SMR63 Radar Flow Velocity and Level Meter

Radar Flow Velocity and Level Meter

  • A
    1. Product and Basic Use
      Q1: What is the product?
      It is a chemical oxygen demand and suspended solids analyzer designed for real-time monitoring in water and digital data output via RS485 communication.

      Q2: What can it measure?
      It mainly measures chemical oxygen demand and also provides turbidity and suspended solids measurement with automatic interference compensation.

      Q3: What is the measurement principle?
      It uses a dual LED light source for measurement. When organic compounds containing conjugated double bonds or polycyclic aromatic hydrocarbons are present in water, they absorb 254 nm ultraviolet light, and the device calculates COD based on the absorption level. At the same time, a 530 nm optical beam system detects turbidity or suspended solids concentration and performs automatic compensation.

      Q4: Does it require chemical reagents?
      No. SMR44 uses optical measurement and does not require chemical reagents, making it suitable for real-time continuous monitoring.

      Q5: What application sites is it suitable for?
      It is suitable for drinking water, surface water, industrial processes, water treatment, wastewater, and other applications.

      Q5: Is a control box required?
      Not necessarily. It has an integrated compact enclosure with a built-in transmitter and sensor. It features a rugged IP68 submersible, corrosion-resistant housing and can be installed directly according to site requirements.

      Q6: Does it support PLC or HMI?
      Yes. It uses the standard Modbus RTU communication protocol and can connect to PLCs, HMIs, or AQUAS data loggers via RS485.

      Q7: Does it have explosion-proof certification?
      Yes. It is certified with IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS explosion-proof certification.

      Q8: Does it have an automatic cleaning function?
      Yes. It can be equipped with an automatic cleaning brush and can also be used with an automatic cleaning device to clean the membrane cap with water or air, reducing maintenance needs caused by optical window contamination.
  • B
    1. Installation Questions
      Q1: What is the operating voltage?
      The operating voltage is 5–12 VDC. Use a stable power supply to prevent equipment damage or unstable communication caused by abnormal voltage.

      Q2: What installation methods are available?
      Immersion, flow-through, and pipe-insertion installation methods are available.

      Q3: What should be noted during installation?
      During installation, avoid pulling or damaging the connection cable. The cable should be long enough and properly secured. It is recommended to install the analyzer at a location with stable water flow to obtain better measurement results.

      Q4: How deep should the sensing surface be submerged?
      During testing, the sensing surface must be fully submerged at least 10 cm below the water surface.

      Q5: How long does it take to stabilize after installation?
      After installation, the analyzer should be allowed to stabilize for at least 5 minutes. The actual stabilization time may vary depending on sample characteristics and installation conditions.

      Q6: Can it be installed close to the bottom of the tank?
      No. The analyzer should not directly contact or be installed too close to the bottom of the tank, as this may cause abnormal measurement due to sensing area obstruction or sediment interference.

      Q7: Is a protective cover recommended during installation?
      Yes. If there is a risk of the sensing area contacting the bottom, impact, or high-load conditions, it is recommended to use a dedicated protective cover, protection tube, or fixing fixture to ensure stable measurement.

      Q8: Is it suitable for outdoor environments without mains power?
      Yes. With its ultra-low power consumption design, it is suitable for battery or solar power supply and can be used in outdoor or remote monitoring sites.
  • C
    1. Communication and Wiring Questions
      Q1: Which communication interface is used?
      RS485 communication interface is used.

      Q2: Which communication protocol is used?
      The standard Modbus RTU communication protocol is used.

      Q3: What are the default communication settings?
      Common communication settings are as follows:
      Communication interface: RS485
      Baud rate: 19200 bps
      Data length: 8 bit
      Parity: None
      Stop bit: 1 bit
      Communication mode: RTU
      The device supports 9600 and 19200 bps.

      Q4: Can it be connected directly to a computer?
      Yes. It can be connected to a computer using CAB06 USB to RS485 Cable, and values can be read or calibration settings configured through a Modbus tool.

      Q5: What is the default device address?
      The factory default device address is 01. If the device address is changed, the device address in the PLC, HMI, host, or reading software must also be updated accordingly.

      Q6: How should RS485 wiring be connected?
      The SMR male connector wiring is as follows:
      PIN 1: GND
      PIN 2: RS485-
      PIN 3: RS485+
      PIN 4: 5–12 VDC
      PIN 5: Not used
      New wire colors are typically: black for GND, white for RS485-, blue for RS485+, and red for 5–12V DC.

      Q7: Can it be connected to an AQUAS data logger?
      Yes. It can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data logging and remote monitoring according to site requirements.

      Q8: What should be checked first if the PLC cannot read values?
      Please check the following in order:
      1. Whether CAB06 can be used to connect to a PC and read data via Modbus
      2. Whether the communication settings of the PLC and SMR21-2 are consistent
      3. Whether the Modbus ID is correct and whether there is an ID conflict
      4. Whether the PLC Modbus address is correct
      5. Whether the power supply is stable at 5–12 VDC
      6. Whether RS485 wiring is correct
      7. Whether the connection has been completed according to the appendix wiring diagram

      Q9: What happens if RS485 D+ / D- are reversed?
      If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
  • D
    1. Modbus Reading Questions
      Q1: Which address should be used to read the main parameter value?
      The main parameter value can be read using PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with the data type set to floating point.

      Q2: Which address should be used to read the secondary parameter value?
      The secondary parameter value can be read using PLC address 2, HEX address 0x0001, Length 2 Words, Function Code 04, with the data type set to floating point.

      Q3: How can the device model be read?
      The device model can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with the data type set to ASCII string.

      Q4: How can the location, area, or zone name be read?
      The location, area, or zone name can be read from PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with the data type set to UTF-16BE string.

      Q5: How can the device model be read?
      The device model can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with the data type set to ASCII string

      Q6: How can latitude and longitude be read?
      Latitude and longitude can be read from PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type set to floating point.

      Q7: How can the cleaning brush be activated?
      If the device is equipped with a cleaning brush function, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.

      Q8: How should floating-point data be decoded?
      SMR21-2 floating-point data follows the IEEE 754 single-precision floating-point format and occupies 4 Bytes. If the decoded value is abnormal, confirm whether the high/low bytes or word order need to be swapped.

      Q9: What should be noted about Modbus CRC?
      Modbus RTU uses CRC16. During transmission, Low Byte is sent first and High Byte last, meaning the low-order byte comes before the high-order byte.

      Q10: What do Function Codes 03, 04, and 10 represent?
      Function Code 03: Read Holding Registers
      Function Code 04: Read Input Registers
      Function Code 10: Write Multiple Holding Registers

  • E
    1. Calibration Questions
      Q1: Is calibration required before initial use?
      In general, no. The device has been calibrated before shipment. During initial use, measurement can begin once the device is properly installed according to the manual.

      Q2: When is recalibration required?
      Recalibration is recommended when on-site readings differ significantly from laboratory measurements, when readings drift after long-term use, when accuracy needs to be confirmed after maintenance or cleaning, or when on-site sample conditions differ significantly from the standard solution.

      Q3: What calibration methods are available?
      Calibration includes temperature calibration, standard calibration, on-site calibration, and background value subtraction. Standard calibration includes zero-point calibration and slope calibration.

      Q4: What should be done before calibration?
      Before calibration, thoroughly clean the entire analyzer and calibration-related vessels using detergent and a soft brush, and confirm that the optical window is fully immersed below the liquid surface. If necessary, activate the cleaning brush to remove attached bubbles and prevent abnormal readings.

      Q5: What tools are required for calibration?
      The following tools are required:
      Calibrated standard thermometer
      Calibration cup
      Magnetic stir bar
      Magnetic stirrer
      Ultrapure water

      Q6: What standard solutions are required for calibration?
      The following can be used:
      SL44-1 chemical oxygen demand standard solution, 50 mg/L, 500 ml
      SL44-2 chemical oxygen demand standard solution, 200 mg/L, 500 ml

      Q7: Which address should the temperature correction value be written to?
      The temperature correction value can be written to PLC address 12289, HEX address 0x3000, Function Code 10, with the data type set to floating point.

      Q8: What water should be used for zero-point calibration?
      Zero-point calibration must use purified water that meets the required specification, such as Type II grade purified water with conductivity below 1.0 µS/cm, or higher-grade purified water.

      Q9: Which address should the zero-point calibration value be written to?
      The zero-point calibration value can be written to PLC address 519, HEX address 0x0206. In addition, input 1 at PLC address 518, HEX address 0x0205, to perform zero-point calibration.

      Q10: What value should be observed during zero-point calibration?
      During zero-point calibration, read the temperature-compensated count value and perform calibration after the value becomes stable. When calibration is successful, the absorbance abs value should display 0 ± 0.5.

      Q11: Which address should the zero-point calibration value be written to?
      The zero-point calibration value can be written to PLC address 20491, HEX address 0x500A. In addition, the absorbance value at PLC address 50, HEX address 0x0031, can be read to confirm whether zero-point calibration has been completed.

      Q12: What standard solution should be used for slope calibration?
      Slope calibration uses a COD standard solution. The recommended concentration should be based on the target measurement value. The standard solutions listed in the manual include 50 mg/L and 200 mg/L COD standard solutions.

      Q13: Which address should the slope calibration value be written to?
      The standard slope calibration value can be written to PLC address 13062, HEX address 0x3305, Function Code 10, with the data type set to floating point.

      Q14: Which address should the on-site calibration value be written to?
      The on-site calibration value can be written to PLC address 13058, HEX address 0x3301, Function Code 10, with the data type set to floating point.

      Q15: What should be noted before on-site calibration?
      Since there may be a time difference between laboratory testing and on-site reading calibration, it is recommended to first observe the concentration trend and select a time period with relatively stable concentration for on-site reading calibration.

      Q16: When should the background value subtraction function be used?
      If the concentration trend of on-site samples has not been confirmed through laboratory testing, the background value subtraction function is not recommended.

      Q17: Which address should the background calibration value be written to?
      The background calibration value can be written to PLC address 28678, HEX address 0x7005, Function Code 10, with the data type set to floating point.

      Q17: Why should large vortices and dense bubbles be avoided during calibration?
      Large vortices and dense bubbles may interfere with optical window and absorbance readings, causing unstable COD readings or calibration errors.

      Q18: Can the cleaning brush be activated during calibration?
      Yes. The cleaning brush can be activated to remove bubbles attached to the optical window and prevent abnormal readings.

  • F
    1. Abnormal Readings and Troubleshooting
      Q1: What should I do if readings keep fluctuating?
      First clean the sensing area, confirm whether the installation depth is sufficient, whether the device is firmly secured, whether bubbles, bottom sludge, or contaminants are interfering on site, and whether the power supply and RS485 communication are normal.

      Q2: What should I do if the reading remains at 0?
      Confirm whether the sample actually contains COD concentration, then check whether the optical window is clean, whether it is immersed deeply enough, and whether calibration is correct. If there is still no response when testing with a standard solution, contact AQUAS.

      Q3: What should I do if the reading remains at the maximum range?
      Possible causes include the sample COD concentration exceeding the measurement range, excessively high suspended solids concentration, optical window contamination, incorrect calibration, or air bubble interference. It is recommended to clean the analyzer first and reconfirm calibration.

      Q4: What should I do if readings are clearly too high or too low?
      Check whether zero-point calibration and slope calibration are correct, whether the optical window is dirty, whether the sample is homogeneous, whether suspended solids concentration is too high, and whether bubbles or sediment are causing interference.

      Q5: What should I do if the response time becomes slower?
      This may be caused by optical window contamination, unstable sample conditions, suspended solids interference, or deposits affecting the sensing area. It is recommended to clean the optical window first and confirm the standard solution reading.

      Q6: What should I do if readings from the same sample are inconsistent?
      Confirm whether the sample is homogeneous, whether stirring is stable, whether stratification or sediment is present, whether bubbles are attached, and whether the calibration procedure is correct.

      Q7: What should I do if the reading does not change?
      Restore factory settings first and repeat the calibration procedure. If the reading still does not change after replacing the standard calibration solution, contact AQUAS technical support.

      Q8: How can I confirm whether the SMR44 light source is functioning properly?
      Insert a sheet of white paper into the optical path and visually check whether the lower optical window emits LED light at a fixed frequency. If there is no response, the light source may be abnormal. Please contact AQUAS.

      Q9: Can excessively high suspended solids affect COD measurement?
      Yes. Excessively high suspended solids concentration may cause abnormal testing or reading deviation. Confirm whether the target substance concentration and suspended solids concentration are within the analyzer measurement range.

      Q10: If on-site readings differ from laboratory measurements, does that mean the device is faulty?
      Not necessarily. Real-time monitoring instruments are directly exposed to the on-site environment, while laboratory samples usually undergo sampling, storage, pretreatment, and standardized procedures. Different test conditions and time points may lead to differences in values.

      Q11: When should AQUAS technical support be contacted?
      If the device still cannot read values or shows no response after checking power supply, wiring, Modbus ID, communication settings, standard solution testing, calibration, cleaning, and installation, it is recommended to contact AQUAS technical support and provide on-site photos, wiring photos, Modbus settings, standard solution test records, and reading screenshots.
  • F
    1. Maintenance and Consumables Questions
      Q1: Is regular maintenance required?
      Yes. SMR44 requires regular cleaning and calibration to ensure long-term stable and accurate measurement.

      Q2: How often should routine maintenance be performed?
      Cleaning is recommended approximately once per week. The cleaning frequency can be adjusted according to on-site water quality conditions.

      Q3: How often should calibration solution be used for calibration?
      Calibration with calibration solution is recommended approximately once every two weeks. The frequency can be adjusted according to on-site water quality conditions.

      Q4: How should the analyzer be stored when not used for a long time?
      If the analyzer will not be used for an extended period, clean it thoroughly and store it dry.

      Q5: What is the recommended maintenance cleaning procedure?
      It is recommended to first clean the entire unit with detergent and a soft brush, then rinse it with tap water and confirm that the optical window and body are free of dirt. If obvious contamination remains, soak it in 5% diluted sulfuric acid for 15 minutes, then place it in purified water for zero-point verification.

      Q6: What conditions should the reading meet during zero-point verification?
      During zero-point verification in purified water, the parameter reading should be less than 5% of full scale. If the reading remains greater than 15% of full scale after acid soaking, return-to-factory maintenance is recommended.

      Q7: What standard solution test error is considered acceptable?
      When testing with a calibration solution, a parameter reading error below 5% can be considered maintenance complete. If the error is between 5% and 20%, single-point slope calibration can be performed. If the error is greater than 20%, check whether there are bubbles on the optical window or whether the standard solution has deteriorated or been contaminated.

      Q8: What consumables are used?
      The main consumables include COD standard solution and purified water.

      Q9: Does the optical window require special protection?
      Yes. The SMR44 optical window uses abrasion-resistant and scratch-resistant sapphire glass with a nano coating to reduce sludge deposition and biological attachment. However, regular cleaning and inspection are still required.

      Q10: What should be noted when cleaning the optical window?
      When cleaning, use a soft cloth, soft brush, or neutral cleaner. Avoid hard objects or strongly corrosive cleaning methods that may scratch the optical surface. If acid soaking is required, follow the maintenance procedure using 5% diluted sulfuric acid and control the soaking time.