SMR07 Conductivity and Temperature Analyzer
Frequently Asked Questions
-
Q1
Product and Basic Use
A- Product and Basic Use
Q1: What is the product?
It is a conductivity and temperature analyzer designed for real-time conductivity and temperature monitoring in water, with digital data output via RS485 communication.
Q2: What models are available?
SMR07 mainly includes:
SMR07-1: conductivity and temperature analyzer
SMR07-3: inductive conductivity and temperature analyzer
Q3: What is the difference between SMR07-1 and SMR07-3?
SMR07-1 is a contact-type conductivity analyzer with selectable replaceable electrodes, such as 2-electrode and 4-electrode types.
SMR07-3 is an inductive conductivity analyzer suitable for high-concentration, acidic or alkaline, or more easily contaminated water sample environments.
Q4: What parameters can it measure?
It mainly measures conductivity and temperature, and includes temperature compensation.
Q5: What is the measurement principle?
SMR07-1 measures the resistance of the water sample through graphite electrodes, then calculates conductivity using the reciprocal of resistance and the cell constant.
SMR07-3 uses the inductive conductivity measurement principle. Current is induced in a closed loop within the solution, and conductivity is calculated according to the current magnitude.
Q6: What application sites is it suitable for?
It is suitable for drinking water, surface water, groundwater, industrial processes, water treatment, wastewater, and other applications.
Q7: Is a control box required?
Not necessarily. SMR07 has an integrated compact enclosure with a built-in transmitter and sensor. It features a rugged IP68 submersible, corrosion-resistant housing and can be installed directly according to site requirements.
Q8: Does it support PLC or HMI?
Yes. SMR07 uses the standard Modbus RTU communication protocol and can connect to PLCs, HMIs, or AQUAS data loggers via RS485.
Q9: Does it have explosion-proof certification?
Yes. SMR07 is certified with IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS explosion-proof certification.
Q10: Is it suitable for outdoor environments without mains power?
Yes. SMR07 features an ultra-low power consumption design and can be used with battery or solar power supply for outdoor or remote monitoring sites.
Q11: Is SMR07-3 suitable for highly contaminated or high-concentration water samples?
Yes. SMR07-3 uses an inductive conductivity design. The electrode is made of acid- and alkali-resistant, corrosion-resistant PEEK material, and its performance is less affected by polarization, oil contamination, surface deposits, and fouling.
- Product and Basic Use
-
Q2
Installation Questions
B- Installation Questions
Q1: What is the operating voltage?
The operating voltage is 5–12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.
Q2: What installation methods are available?
Immersion, flow-through, and pipe-insertion installation methods are available.
Q3: What should be noted during installation?
During installation, avoid pulling or damaging the connection cable. The cable should be long enough and properly secured. It is recommended to install the analyzer at a location with stable water flow to obtain better measurement results.
Q4: How deep should the sensing surface be submerged?
During testing, the sensing surface must be fully submerged at least 10 cm below the water surface.
Q5: How long does it take to stabilize after installation?
After installation, the analyzer should be allowed to stabilize for at least 5 minutes. The actual stabilization time may vary depending on sample characteristics and installation conditions.
Q6: Can it be installed close to the bottom of the tank?
No. The analyzer should not directly contact or be installed too close to the bottom of the tank, as this may cause abnormal measurement due to sensing area obstruction or sediment interference.
Q7: Does it need to be vertically fixed during on-site installation?
Yes. During on-site installation, it is recommended to vertically fix the analyzer and use a protection tube and mounting clamps to secure the device, preventing impact and installation displacement.
Q8: What should be noted when installing outdoors or in high-load environments?
When installed outdoors or in high-load environments, it is recommended to use AQUAS dedicated protective covers and fixtures to improve overall protection and measurement stability.
Q9: How should SMR07 be set up for laboratory testing?
The analyzer should be connected to a computer, vertically fixed on a stand, inserted into the calibration container, and the electrode protection bottle should be removed. The electrode sensing area must be fully immersed in the sample water while avoiding contact with the bottom or proximity to the container wall.
Q10: Is stirring required during laboratory testing?
It is recommended to place a 2.5 cm magnetic stir bar at the bottom of the container and use a magnetic stirrer to maintain stable rotation at approximately 300 rpm, preventing sample stratification and sediment from affecting the results.
- Installation Questions
-
Q3
Communication and Wiring Questions
C- Communication and Wiring Questions
Q1: Which communication interface is used?
RS485 communication interface is used.
Q2: Which communication protocol is used?
The standard Modbus RTU communication protocol is used.
Q3: What are the default communication settings?
Common communication settings are as follows:
Communication interface: RS485
Baud rate: 19200 bps
Data length: 8 bit
Parity: None
Stop bit: 1 bit
Communication mode: RTU
The device supports 9600 and 19200 bps.
Q4: Can it be connected directly to a computer?
Yes. It can be connected to a computer using CAB06 USB to RS485 Cable, and values can be read, calibrated, and parameters configured through a Modbus tool or AQCFG.
Q5: What is the default device address?
The factory default device address is 01. If the device address is changed, the device address in the PLC, HMI, host, or reading software must also be updated accordingly.
Q6: How should RS485 wiring be connected?
The SMR male connector wiring is as follows:
PIN 1: GND
PIN 2: RS485-
PIN 3: RS485+
PIN 4: 5–12 VDC
PIN 5: Not used
New wire colors are typically: black for GND, white for RS485-, blue for RS485+, and red for 5–12V DC.
Q7: Can it be connected to an AQUAS data logger?
Yes. SMR07 can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data logging and remote monitoring according to site requirements.
Q8: What should be checked first if the PLC cannot read values?
Please check the following in order:
1. Whether CAB06 can be used to connect to a PC and read data via Modbus
2. Whether the communication settings of the PLC and SMR07 are consistent
3. Whether the Modbus ID is correct and whether there is an ID conflict
4. Whether the PLC Modbus address is correct
5. Whether the power supply is stable at 5–12 VDC
6. Whether RS485 wiring is correct
7. Whether the connection has been completed according to the appendix wiring diagram
Q9: What happens if RS485 D+ / D- are reversed?
If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
- Communication and Wiring Questions
-
Q4
Modbus Reading Questions
D- Modbus Reading Questions
Q1: Which address should be used to read the main parameter value?
The main parameter value can be read using PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with the data type set to floating point.
Q2: How can the temperature value be read?
The temperature value can be read from PLC address 2, HEX address 0x0001, Function Code 04, with the data type set to floating point.
Q3: How can the location, area, or zone name be read?
The location, area, or zone name can be read from PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with the data type set to UTF-16BE string.
Q4: How can the device model be read?
The device model can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with the data type set to ASCII string
Q5: How can latitude and longitude be read?
Latitude and longitude can be read from PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type set to floating point.
Q6: How can the cleaning brush be activated?
If the device is equipped with a cleaning brush function, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.
Q7: How should floating-point data be decoded?
SMR21-2 floating-point data follows the IEEE 754 single-precision floating-point format and occupies 4 Bytes. If the decoded value is abnormal, confirm whether the high/low bytes or word order need to be swapped.
Q8: What should be noted about Modbus CRC?
Modbus RTU uses CRC16. During transmission, Low Byte is sent first and High Byte last, meaning the low-order byte comes before the high-order byte.
Q9: What do Function Codes 03, 04, and 10 represent?
Function Code 03: Read Holding Registers
Function Code 04: Read Input Registers
Function Code 10: Write Multiple Holding Registers
- Modbus Reading Questions
-
Q5
Calibration Questions
E- Calibration Questions
Q1: Is calibration required before initial use?
In general, no. SMR07 has been calibrated before shipment. During initial use, measurement can begin once the device is properly installed according to the manual. If accuracy needs to be confirmed before use, standard calibration can be performed.
Q2: When is recalibration required?
Recalibration is recommended when on-site readings differ significantly from standard solution or laboratory measurements, when readings drift after long-term use, after electrode replacement, when accuracy needs to be confirmed after maintenance or cleaning, or when on-site sample conditions differ significantly from the standard solution.
Q3: What calibration items are available?
Calibration items include temperature calibration, standard zero-point calibration, standard slope calibration, and on-site calibration.
Q4: What should be done before calibration?
The analyzer must be thoroughly cleaned before calibration. It is recommended to remove the electrode protective cap, rinse the surface with tap water, clean with a soft brush and neutral detergent if necessary, dry all parts of the body as much as possible, and shake the electrode dry.
Q5: What tools are required for calibration?
The following tools are required:
Calibrated standard thermometer
WTW conductivity meter
Calibration cup
Magnetic stir bar and magnetic stirrer Ultrapure water
Q6: What standard solutions are required for calibration?
Conductivity 1.413 mS/cm standard solution and 35.1 ppt standard solution can be used.
Q7: How is temperature calibration performed?
Pour purified water or tap water into the calibration cup, then place the analyzer and standard thermometer into the cup. Wait at least 2 minutes and confirm that the temperature is stable. Then write “real-time temperature value − thermometer reading” as the temperature correction value.
Q8: Which address should the temperature correction value be written to?
The temperature correction value can be written to PLC address 714, HEX address 0x02C9, Function Code 10, with the data type set to floating point.
Q9: What water should be used for zero-point calibration?
Zero-point calibration must use purified water that meets the required specification, such as Type II grade purified water with conductivity below 1.0 µS/cm, or higher-grade purified water.
Q10: How is zero-point calibration performed?
Place purified water and a magnetic stir bar into the calibration container, immerse the analyzer to the appropriate depth, and start the magnetic stirrer at a constant speed while avoiding large vortices and dense bubbles. Then enter 0 into the zero-point calibration value, enter 1 to perform zero-point calibration, and finally query the real-time conductivity value to confirm whether calibration has been completed.
Q11: Which address should the zero-point calibration value be written to?
The zero-point calibration value can be written to PLC address 519, HEX address 0x0206, Function Code 10, with the data type set to floating point.
Q12: Which address should be written to perform zero-point calibration?
Input 1 at PLC address 518, HEX address 0x0205, Function Code 10, to perform zero-point calibration.
Q13: How is SMR07 slope calibration performed?
Place the standard solution and magnetic stir bar into the calibration container, immerse the analyzer to the appropriate depth, start the magnetic stirrer at a constant speed, and wait until the real-time conductivity value stabilizes. Then enter the standard solution concentration into the slope calibration value, and enter 1 to perform slope calibration.
Q14: Which address should the slope calibration value be written to?
The slope calibration value can be written to PLC address 521, HEX address 0x0208, Function Code 10, with the data type set to floating point.
Q15: Which address should be written to perform slope calibration?
Input 1 at PLC address 520, HEX address 0x0207, Function Code 10, to perform slope calibration.
Q16: What concentration is recommended for slope calibration?
It is recommended to select the standard solution concentration according to the target measurement value. The chemicals listed in the manual include conductivity 1.413 mS/cm standard solution and 35.1 ppt standard solution.
Q17: What are Zero lock and Span lock?
Zero lock and Span lock are locked values in the calibration logic, used for the zero-point reference and slope reference. Zero lock can be read from PLC address 733, HEX address 0x02DC; Span lock can be read from PLC address 734, HEX address 0x02DD.
Q18: What should be noted before on-site calibration?
After the analyzer is installed on site, wait until the reading stabilizes. The target calibration value should preferably refer to laboratory test results. After successful calibration, the slope value will be updated.
Q19: Why should large vortices and dense bubbles be avoided during calibration?
Large vortices and dense bubbles may interfere with sensor readings, causing unstable conductivity values and affecting calibration results.
- Calibration Questions
-
Q7
Abnormal Readings and Troubleshooting
F- Abnormal Readings and Troubleshooting
Q1: What should I do if readings keep fluctuating?
First clean the sensing area, check whether the water inlet is blocked, whether air bubbles remain, whether the installation is secure, whether the on-site water flow is excessively disturbed, and whether the signal cable is affected by interference from power cables.
Q1: What should I do if the reading remains at 0?
Confirm whether the sample and standard solution are normal, then check whether the electrode is clean, whether it is immersed deeply enough, and whether calibration is correct. If there is still no response when testing with a standard solution, contact AQUAS.
Q3: What should I do if the reading remains at the maximum range?
Possible causes include incorrect calibration, electrode contamination, electrode aging, air bubble interference, or sensing area obstruction. It is recommended to clean the analyzer first and reconfirm standard calibration.
Q4: What should I do if readings are clearly too high or too low?
Check whether standard calibration is correct, whether the electrode is dirty or damaged, whether the sample is homogeneous, and whether bubbles or bottom sludge are causing interference. Recalibrate if necessary.
Q5: What should I do if the response time becomes slower?
This may be caused by electrode aging, contamination in the sensing area, unstable sample conditions, or degraded electrode performance. It is recommended to clean the electrode first and confirm the standard solution reading. Activate or replace the electrode if necessary.
Q6: What should I do if readings from the same sample are inconsistent?
Confirm whether the sample is homogeneous, whether stirring is stable, whether stratification or sediment is present, whether bubbles are attached, and whether the calibration procedure is correct.
Q7: What should I do if the reading does not change?
Restore factory settings first and repeat the calibration procedure. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.
Q8: If on-site readings differ from laboratory measurements, does that mean the device is faulty?
Not necessarily. Real-time monitoring instruments are directly exposed to the on-site environment, while laboratory samples usually undergo sampling, storage, pretreatment, and standardized procedures. Different test conditions and time points may lead to differences in values.
Q9: Do air bubbles affect SMR07 readings?
Yes. Air bubbles may interfere with sensor readings, causing unstable conductivity values or calibration errors. Stable water flow, correct installation, and avoiding dense bubbles attached to the sensing area are recommended.
Q10: Do bottom sludge or sediments affect SMR07 readings?
Yes. If the analyzer is buried by bottom sludge or the sensing area is blocked by sediment, readings may become abnormal, unstable, or slower to respond.
Q11: Can the SMR07 electrode be activated?
Yes. The manual recommends the following EC electrode activation procedure: first rinse the electrode surface with purified water, then soak it in 0.15M NaOH for at least 30 minutes, rinse with clean water, and then soak it in 0.15M HCl for at least 30 minutes. The order must be alkaline first, then acidic, and must not be changed. After activation, perform calibration according to the standard procedure and observe whether the values are normal.
Q12: What should I do if the reading is still incorrect after electrode activation?
If the reading is still incorrect after activation and calibration, electrode replacement is recommended. If the new electrode is still abnormal, contact AQUAS.
Q13: When should AQUAS technical support be contacted?
If the device still cannot read values or shows no response after checking power supply, wiring, Modbus ID, communication settings, standard solution testing, calibration, cleaning, electrode activation, and electrode inspection, it is recommended to contact AQUAS technical support and provide on-site photos, wiring photos, Modbus settings, standard solution test records, and reading screenshots.
- Abnormal Readings and Troubleshooting
-
Q8
Maintenance Questions
F- Maintenance Questions
Q1: Is regular maintenance required?
Yes. To keep the analyzer in optimal operating condition, regular maintenance and cleaning are required, especially in environments where oil, sediment, or other contaminants may adhere to the device.
Q2: How often should it be cleaned?
Cleaning is recommended approximately once per week. The cleaning frequency can be adjusted according to on-site water quality conditions.
Q3: How often should calibration solution be used for calibration?
Calibration with calibration solution is recommended approximately once every two weeks. The frequency can be adjusted according to on-site water quality conditions.
Q4: How should the analyzer be stored when not used for a long time?
If the analyzer will not be used for an extended period, clean it thoroughly and store it dry.
Q5: What is the recommended maintenance cleaning procedure?
It is recommended to first clean the entire unit with purified water and a soft brush, then rinse it again with purified water and confirm that the electrode and body are free of dirt. If obvious contamination remains, soak it in 5% diluted sulfuric acid for 15 minutes, then place it in purified water for zero-point verification.
Q6: What conditions should the reading meet during zero-point verification?
During zero-point verification in purified water, the parameter reading should be less than 5% of full scale. If the reading remains greater than 15% of full scale after acid soaking, return-to-factory maintenance is recommended.
Q7: What standard solution test error is considered acceptable?
When testing with a calibration solution, a parameter reading error below 5% can be considered maintenance complete. If the error is between 5% and 20%, single-point slope calibration can be performed. If the error is greater than 20%, check whether there are bubbles in the sensing area or whether the standard solution has deteriorated or been contaminated.
Q8: What consumables are used?
The main consumables include conductivity standard solution and purified water.
Q9: When should the electrode be replaced?
Electrode replacement is recommended when the electrode response becomes slower, readings are unstable, electrode performance decreases significantly, or physical damage such as cracks, wear, or breakage is observed.
Q10: Can an electrode that has been used for more than one year continue to be used?
If the electrode has been used for more than one year, it may have aged. Even if readings are still available, it is recommended to evaluate whether replacement is required based on standard solution calibration and reading stability.
- Maintenance Questions