SMR12 Total Suspended Solids (TSS) and Mixed Liquor Suspended (MLSS)Analyzer
Frequently Asked Questions
-
Q1
Product and Basic Use
A- Product and Basic Use
Q1: What is the product?
It is a total suspended solids TSS analyzer designed for real-time monitoring of total suspended solids concentration in water, with digital data output via RS485 communication.
Q2: What can it measure?
It is mainly used to measure total suspended solids concentration in water and is suitable for applications that require real-time monitoring of sludge, suspended particles, or changes in water turbidity.
Q5: What is the measurement principle?
It uses a built-in LED light source to emit 880 nm near-infrared light. After the light beam is scattered by suspended particles in the water sample, the detector measures the signal intensity and converts it into total suspended solids concentration. Since 880 nm near-infrared light is used, measurement is less affected by sample color.
Q6: What application sites is it suitable for?
It is suitable for drinking water, surface water, groundwater, industrial processes, water treatment, wastewater, and other applications.
Q7: Is a control box required?
Not necessarily. SMR12 has an integrated compact enclosure with a built-in transmitter and sensor. It features a rugged IP68 submersible, corrosion-resistant housing and can be installed directly according to site requirements.
Q8: Does it support PLC or HMI?
Yes. SMR12 uses the standard Modbus RTU communication protocol and can connect to PLCs, HMIs, or AQUAS data loggers via RS485.
Q9: Does it have explosion-proof certification?
Yes. SMR12 is certified with IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS explosion-proof certification.
Q9: Does it have an automatic cleaning function?
Yes. SMR12 has an automatic cleaning brush, and the cleaning interval can be configured to reduce measurement interference caused by microorganisms, sludge, and air bubble attachment.
Q10: What are the features of the optical window material?
The optical window uses abrasion-resistant and scratch-resistant sapphire glass with a nano coating, reducing sludge deposition and biological attachment while improving cleaning and maintenance convenience.
Q11: Is it suitable for outdoor environments without mains power?
Yes. SMR12 features an ultra-low power consumption design and can be used with battery or solar power supply for outdoor or remote monitoring sites.
- Product and Basic Use
-
Q2
Installation Questions
B- Installation Questions
Q1: What is the operating voltage?
The operating voltage is 5–12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.
Q2: What installation methods are available?
Immersion, flow-through, and pipe-insertion installation methods are available.
Q3: What should be noted during installation?
During installation, avoid pulling or damaging the connection cable. The cable should be long enough and properly secured. It is recommended to install the analyzer at a location with stable water flow to obtain better measurement results.
Q4: How deep should the sensing surface be submerged?
During testing, the sensing surface must be fully submerged at least 10 cm below the water surface.
Q5: How long does it take to stabilize after installation?
After installation, the analyzer should be allowed to stabilize for at least 5 minutes. The actual stabilization time may vary depending on sample characteristics and installation conditions.
Q6: Can it be installed close to the bottom of the tank?
No. The analyzer should not directly contact or be installed too close to the bottom of the tank, as this may cause abnormal measurement due to sensing area obstruction or sediment interference.
Q7: How far should the sensing optical window be from the bottom of the tank or channel?
It is recommended that the sensing optical window be at least 30 cm away from the bottom of the tank or channel to prevent sediments from entering the optical field of view and causing errors.
Q8: How much clearance should be kept around the analyzer?
It is recommended to keep a 30 cm clearance around the analyzer to prevent walls, turbulent flow, bottom sludge, or other structures from affecting the measurement results.
Q9: Why is it recommended to install isolation plates?
It is recommended to install two layers of isolation plates to reduce the impact of air bubbles on value stability.
Q10: Is there a recommended range for on-site flow conditions?
It is recommended to maintain sample flow velocity between 0.1 and 10 m/s. The analyzer can be used in open water bodies, open channels, and pressurized pipeline systems.
Q11: What is the on-site pressure limit?
The recommended operating pressure should not exceed 10 kgf/cm² to prevent sensor damage caused by overpressure.
Q12: What should be noted when installing outdoors or in high-load environments?
When installed outdoors or in high-load environments, it is recommended to use AQUAS dedicated protective covers, protection tubes, and mounting clamps to improve protection and measurement stability.
Q13: How should SMR12 be set up for laboratory testing?
The analyzer should be connected to a computer, vertically fixed on a stand, and inserted into the calibration container. It is recommended to use the AQUAS calibration cup, with a diameter of 85 mm and a height of 220 mm.
Q14: How far should the sensing optical window be from the bottom of the container during laboratory testing?
The sensing optical window should be fully immersed in the water sample and at least 200 mm away from the bottom of the container to avoid optical reflection interference.
Q15: Is stirring required during laboratory testing?
It is recommended to place a 2.5 cm magnetic stir bar and use a magnetic stirrer to maintain stable rotation at approximately 300 rpm to ensure sample homogeneity.
Q16: What optical interference should be avoided during laboratory testing?
It is recommended to use an opaque container or avoid direct light exposure to reduce interference from external light on optical measurement.
- Installation Questions
-
Q3
Communication and Wiring Questions
C- Communication and Wiring Questions
Q1: Which communication interface is used?
RS485 communication interface is used.
Q2: Which communication protocol is used?
The standard Modbus RTU communication protocol is used.
Q3: What are the default communication settings?
Common communication settings are as follows:
Communication interface: RS485
Baud rate: 19200 bps
Data length: 8 bit
Parity: None
Stop bit: 1 bit
Communication mode: RTU
The device supports 9600 and 19200 bps.
Q4: Can it be connected directly to a computer?
Yes. It can be connected to a computer using CAB06 USB to RS485 Cable, and values can be read, calibrated, and parameters configured through a Modbus tool or AQCFG.
Q5: If AO analog output is available, what is PIN 5?
For the AO version, PIN 5 is the AO analog output wire, and the wire color is gray. For general non-AO versions, PIN 5 is not used.
Q6: What is the default device address?
The factory default device address is 01. If the device address is changed, the device address in the PLC, HMI, host, or reading software must also be updated accordingly.
Q7: How should RS485 wiring be connected?
The SMR male connector wiring is as follows:
PIN 1: GND
PIN 2: RS485-
PIN 3: RS485+
PIN 4: 5–12 VDC
PIN 5: Not used
New wire colors are typically: black for GND, white for RS485-, blue for RS485+, and red for 5–12V DC.
Q7: Can it be connected to an AQUAS data logger?
Yes. SMR12 can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data logging and remote monitoring according to site requirements.
Q8: What should be checked first if the PLC cannot read values?
Please check the following in order:
1. Whether CAB06 can be used to connect to a PC and read data via Modbus
2. Whether the communication settings of the PLC and SMR02 are consistent
3. Whether the Modbus ID is correct and whether there is an ID conflict
4. Whether the PLC Modbus address is correct
5. Whether the power supply is stable at 5–12 VDC
6. Whether RS485 wiring is correct
7. Whether the connection has been completed according to the appendix wiring diagram
Q9: What happens if RS485 D+ / D- are reversed?
If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
- Communication and Wiring Questions
-
Q4
Modbus Reading Questions
D- Modbus Reading Questions
Q1: Which address should be used to read the total suspended solids value?
The main parameter value can be read using PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with the data type set to floating point.
Q2: Which address should be used to read the secondary parameter value?
The secondary parameter value can be read from PLC address 2, HEX address 0x0001, Length 2 Words, Function Code 04, with the data type set to floating point.
Q3: How can the location, area, or zone name be read?
The location, area, or zone name can be read from PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with the data type set to UTF-16BE string.
Q4: How can the device model be read?
The device model can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with the data type set to ASCII string
Q5: How can latitude and longitude be read?
Latitude and longitude can be read from PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type set to floating point.
Q6: How can the cleaning brush be activated?
If the device is equipped with a cleaning brush function, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.
Q7: How should floating-point data be decoded?
SMR12 floating-point data follows the IEEE 754 single-precision floating-point format and occupies 4 Bytes. If the decoded value is abnormal, confirm whether the high/low bytes or word order need to be swapped.
Q8: What should be noted about Modbus CRC?
Modbus RTU uses CRC16. During transmission, Low Byte is sent first and High Byte last, meaning the low-order byte comes before the high-order byte.
Q9: What do Function Codes 03, 04, and 10 represent?
Function Code 03: Read Holding Registers
Function Code 04: Read Input Registers
Function Code 10: Write Multiple Holding Registers
- Modbus Reading Questions
-
Q5
Calibration Questions
E- Calibration Questions
Q1: Is calibration required before initial use?
In general, no. SMR12 has been calibrated before shipment. During initial use, measurement can begin once the device is properly installed according to the manual.
Q2: When is recalibration required?
Recalibration is recommended when on-site readings differ significantly from laboratory measurements, when readings drift after long-term use, when accuracy needs to be confirmed after maintenance or cleaning, or when on-site sample conditions differ significantly from the standard solution.
Q3: What calibration items are available?
SMR12 calibration includes standard calibration, on-site calibration, and AO calibration. Standard calibration is divided into zero-point calibration and slope calibration.
Q4: What should be done before calibration?
The analyzer must be thoroughly cleaned before calibration. It is recommended to rinse the surface with tap water, clean with a soft brush and neutral detergent if necessary, rinse again with purified water, and finally dry all parts of the body as much as possible.
Q5: What tools are required for calibration?
The following tools are required:
Calibration container, preferably the AQUAS dedicated calibration cup
Wash bottle
Lens cleaning paper
Magnetic stirrer
2.5 cm magnetic stir bar
Stand and mounting clamp
Q6: What standard solutions are required for calibration?
The following can be used:
SL12-1 total suspended solids standard solution, 500 mg/L, 1000 ml
SL12-2 total suspended solids standard solution, 1000 mg/L, 1000 ml
SL12-3 total suspended solids standard solution, 5000 mg/L, 1000 ml
Ultrapure water
Q7: What water should be used for zero-point calibration?
Zero-point calibration must use purified water that meets the required specification, such as Type II grade purified water with conductivity below 1.0 µS/cm, or higher-grade purified water.
Q8: How is zero-point calibration performed?
Place purified water and a magnetic stir bar into the calibration container, immerse the analyzer to the appropriate depth, and start the magnetic stirrer at a constant speed while avoiding large vortices and dense bubbles. Then enter 0 into the zero-point calibration value, enter 1 at the zero-point calibration button to perform zero-point calibration, and finally query the real-time total suspended solids value to confirm whether calibration has been completed.
Q9: Which address should the zero-point calibration value be written to?
The zero-point calibration value can be written to PLC address 12807, HEX address 0x3206, Function Code 10, with the data type set to floating point.
Q10: Which address should be written to perform zero-point calibration?
Input 1 at PLC address 12806, HEX address 0x3205, Function Code 10, to perform zero-point calibration.
Q11: What standard solution should be used for slope calibration?
Slope calibration uses a total suspended solids standard solution. It is recommended to select the standard solution concentration according to the target measurement value. The standard solutions listed in the manual include 500 mg/L, 1000 mg/L, and 5000 mg/L.
Q12: How is slope calibration performed?
Place the standard solution and magnetic stir bar into the calibration container, immerse the analyzer to the appropriate depth, start the magnetic stirrer at a constant speed, and wait until the real-time total suspended solids value stabilizes. Then enter the standard solution concentration into the slope calibration value, and enter 1 at the slope calibration button to perform slope calibration.
Q13: Which address should the slope calibration value be written to?
The slope calibration value can be written to PLC address 12809, HEX address 0x3208, Function Code 10, with the data type set to floating point.
Q14: Which address should be written to perform slope calibration?
Input 1 at PLC address 12808, HEX address 0x3207, Function Code 10, to perform slope calibration.
Q15: Can the cleaning brush be activated during calibration?
Yes. The cleaning brush can be activated to remove bubbles attached to the optical window and prevent abnormal readings.
Q16: Why should large vortices and dense bubbles be avoided during calibration?
Large vortices and dense bubbles may interfere with optical window readings, causing unstable total suspended solids readings or calibration errors.
Q17: What are Zero lock and Span lock?
Zero lock and Span lock are locked values in the calibration logic, used for the zero-point reference and slope reference. Zero lock can be read from PLC address 13022, HEX address 0x32DD; Span lock can be read from PLC address 13021, HEX address 0x32DC.
Q18: What should be noted before on-site calibration?
After the analyzer is installed on site, wait until the reading stabilizes. The target calibration value should preferably refer to laboratory test results. After successful calibration, the slope value will be updated.
Q19: What is single-point slope calibration?
Single-point slope calibration is a quick adjustment based on a known concentration at a single time point. It can quickly correct instrument deviation and is suitable for short-term rapid calibration.
Q20: What are trend monitoring and regression calibration?
Trend monitoring and regression calibration use long-term data collection and analysis for correction. They can continuously optimize instrument performance, compensate for instantaneous errors and medium- to long-term drift, and are suitable for long-term monitoring sites.
Q21: Will the standard solution reading still be correct after on-site calibration?
Not necessarily. The manual notes that after on-site slope calibration is performed, the original standard reading may no longer be correct and should be reconfirmed or recalibrated with standard solution.
- Calibration Questions
-
Q7
Abnormal Readings and Troubleshooting
F- Abnormal Readings and Troubleshooting
Q1: What should I do if readings keep fluctuating?
First clean the sensing area, check whether the water inlet is blocked, whether air bubbles remain, whether the installation is secure, whether the on-site water flow is excessively disturbed, and whether the signal cable is affected by interference from power cables.
Q1: What should I do if the reading remains at 0?
Confirm whether the sample and standard solution are normal, then check whether the electrode is clean, whether it is immersed deeply enough, and whether calibration is correct. If there is still no response when testing with a standard solution, contact AQUAS.
Q3: What should I do if the reading remains at the maximum range?
Possible causes include incorrect calibration, electrode contamination, electrode aging, air bubble interference, or sensing area obstruction. It is recommended to clean the analyzer first and reconfirm standard calibration.
Q4: What should I do if readings are clearly too high or too low?
Check whether standard calibration is correct, whether the electrode is dirty or damaged, whether the sample is homogeneous, and whether bubbles or bottom sludge are causing interference. Recalibrate if necessary.
Q5: What should I do if the response time becomes slower?
This may be caused by electrode aging, contamination in the sensing area, unstable sample conditions, or degraded electrode performance. It is recommended to clean the electrode first and confirm the standard solution reading. Activate or replace the electrode if necessary.
Q6: What should I do if readings from the same sample are inconsistent?
Confirm whether the sample is homogeneous, whether stirring is stable, whether stratification or sediment is present, whether bubbles are attached, and whether the calibration procedure is correct.
Q7: What should I do if the reading does not change?
Restore factory settings first and repeat the calibration procedure. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.
Q8: If on-site readings differ from laboratory measurements, does that mean the device is faulty?
Not necessarily. Real-time monitoring instruments are directly exposed to the on-site environment, while laboratory samples usually undergo sampling, storage, pretreatment, and standardized procedures. Different test conditions and time points may lead to differences in values.
Q9: Do air bubbles affect SMR07 readings?
Yes. Air bubbles may interfere with sensor readings, causing unstable conductivity values or calibration errors. Stable water flow, correct installation, and avoiding dense bubbles attached to the sensing area are recommended.
Q10: Do bottom sludge or sediments affect SMR07 readings?
Yes. If the analyzer is buried by bottom sludge or the sensing area is blocked by sediment, readings may become abnormal, unstable, or slower to respond.
Q11: Can the SMR07 electrode be activated?
Yes. The manual recommends the following EC electrode activation procedure: first rinse the electrode surface with purified water, then soak it in 0.15M NaOH for at least 30 minutes, rinse with clean water, and then soak it in 0.15M HCl for at least 30 minutes. The order must be alkaline first, then acidic, and must not be changed. After activation, perform calibration according to the standard procedure and observe whether the values are normal.
Q12: What should I do if the reading is still incorrect after electrode activation?
If the reading is still incorrect after activation and calibration, electrode replacement is recommended. If the new electrode is still abnormal, contact AQUAS.
Q13: When should AQUAS technical support be contacted?
If the device still cannot read values or shows no response after checking power supply, wiring, Modbus ID, communication settings, standard solution testing, calibration, cleaning, electrode activation, and electrode inspection, it is recommended to contact AQUAS technical support and provide on-site photos, wiring photos, Modbus settings, standard solution test records, and reading screenshots.
- Abnormal Readings and Troubleshooting
-
Q8
AO Analog Output Questions
F- AO Analog Output Questions
Q1: Does SMR12 support AO 4–20 mA output?
If the AO version is selected, the AO output range can be configured so that 4–20 mA corresponds to the specified measurement range.
Q2: What equipment is required for AO calibration?
AO calibration requires an SMR instrument, suitable power cables, a USB to RS485 converter, CAB06 recommended, and a digital multimeter or current calibrator.
Q3: Are reagents required for AO calibration?
No. AO calibration is analog output current calibration and does not require reagents.
Q4: What is the AO1 calibration wiring?
AO1 wiring is as follows:
Brown: AO1+, connect to multimeter mA+
White: AO1−, connect to multimeter COM / mA−
Q5: What is the AO2 calibration wiring?
AO2 wiring is as follows:
Blue: AO2+, connect to multimeter mA+
Black: AO2−, connect to multimeter COM / mA−
Q6: What mode should the multimeter be set to during AO calibration?
The multimeter should be set to DC mA mode, and the test leads must be inserted into the correct mA measurement terminals.
Q7: What does a negative value on the multimeter indicate during AO measurement?
If the multimeter displays a negative value, the polarity may be reversed. Swap the multimeter wiring and measure again.
Q8: Which addresses should AO Min / Max be written to?
The AO output range can be written to:
Min: PLC address 31241, HEX address 0x7A08
Max: PLC address 31242, HEX address 0x7A09
Q9: Which address should the AO current offset be written to?
The AO current offset can be written to PLC address 31236, HEX address 0x7A03, Function Code 10, with the data type set to floating point.
- AO Analog Output Questions
-
Q9
Maintenance Questions
F- Maintenance Questions
Q1: What should I do if readings keep fluctuating?
First clean the sensing area, confirm whether the installation depth is sufficient, whether the device is firmly secured, whether there are air bubbles, bottom sludge, or structural interference around the device, and whether the power supply and RS485 communication are normal.
Q2: What should I do if the reading remains at 0?
Confirm whether the sample actually contains suspended solids, then check whether the optical window is clean, whether it is immersed deeply enough, and whether calibration is correct. If there is still no response when testing with a standard solution, contact AQUAS.
Q3: What should I do if the reading remains at the maximum range?
Possible causes include sample suspended solids concentration exceeding the measurement range, optical window contamination, incorrect calibration, air bubble interference, or sensing area obstruction. It is recommended to clean the analyzer first and reconfirm calibration.
Q4: What should I do if readings are clearly too high or too low?
Check whether zero-point calibration and slope calibration are correct, whether the optical window is dirty, whether the sample is homogeneous, and whether bubbles or bottom sludge are causing interference. Recalibrate if necessary.
Q5: What should I do if the response time becomes slower?
This may be caused by optical window contamination, unstable sample conditions, or deposits affecting the sensing area. It is recommended to clean the optical window first and confirm the standard solution reading.
Q6: What should I do if readings from the same sample are inconsistent?
Confirm whether the sample is homogeneous, whether stirring is stable, whether stratification or sediment is present, whether bubbles are attached, and whether the calibration procedure is correct.
Q7: What should I do if the reading does not change?
Restore factory settings first and repeat the calibration procedure. You can also try covering the sensing area by hand to confirm whether the value changes. If there is no change, the light source may be damaged or the optical window may be heavily contaminated. Please contact AQUAS.
Q8: If on-site readings differ from laboratory measurements, does that mean the device is faulty?
Not necessarily. Real-time monitoring instruments are directly exposed to the on-site environment, while laboratory samples usually undergo sampling, storage, pretreatment, and standardized procedures. Different test conditions and time points may lead to differences in values. Consistency can be improved through on-site calibration or trend monitoring.
Q9: Do air bubbles affect SMR12 readings?
Yes. Air bubbles may interfere with optical window readings, causing unstable total suspended solids readings. Isolation plates, stable water flow, correct installation position, or activation of the cleaning brush can help reduce the effect.
Q10: Do bottom sludge or sediments affect SMR12 readings?
Yes. If the sensing area is buried by bottom sludge or sediments enter the optical field of view, readings may become too high, too low, or unstable.
Q11: When should AQUAS technical support be contacted?
If the device still cannot read values or shows no response after checking power supply, wiring, Modbus ID, communication settings, standard solution testing, calibration, cleaning, and installation, it is recommended to contact AQUAS technical support and provide on-site photos, wiring photos, Modbus settings, standard solution test records, and reading screenshots.
- Maintenance Questions