SMR04 pH and Temperature Analyzer

Frequently Asked Questions

  • A
    1. Product and Basic Use
      Q1: What is the product?
      It is an AQUAS pH and temperature analyzer designed for real-time pH and temperature monitoring in water, with digital data output via RS485 communication.

      Q2: What models are available?
      SMR04 mainly includes:
      SMR04-3: pH and temperature analyzer
      SMR04-5: differential pH and temperature analyzer

      Q3: What is the difference between SMR04-3 and SMR04-5?
      SMR04-3 is a standard pH and temperature analyzer with a replaceable electrode and standard M12 connector.
      SMR04-5 is a differential pH and temperature analyzer with a differential electrode design, suitable for heavily polluted or harsh on-site environments.


      Q4: What parameters can it measure?
      It mainly measures pH value and temperature, and includes temperature compensation.

      Q5: What is the measurement principle?
      SMR04-3 measures pH by detecting the potential change generated on the electrode through hydrogen ion response, followed by signal processing and temperature compensation.
      SMR04-5 uses a differential design, measuring pH with two glass electrodes relative to a third metal electrode in a differential configuration, improving contamination resistance and measurement stability.

      Q6: What application sites is it suitable for?
      It is suitable for drinking water, surface water, groundwater, industrial processes, water treatment, wastewater, and other applications.

      Q7: Is a control box required?
      Not necessarily. SMR04 has an integrated compact enclosure with a built-in transmitter and sensor. It features a rugged IP68 submersible, corrosion-resistant housing and can be installed directly according to site requirements.

      Q8: Does it support PLC or HMI?
      Yes. SMR04 uses the standard Modbus RTU communication protocol and can connect to PLCs, HMIs, or AQUAS data loggers via RS485.

      Q9: Does it have explosion-proof certification?
      Yes. SMR04 is certified with IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS explosion-proof certification.

      Q10: How long is the electrode service life?
      According to the SMR04-5 manual, the electrode service life can be up to more than 12 months. However, the actual service life may vary depending on on-site water quality, contamination level, maintenance frequency, and operating environment.
  • B
    1. Installation Questions
      Q1: What is the operating voltage?
      The operating voltage is 5–12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.

      Q2: What installation methods are available?
      Immersion, flow-through, and pipe-insertion installation methods are available.

      Q3: What should be noted during installation?
      During installation, avoid pulling or damaging the connection cable. The cable should be long enough and properly secured. It is recommended to install the analyzer at a location with stable water flow to obtain better measurement results.

      Q4: How deep should the sensing surface be submerged?
      During testing, the sensing surface must be fully submerged at least 10 cm below the water surface.

      Q5: How long does it take to stabilize after installation?
      After installation, the analyzer should be allowed to stabilize for at least 5 minutes. The actual stabilization time may vary depending on sample characteristics and installation conditions.

      Q6: Can it be installed close to the bottom of the tank?
      No. The analyzer should not directly contact or be installed too close to the bottom of the tank, as this may cause abnormal measurement due to sensing area obstruction or sediment interference.

      Q7: Does it need to be vertically fixed during on-site installation?
      Yes. During on-site installation, it is recommended to vertically fix the analyzer and use a protection tube and mounting clamps to secure the device, preventing impact and installation displacement.

      Q8: What should be noted when installing outdoors or in high-load environments?
      When installed outdoors or in high-load environments, it is recommended to use AQUAS dedicated protective covers and fixtures to improve overall protection and measurement stability.

      Q9: How should SMR04 be set up for laboratory testing?
      The analyzer should be connected to a computer, vertically fixed on a stand, inserted into the calibration container, and the electrode protection bottle should be removed. The electrode sensing area must be fully immersed in the sample water while avoiding contact with the bottom or proximity to the container wall.

      Q10: Is stirring required during laboratory testing?
      It is recommended to place a 2.5 cm magnetic stir bar at the bottom of the container and use a magnetic stirrer to maintain stable rotation at approximately 300 rpm, preventing sample stratification and sediment from affecting the results.
  • C
    1. Communication and Wiring Questions
      Q1: Which communication interface is used?
      RS485 communication interface is used.

      Q2: Which communication protocol is used?
      The standard Modbus RTU communication protocol is used.

      Q3: What are the default communication settings?
      Common communication settings are as follows:
      Communication interface: RS485
      Baud rate: 19200 bps
      Data length: 8 bit
      Parity: None
      Stop bit: 1 bit
      Communication mode: RTU
      The device supports 9600 and 19200 bps.

      Q4: Can it be connected directly to a computer?
      Yes. It can be connected to a computer using CAB06 USB to RS485 Cable, and values can be read, calibrated, and parameters configured through a Modbus tool or AQCFG.

      Q5: What is the default device address?
      The factory default device address is 01. If the device address is changed, the device address in the PLC, HMI, host, or reading software must also be updated accordingly.

      Q6: How should RS485 wiring be connected?
      The SMR male connector wiring is as follows:
      PIN 1: GND
      PIN 2: RS485-
      PIN 3: RS485+
      PIN 4: 5–12 VDC
      PIN 5: Not used
      New wire colors are typically: black for GND, white for RS485-, blue for RS485+, and red for 5–12V DC.

      Q7: Can it be connected to an AQUAS data logger?
      Yes. SMR02 can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data logging and remote monitoring according to site requirements.

      Q8: What should be checked first if the PLC cannot read values?
      Please check the following in order:
      1. Whether CAB06 can be used to connect to a PC and read data via Modbus
      2. Whether the communication settings of the PLC and SMR02 are consistent
      3. Whether the Modbus ID is correct and whether there is an ID conflict
      4. Whether the PLC Modbus address is correct
      5. Whether the power supply is stable at 5–12 VDC
      6. Whether RS485 wiring is correct
      7. Whether the connection has been completed according to the appendix wiring diagram

      Q9: What happens if RS485 D+ / D- are reversed?
      If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
  • D
    1. Modbus Reading Questions
      Q1: Which address should be used to read the main parameter value?
      The main parameter value can be read using PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with the data type set to floating point.

      Q2: How can the temperature value be read?
      The temperature value can be read from PLC address 2, HEX address 0x0001, Function Code 04, with the data type set to floating point.

      Q3: How can the location, area, or zone name be read?
      The location, area, or zone name can be read from PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with the data type set to UTF-16BE string.

      Q4: How can the device model be read?
      The device model can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with the data type set to ASCII string

      Q5: How can latitude and longitude be read?
      Latitude and longitude can be read from PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type set to floating point.

      Q6: How can the cleaning brush be activated?
      If the device is equipped with a cleaning brush function, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.

      Q7: How should floating-point data be decoded?
      SMR21-2 floating-point data follows the IEEE 754 single-precision floating-point format and occupies 4 Bytes. If the decoded value is abnormal, confirm whether the high/low bytes or word order need to be swapped.

      Q8: What should be noted about Modbus CRC?
      Modbus RTU uses CRC16. During transmission, Low Byte is sent first and High Byte last, meaning the low-order byte comes before the high-order byte.

      Q9: What do Function Codes 03, 04, and 10 represent?
      Function Code 03: Read Holding Registers
      Function Code 04: Read Input Registers
      Function Code 10: Write Multiple Holding Registers

  • E
    1. Calibration Questions
      Q1: Is calibration required before initial use?
      In general, no. SMR04 has been calibrated before shipment. During initial use, measurement can begin once the device is properly installed according to the manual. If accuracy needs to be confirmed before use, standard calibration can be performed.

      Q2: When is recalibration required?
      Recalibration is recommended when on-site readings differ significantly from standard solution or laboratory measurements, when readings drift after long-term use, after electrode replacement, when accuracy needs to be confirmed after maintenance or cleaning, or when on-site sample conditions differ significantly from the standard solution.

      Q3: What calibration items are available?
      Calibration items include temperature calibration and pH standard calibration.

      Q4: What should be done before calibration?
      The analyzer must be thoroughly cleaned before calibration. It is recommended to remove the electrode protective cap, rinse the surface with tap water, clean with a soft brush and neutral detergent if necessary, dry all parts of the body as much as possible, and shake the electrode dry.

      Q5: What tools are required for calibration?
      The following tools are required:
      Calibrated standard thermometer
      Calibration cup
      Magnetic stir bar and magnetic stirrer Ultrapure water

      Q6: What standard solutions are required for calibration?
      pH standard solutions are required, such as pH 4, pH 7, and pH 10.

      Q7: How is temperature calibration performed?
      Pour purified water or tap water into the calibration cup, then place the analyzer and standard thermometer into the cup. Wait at least 2 minutes and confirm that the temperature is stable. Then write “real-time temperature value − thermometer reading” as the temperature correction value.

      Q8: Which address should the temperature correction value be written to?
      The temperature correction value can be written to PLC address 12289, HEX address 0x3000, Function Code 10, with the data type set to floating point.

      Q9: How is pH standard calibration performed?
      Using pH 7 as an example, place the pH 7 standard solution and magnetic stir bar into the calibration container, immerse the analyzer to the appropriate depth, start the magnetic stirrer at a constant speed, and wait until the real-time pH value stabilizes. Then enter the standard solution concentration into the slope calibration value. After completion, query the real-time pH value again to confirm the calibration result.

      Q10: Which address should the slope calibration value be written to?
      The pH slope calibration value can be written to PLC address 518, HEX address 0x0205, Function Code 10, with the data type set to floating point.

      Q11: Which address should be used to read the pH millivolt value?
      The pH millivolt value can be read from PLC address 49, HEX address 0x0030, Function Code 04, with the data type set to floating point.

      Q12: Which address should be written for pH potential calibration?
      For pH potential calibration, input 1 at PLC address 289, HEX address 0x0120, Function Code 10, to write the current potential value.

      Q13: How can I confirm whether the pH potential value has been written?
      Read PLC address 258, HEX address 0x0101, Function Code 04, to confirm whether the calibration potential value has been written.

      Q14: Which addresses should be used to read the potential slope values for pH 4 and pH 10?
      The potential slope value from pH 7 to pH 4 can be read from PLC address 262, HEX address 0x0105.
      The potential slope value from pH 7 to pH 10 can be read from PLC address 263, HEX address 0x0106.

      Q15: What range should the mV value fall within during pH calibration?
      The reference ranges are as follows:
      pH 7: approximately 0 ± 25 mV
      pH 4: approximately 175 ± 25 mV
      pH 10: approximately -175 ± 25 mV
      The potential slope should be between -56 mV/pH and -61 mV/pH.
      Q16: How deep should the electrode be immersed during pH calibration?
      When immersing the electrode in the standard solution, ensure that the front end of the electrode is at least 3 cm below the liquid surface and that the electrode does not touch the bottom of the calibration cup.

      Q17: Why should large vortices and dense bubbles be avoided during calibration?
      Large vortices and dense bubbles may interfere with electrode readings, causing unstable pH or mV values and affecting calibration results.

      Q18: What should be noted when cleaning the pH electrode?
      When cleaning the electrode, rinse the pH electrode with purified water and dry it with lens cleaning paper to prevent residual standard solution or contaminants from affecting the next calibration step.

  • F
    1. Abnormal Readings and Troubleshooting
      Q1: What should I do if readings keep fluctuating?
      First clean the sensing area, check whether the water inlet is blocked, whether air bubbles remain, whether the installation is secure, whether the on-site water flow is excessively disturbed, and whether the signal cable is affected by interference from power cables.

      Q1: What should I do if the reading remains at 0?
      Confirm whether the sample and standard solution are normal, then check whether the electrode is clean, whether it is immersed deeply enough, and whether calibration is correct. If there is still no response when testing with a standard solution, contact AQUAS.

      Q3: What should I do if the reading remains at the maximum range?
      Possible causes include incorrect calibration, electrode contamination, electrode aging, air bubble interference, or sensing area obstruction. It is recommended to clean the analyzer first and reconfirm standard calibration.

      Q4: What should I do if readings are clearly too high or too low?
      Check whether standard calibration is correct, whether the electrode is dirty or damaged, whether the sample is homogeneous, and whether bubbles or bottom sludge are causing interference. Recalibrate if necessary.

      Q4: What should I do if the response time becomes slower?
      This may be caused by electrode aging, contamination in the sensing area, unstable sample conditions, or degraded electrode performance. It is recommended to clean the electrode first and confirm the standard solution reading. Activate or replace the electrode if necessary.

      Q5: What should I do if readings from the same sample are inconsistent?
      Confirm whether the sample is homogeneous, whether stirring is stable, whether stratification or sediment is present, whether bubbles are attached, and whether the calibration procedure is correct.

      Q7: What should I do if the reading does not change?
      Restore factory settings first and repeat the calibration procedure. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.

      Q8: If on-site readings differ from laboratory measurements, does that mean the device is faulty?
      Not necessarily. Real-time monitoring instruments are directly exposed to the on-site environment, while laboratory samples usually undergo sampling, storage, pretreatment, and standardized procedures. Different test conditions and time points may lead to differences in values.

      Q9: Do cracks or wear on the electrode affect readings?
      Yes. If the electrode is cracked, damaged, or worn, readings may become unstable, response may slow down, or calibration may fail. The electrode should be replaced immediately.

      Q10: Can the pH electrode be activated?
      Yes. The manual recommends the following electrode activation procedure: first rinse the electrode surface with purified water, then soak it in 0.15M NaOH for at least 30 minutes, rinse with clean water, and then soak it in 0.15M HCl for at least 30 minutes. The order must be alkaline first, then acidic, and must not be changed. After activation, perform calibration according to the standard procedure and observe whether the mV value is normal.

      Q11: What should I do if the reading is still incorrect after electrode activation?
      If the reading is still incorrect after activation and calibration, electrode replacement is recommended. If abnormalities remain after electrode replacement, contact AQUAS.

      Q12: When should AQUAS technical support be contacted?
      If the device still cannot read values or shows no response after checking power supply, wiring, Modbus ID, communication settings, standard solution testing, calibration, cleaning, electrode activation, and electrode inspection, it is recommended to contact AQUAS technical support and provide on-site photos, wiring photos, Modbus settings, standard solution test records, and reading screenshots.
  • F
    1. Maintenance Questions
      Q1: Is regular maintenance required?
      Yes. To keep the level meter in optimal operating condition, regular maintenance and cleaning are required, especially in environments where oil, sediment, or other contaminants may adhere to the device.

      Q2: How often should routine maintenance be performed?
      Routine maintenance is recommended at least once per week. The actual frequency can be adjusted according to on-site water quality, bottom sludge, oil and grease, and the monitoring environment.

      Q3: What should be included in routine maintenance?
      Routine maintenance should include:
      Cleaning the analyzer surface and sensing area
      Checking the battery level, if using a battery-powered model
      Checking consumable status
      Checking whether cables and mounting fixtures are secure
      Confirming whether data transmission is stable

      Q4: How often should periodic maintenance be performed?
      Periodic maintenance is recommended at least once per month.

      Q5: What is included in periodic maintenance?
      Periodic maintenance includes trend calibration and standard calibration. Long-term data trends can be used to determine whether instrument adjustment is required.

      Q6: What consumables are used?
      The main consumables include pH standard solutions and purified water. Electrodes should also be replaced periodically depending on usage conditions.

      Q7: When should the electrode be replaced?
      Electrode replacement is recommended when the electrode response becomes slower, readings are unstable, the electrode slope decreases significantly, or physical damage such as cracks or breakage is observed.

      Q8: Can an electrode that has been used for more than one year continue to be used?
      If the electrode has been used for more than one year, it may have aged. Even if readings are still available, it is recommended to evaluate whether replacement is required based on standard solution calibration, mV range, and slope performance.

      Q9: How should pH, ORP, and Ref electrodes be stored when not used for a long time?
      When the electrodes are not used for an extended period, clean the sensor thoroughly and store the pH, ORP, and Ref electrodes immersed in 3M KCl storage solution.

      Q10: Is calibration required after replacing an electrode?
      Yes, it is recommended. After replacing an electrode, confirm that all connections are correctly installed and perform calibration again to ensure accurate measurement results.