8: Can bubbles affect readings?
      Yes. Bubbles may interfere with sensor readings, causing unstable TDS values or calibration errors. Stabilize the water flow, install the device properly, and avoid dense bubbles adhering to the sensing area.

      Q9: Can sludge or sediment affect SMR08 readings?
      Yes. If the analyzer is buried in sludge or the sensing area is covered by sediment, readings may become abnormal, unstable, or slow to respond.

      Q10: Can the SMR08 electrode be activated?
      Yes. The manual recommends the following EC electrode activation procedure: first clean the electrode surface with purified water, then soak it in 0.15M NaOH for at least 30 minutes. Rinse with clean water, then soak it in 0.15M HCl for at least 30 minutes. The sequence must be alkali first, then acid, and must not be changed. After activation, calibrate according to the standard procedure and observe whether the values return to normal.

      Q11: What should be done if readings are still incorrect after electrode activation?
      If readings remain incorrect after activation and calibration, electrode replacement is recommended. If a new electrode is still abnormal, contact AQUAS.

      Q12: If field readings differ from laboratory measurements, does this indicate device failure?
      Not necessarily. Real-time monitoring instruments are directly exposed to field conditions, while laboratory samples typically undergo sampling, storage, pretreatment, and standardized procedures. Differences in test conditions and timing may cause discrepancies. Field calibration can improve consistency.

      Q13: When should AQUAS technical support be contacted?
      If the device still cannot be read or does not respond after checking power supply, wiring, Modbus ID, communication settings, standard solution testing, calibration, cleaning, electrode activation, and installation, contact AQUAS technical support. Provide site photos, wiring photos, Modbus settings, standard solution test records, and reading screenshots.
  • F
    1. Maintenance and Consumables
      Q1: Is periodic maintenance required?
      Yes. To keep the analyzer in optimal operating condition, periodic cleaning and calibration are required, especially in environments where grease, sediment, or other contaminants can easily adhere to the device.

      Q2: How often should cleaning be performed?
      Cleaning is recommended approximately once per week. The cleaning frequency may be adjusted according to site water quality conditions.

      Q3: How often should calibration solution be used for calibration?
      Calibration using calibration solution is recommended approximately once every two weeks. The frequency may be adjusted according to site water quality conditions.

      Q4: How should the analyzer be stored when not used for an extended period?
      If the analyzer will not be used for an extended period, clean it thoroughly and store it dry.

      Q5: What maintenance cleaning procedure is recommended?
      It is recommended to clean the entire unit with purified water and a soft brush, then rinse with purified water and confirm that the electrode and body are free of dirt. If obvious contamination is present, soak the unit in 5% dilute sulfuric acid for 15 minutes, then place it in purified water for zero-point verification.

      Q6: What reading condition should be met during zero-point verification?
      When performing zero-point verification in purified water, the measured value should be less than 5% of full scale. If the reading remains greater than 15% of full scale after acid soaking, return for service and maintenance is recommended.

      Q7: What standard solution test error is considered acceptable?
      When testing with a calibration solution, a measured value error of less than 5% may be considered acceptable after maintenance. If the error is between 5% and 20%, single-point slope calibration may be performed. If the error exceeds 20%, check whether bubbles are present in the sensing area and whether the standard solution has deteriorated or become contaminated.

      Q8: What consumables are required?
      The main consumables include TDS standard solutions and purified water.

      Q9: When should the electrode be replaced?
      Electrode replacement is recommended when the electrode response becomes slow, readings become unstable, electrode performance noticeably declines, or physical damage such as cracks, wear, or breakage is observed.

      Q10: Can an electrode still be used after more than one year of service?
      If the electrode has been used for more than one year, it may have aged. Even if it can still provide readings, replacement should be evaluated based on standard solution calibration results and reading stability.