SMR23-2 Ammonium (NH4-N/NH4+) and pH Analyzer
Frequently Asked Questions
-
Q1
Product and Basic Use
A- Product and Basic Use
Q1: What is the product?
SMR23-2 is an AQUAS ammonia nitrogen NH4-N / NH4+ and pH analyzer. It outputs data through RS485 digital communication and supports remote status display, calibration, configuration, and fault diagnosis.
Q2: What parameters can it measure?
It mainly measures ammonia nitrogen NH4-N / NH4+, pH, and temperature. It also supports potassium ion K+ compensation to improve ammonia nitrogen measurement accuracy.
Q3: What is the measurement principle?
It uses the ion-selective electrode method. A selective membrane allows specific ions to pass through and generate an electrochemical potential, which is then converted into concentration according to the Nernst equation.
Q4: What application sites is it suitable for?
It is suitable for water quality monitoring applications such as drinking water, water treatment, wastewater, and aquaculture.
Q5: Is a control box required?
Not necessarily. SMR23-2 has an integrated compact enclosure with a built-in transmitter and sensor. It features a rugged IP68 submersible, corrosion-resistant housing and can be installed directly according to site requirements.
Q6: Does it support PLC or HMI?
Yes. SMR23-2 uses the standard Modbus RTU communication protocol and can connect to PLCs, HMIs, or AQUAS data loggers via RS485.
Q7: Does it have explosion-proof certification?
Yes. SMR23-2 is certified with IECEx CoC / ATEX Cat.1, Ex ia IIB T4/T5 Ga/TS explosion-proof certification.
Q8: How long is the electrode service life?
According to the manual, the electrode service life can be up to more than 12 months. However, the actual service life may vary depending on on-site water quality, contamination level, maintenance frequency, and operating environment.
- Product and Basic Use
-
Q2
Installation Questions
B- Installation Questions
Q1: What is the operating voltage?
The operating voltage is 5–12 VDC. The power supply must be stable to prevent equipment damage or unstable communication caused by abnormal voltage.
Q2: What installation methods are available?
Immersion, flow-through, and pipe-insertion installation methods are available.
Q3: What should be noted during installation?
During installation, avoid pulling or damaging the connection cable. The cable should be long enough and properly secured. It is recommended to install the analyzer at a location with stable water flow to obtain better measurement results.
Q4: How deep should the sensing surface be submerged?
During testing, the sensing surface must be fully submerged at least 10 cm below the water surface.
Q5: How long does it take to stabilize after installation?
After installation, the analyzer should be allowed to stabilize for at least 5 minutes. The actual stabilization time may vary depending on sample characteristics and installation conditions.
Q6: Can it be installed close to the bottom of the tank?
No. The analyzer should not directly contact or be installed too close to the bottom of the tank, as this may cause abnormal measurement due to sensing area obstruction or sediment interference.
Q7: Does it need to be vertically fixed during on-site installation?
Yes. During on-site installation, it is recommended to vertically fix the analyzer and use a protection tube and mounting clamps to secure the device, preventing impact and installation displacement.
Q8: What should be noted when installing outdoors or in high-load environments?
When installed outdoors or in high-load environments, it is recommended to use AQUAS dedicated protective covers and fixtures to improve overall protection and measurement stability.
Q9: How should SMR23-2 be set up for laboratory testing?
The analyzer should be connected to a computer, vertically fixed on a stand, inserted into the calibration container, and the electrode protection bottle should be removed. The electrode sensing area must be fully immersed in the sample water while avoiding contact with the bottom or proximity to the container wall.
Q10: Is stirring required during laboratory testing?
It is recommended to place a 2.5 cm magnetic stir bar at the bottom of the container and use a magnetic stirrer to maintain stable rotation at approximately 300 rpm, preventing sample stratification and sediment from affecting the results.
- Installation Questions
-
Q3
Communication and Wiring Questions
C- Communication and Wiring Questions
Q1: Which communication interface is used?
RS485 communication interface is used.
Q2: Which communication protocol is used?
The standard Modbus RTU communication protocol is used.
Q3: What are the default communication settings?
Common communication settings are as follows:
Communication interface: RS485
Baud rate: 19200 bps
Data length: 8 bit
Parity: None
Stop bit: 1 bit
Communication mode: RTU
The device supports 9600 and 19200 bps.
Q4: Can it be connected directly to a computer?
Yes. It can be connected to a computer using CAB06 USB to RS485 Cable, and values can be read or calibration settings configured through a Modbus tool.
Q5: What is the default device address?
The factory default device address is 01. If the device address is changed, the device address in the PLC, HMI, host, or reading software must also be updated accordingly.
Q6: How should RS485 wiring be connected?
The SMR male connector wiring is as follows:
PIN 1: GND
PIN 2: RS485-
PIN 3: RS485+
PIN 4: 5–12 VDC
PIN 5: Not used
New wire colors are typically: black for GND, white for RS485-, blue for RS485+, and red for 5–12V DC.
Q7: Can it be connected to an AQUAS data logger?
Yes. SMR23-2 can be connected to AQUAS data loggers such as ECO, HMI1, HMI2, and FLO for data logging and remote monitoring according to site requirements.
Q8: What should be checked first if the PLC cannot read values?
Please check the following in order:
1. Whether CAB06 can be used to connect to a PC and read data via Modbus
2. Whether the communication settings of the PLC and SMR21-2 are consistent
3. Whether the Modbus ID is correct and whether there is an ID conflict
4. Whether the PLC Modbus address is correct
5. Whether the power supply is stable at 5–12 VDC
6. Whether RS485 wiring is correct
7. Whether the connection has been completed according to the appendix wiring diagram
Q9: What happens if RS485 D+ / D- are reversed?
If RS485+ and RS485- are reversed, the PLC, HMI, or PC may be unable to communicate with the analyzer.
- Communication and Wiring Questions
-
Q4
Modbus Reading Questions
D- Modbus Reading Questions
Q1: Which address should be used to read the main parameter value?
The main parameter value can be read using PLC address 1, HEX address 0x0000, Length 2 Words, Function Code 04, with the data type set to floating point.
Q2: Which address should be used to read the secondary parameter value?
The secondary parameter value can be read using PLC address 2, HEX address 0x0001, Length 2 Words, Function Code 04, with the data type set to floating point.
Q3: How can the real-time temperature value be read?
The real-time temperature value can be read from PLC address 3, HEX address 0x0002, Function Code 04, with the data type set to floating point.
Q4: How can the real-time pH value be read?
The real-time pH value can be read from PLC address 4, HEX address 0x0003, Function Code 04, with the data type set to floating point.
Q5: How can the device model be read?
The device model can be read from PLC address 4098, HEX address 0x1001, Length 20 Words, Function Code 04, with the data type set to ASCII string
Q6: How can the location, area, or zone name be read?
The location, area, or zone name can be read from PLC address 9, HEX address 0x0008, Length 41 Words, Function Code 03, with the data type set to UTF-16BE string.
Q7: How can latitude and longitude be read?
Latitude and longitude can be read from PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type set to floating point.
Q8: How can the cleaning brush be activated?
If the device is equipped with a cleaning brush function, write to PLC address 5, HEX address 0x0004, Function Code 10. The data definition is 0 for standby and 1 for activation.
Q9: How should floating-point data be decoded?
SMR21-2 floating-point data follows the IEEE 754 single-precision floating-point format and occupies 4 Bytes. If the decoded value is abnormal, confirm whether the high/low bytes or word order need to be swapped.
Q10: What should be noted about Modbus CRC?
Modbus RTU uses CRC16. During transmission, Low Byte is sent first and High Byte last, meaning the low-order byte comes before the high-order byte.
Q11: What do Function Codes 03, 04, and 10 represent?
Function Code 03: Read Holding Registers
Function Code 04: Read Input Registers
Function Code 10: Write Multiple Holding Registers
- Modbus Reading Questions
-
Q5
Calibration Questions
E- Calibration Questions
Q1: Is calibration required before initial use?
In general, no. The device has been calibrated before shipment. During initial use, measurement can begin once the device is properly installed according to the manual.
Q2: When is recalibration required?
Recalibration is recommended when the on-site water level differs significantly from the system display value, after reinstallation, after installation height changes, when readings drift over time, or when accuracy needs to be verified after maintenance.
Q3: What calibration items are available?
Calibration items include temperature calibration, NH4-N standard calibration, K+ calibration, K+ compensation calibration, on-site calibration, background value subtraction, and pH calibration.
Q4: What should be done before calibration?
The analyzer must be thoroughly cleaned before calibration. It is recommended to remove the electrode protective cap, rinse the surface with tap water, clean with a soft brush and neutral detergent if necessary, dry all parts of the body as much as possible, and shake the electrodes dry.
Q5: What tools are required for calibration?
The following tools are required:
Calibrated standard thermometer
Calibration cup
Magnetic stir bar and magnetic stirrer
Ultrapure water
Q6: What standard solutions are required for calibration?
The following can be used:
SL23-1 ammonia nitrogen NH4-N / NH4+ and pH standard solution, 50 mg/L, 500 ml
SL23-2 ammonia nitrogen NH4-N / NH4+ and pH standard solution, 500 mg/L, 500 ml
SL23-3 four-in-one low-concentration calibration solution, 500 ml
SL23-4 four-in-one high-concentration calibration solution, 500 ml
Q7: How is temperature calibration performed?
Pour purified water or tap water into the calibration cup, then place the analyzer and standard thermometer into the cup. Wait at least 2 minutes and confirm that the temperature is stable. Then write “real-time temperature value − thermometer reading” as the temperature correction value.
Q8: Which address should the temperature calibration value be written to?
The temperature correction value can be written to PLC address 28680, HEX address 0x7007, Function Code 10, with the data type set to floating point.
Q9: Which concentrations are used for NH4-N standard calibration?
Standard calibration uses two concentrations: 50 mg/L and 500 mg/L.
Q10: How long should I wait during NH4-N standard calibration?
After stirring starts, wait approximately 15 minutes until the real-time NH4+mV value becomes stable before performing calibration.
Q11: Which address should the first-point NH4-N calibration value be written to?
The first-point NH4+mV calibration value can be written to PLC address 580, HEX address 0x0243.
Q12: Which address should the second-point NH4-N calibration value be written to?
The second-point NH4+mV calibration value can be written to PLC address 581, HEX address 0x0244.
Q13: How deep should the electrode be immersed during NH4-N calibration?
The front end of the electrode must be immersed at least 3 cm below the standard solution surface and must not touch the bottom of the calibration cup.
Q14: Which concentrations are used for K+ calibration?
K+ calibration also uses two concentrations, 50 mg/L and 500 mg/L, for first-point and second-point calibration.
Q15: Which address should the first-point K+mV calibration value be written to?
The first-point K+mV calibration value can be written to PLC address 585, HEX address 0x0248.
Q16: Which address should the second-point K+mV calibration value be written to?
The second-point K+mV calibration value can be written to PLC address 586, HEX address 0x0249.
Q17: Can K+ compensation calibration be performed by the user?
No. The manual states that the potassium ion compensation electrode has already completed compensation calibration before shipment, and customers are strongly advised not to perform this calibration themselves.
Q18: What mixed calibration solution is used for K+ compensation calibration?
K+ compensation calibration uses a mixed calibration solution of 10 ppm NH4+ and 1000 ppm K+. However, this step is not recommended for customers to perform by themselves.
Q19: Which address should the on-site calibration value be written to?
The on-site calibration value can be written to PLC address 13058, HEX address 0x3301.
Q20: When should the background value subtraction function be used?
If the concentration trend of on-site samples has not been confirmed through laboratory testing, the background value subtraction function is not recommended.
Q21: Which standard solution is used for pH calibration?
The manual example uses pH 7 standard solution for calibration. pH 4 and pH 10 calibration can also be performed as needed.
Q22: Which address should the pH slope calibration value be written to?
The pH slope calibration value can be written to PLC address 1142, HEX address 0x0475
Q23: Which address should be used for pH potential calibration?
For pH potential calibration, input 1 at PLC address 1138, HEX address 0x0471, to write the current potential value.
Q24: What range should the mV value fall within during pH calibration?
The reference ranges are as follows:
pH 7: approximately 0 ± 25 mV
pH 4: approximately 175 ± 20 mV
pH 10: approximately -175 ± 25 mV
The potential slope should be between -56 mV/pH and -61 mV/pH.
Q25: Why should large vortices and dense bubbles be avoided during calibration?
Large vortices and dense bubbles may interfere with electrode readings, causing unstable mV or concentration readings and affecting calibration results.
- Calibration Questions
-
Q7
Abnormal Readings and Troubleshooting
F- Abnormal Readings and Troubleshooting
Q1: What should I do if readings keep fluctuating?
First clean the sensing area, confirm whether the installation depth is sufficient, whether the device is firmly secured, whether bubbles, bottom sludge, or contaminants are interfering on site, and whether the power supply and RS485 communication are normal.
Q2: What should I do if the reading remains at 0?
Confirm whether the sample actually contains the target concentration, then check whether the electrode is clean, whether it is immersed deeply enough, and whether calibration is correct. If there is still no response when testing with a standard solution, contact AQUAS.
Q3: What should I do if the reading remains at the maximum range?
Possible causes include sample concentration exceeding the measurement range, incorrect calibration, electrode contamination, or interference in the sensing area. It is recommended to clean the analyzer first and reconfirm calibration.
Q4: What should I do if readings are clearly too high or too low?
Check whether standard calibration and on-site calibration are correct, whether the electrode is dirty or damaged, whether the sample is homogeneous, and whether bubbles or bottom sludge are causing interference. Recalibrate if necessary.
Q5: What should I do if the response time becomes slower?
This may be caused by electrode aging, contamination in the sensing area, unstable sample conditions, or degraded electrode performance. It is recommended to clean the electrode first and confirm the standard solution reading. Replace the electrode if necessary.
Q6: What should I do if readings from the same sample are inconsistent?
Confirm whether the sample is homogeneous, whether stratification or sediment is present, whether stirring is stable, whether bubbles are attached, and whether the calibration procedure is correct.
Q7: What should I do if the reading does not change?
Restore factory settings first and repeat the calibration procedure. If the reading still does not change after testing with different standard solutions, contact AQUAS technical support.
Q8: If on-site readings differ from laboratory measurements, does that mean the device is faulty?
Not necessarily. Real-time monitoring instruments are directly exposed to the on-site environment, while laboratory samples usually undergo sampling, storage, pretreatment, and standardized procedures. Different test conditions may lead to differences in values. Consistency can be improved through on-site calibration.
Q9: Do cracks or wear on the electrode affect readings?
Yes. If the electrode is cracked, damaged, or worn, readings may become unstable, response may slow down, or calibration may fail. Electrode replacement should be considered.
Q10: When should AQUAS technical support be contacted?
If the device still cannot read values or shows no response after checking power supply, wiring, Modbus ID, communication settings, standard solution testing, calibration, cleaning, and electrode inspection, it is recommended to contact AQUAS technical support and provide on-site photos, wiring photos, Modbus settings, standard solution test records, and reading screenshots.
- Abnormal Readings and Troubleshooting
-
Q8
Maintenance and Consumables Questions
F- Maintenance and Consumables Questions
Q1: Is regular maintenance required?
Yes. To keep the analyzer in optimal operating condition, regular maintenance and cleaning are required, especially in environments where oil, sediment, or other contaminants easily adhere to the device.
Q2: How often should routine maintenance be performed?
Routine maintenance is recommended at least once per week. The actual frequency can be adjusted according to on-site water quality, contamination level, and monitoring environment.
Q3: What should be included in routine maintenance?
Routine maintenance should include:
Cleaning the analyzer surface and sensing area
Checking the battery level, if using a battery-powered model
Checking consumable status
Checking whether cables and mounting fixtures are secure
Confirming whether data transmission is stable
Q4: How often should periodic maintenance be performed?
Periodic maintenance is recommended at least once per month.
Q5: What is included in periodic maintenance?
Periodic maintenance includes trend calibration and standard calibration. Long-term data trends can be used to determine whether instrument adjustment is required.
Q6: What consumables are used?
The main consumables include standard solutions, purified water, and electrodes.
Q7: When should the electrode be replaced?
Electrode replacement is recommended when the electrode response becomes slower, readings are unstable, the electrode slope decreases significantly, or physical damage such as cracks or breakage is observed.
Q8: How should K+ or Ref electrodes be stored when not used for a long time?
When the electrodes are not used for an extended period, clean the sensor thoroughly and store the K+ and Ref electrodes immersed in 3M KCl storage solution.
Q9: Is calibration required after replacing an electrode?
Yes, it is recommended. After replacing an electrode, confirm that all connections are correctly installed and perform calibration again to ensure accurate measurement results.
- Maintenance and Consumables Questions