MAX Multiparameter Water Quality Sonde

Frequently Asked Questions

  • A
    1. General Product Questions
      Q1: About product?
      MAX is an AQUAS multi-parameter water quality analyzer. Depending on the selected sensors, it can simultaneously monitor multiple water quality parameters, including pH, ORP, conductivity, dissolved oxygen, turbidity, suspended solids, ammonium nitrogen, nitrate nitrogen, potassium ions, chloride ions, water level, temperature, chlorophyll-a, blue-green algae, CDOM, oil in water, BOD, color, and more.

      Q2: What application sites is it suitable for?
      MAX is suitable for water quality monitoring in groundwater, sewers, surface water, detention ponds, rainwater, storage tanks, reservoirs, rivers, lakes, wetlands, coastal estuaries, aquaculture sites, and other environments.

      Q3: What is the waterproof rating?
      MAX features a rugged IP68-rated submersible enclosure, making it suitable for long-term installation in wet, rainy, or underwater monitoring environments.

      Q4: Does it support remote monitoring?
      Yes. MAX can be integrated with AQUAS data loggers, AQWEB, AQOPC, customer platforms, or PLC systems through RS485 serial communication.

      Q5: Which communication protocol does it use?
      MAX uses RS485 communication and supports the standard Modbus RTU protocol.

      Q6: Can it be connected directly to a PLC?
      Yes. MAX can be connected to a PLC via RS485. However, the Modbus ID, baud rate, parity, data bits, and stop bit must be consistent with the PLC-side settings.
  • A
    1. Installation Questions
      Q1: What is the operating voltage?
      The operating voltage is 5–12VDC. For some configurations, 12VDC may be recommended to ensure optimal performance.

      Q2: What should be noted during installation?
      During installation, avoid pulling or straining the cable. The cable should be long enough and properly secured. It is recommended to install the device at a location with stable water flow to obtain better measurement results.

      Q3: What should be removed before installing the electrode?
      If the electrode has a protective solution cap or protective sleeve, it must be removed before installation to avoid affecting measurement performance.

      Q4: How deep should the sensor be submerged?
      When testing Type A sensors, the sensing surface must be submerged at least 10 cm below the liquid surface.

      Q5: How long does it take to stabilize after installation?
      Type A sensors typically require at least 20 minutes to stabilize after installation. The actual stabilization time may vary depending on sample characteristics and installation conditions.

      Q6: Can it be installed horizontally?
      Installation too close to a horizontal position is not recommended. When using Type A sensors, MAX should be installed perpendicular to the horizontal plane. A certain degree of tilt is acceptable, but an angle too close to horizontal may cause abnormal readings.

      Q7: Can the protective cap be used as a calibration cup?
      Yes. After removing the protective cap base and installing it on the other end of the protection tube, it can be used as a calibration cup. Before reinstalling the device, confirm that the bottom cap has been moved back to the opening side.
  • B
    1. Connection and AQCFG Questions
      Q1: How does MAX connect to a computer?
      MAX can be connected directly to a computer using CAB06 Cable. Values can then be read and settings configured through AQCFG software.

      Q2: What should I do if CAB06 cannot power or drive MAX properly?
      If CAB06 cannot properly power or drive MAX, it is recommended to use external 12VDC power wiring and confirm that the RS485 and power wiring are correct.

      Q3: What are the default communication settings?
      The default MAX communication settings are as follows:
      Modbus ID: 1
      Baud rate: 19200
      Parity: None
      Data Bits: 8
      Stop Bit: 1

      Q4: Which baud rates are supported?
      MAX supports 9600 and 19200 bps.

      Q5: What should I do if AQCFG cannot find the device?
      Please check the following in order:
      1. Whether CAB06 is connected correctly
      2. Whether the correct COM Port is selected
      3. Whether the Modbus ID is within the search range
      4. Whether the baud rate, parity, data bits, and stop bit are correct
      5. Whether MAX has sufficient power supply
      6. Whether there is an ID conflict

      Q6: What do the AQCFG communication status colors indicate?
      Green indicates connected, and red indicates disconnected.

      Q7: How should the ID be set when MAX is connected to an AQUAS data logger?
      When MAX is connected to an AQUAS data logger, the ID should be set to 1.

      Q8: Which device IDs are occupied by MAX1, MAX2, and MAX3?
      When integrated with an AQUAS host, MAX1 occupies IDs 1–4, MAX2 occupies IDs 1–8, and MAX3 occupies IDs 1–12. If other analyzers are used at the same time, their IDs must avoid the above ranges.
  • C
    1. Calibration Questions
      Q1: Is calibration required before initial use?
      In general, no. MAX is calibrated before shipment. During initial use, measurement can begin once the device is properly installed according to the manual and the electrode protective sleeve is removed.

      Q2: When is recalibration required?
      Recalibration is recommended when readings differ significantly from on-site or laboratory measurements, after electrode replacement, after long-term use, after maintenance, or when trend data shows drift.

      Q3: What should be done before calibration?
      MAX should be thoroughly cleaned before calibration, especially the electrodes and sensors. It is recommended to rinse with tap water first, use a soft brush and neutral detergent if necessary, rinse again with purified water, and then dry all parts of the device as much as possible.

      Q4: Which standard solutions are required for pH calibration?
      pH calibration requires pH 7, pH 4, and pH 10 standard solutions.

      Q5: How deep should the pH electrode be immersed during calibration?
      The front end of the pH electrode must be immersed at least 3 cm below the standard solution surface and must not touch the bottom of the calibration cup.

      Q6: What standard solution is required for ORP calibration?
      ORP calibration uses a 220 mV ORP standard solution.

      Q7: What are the stability conditions for ORP calibration?
      The ORP reading variation must be less than 1 mV per minute, and the potential during calibration should be approximately 220 ± 20 mV.

      Q8: What standard solution is required for EC conductivity calibration?
      EC calibration requires a 1.413 mS/cm standard solution.

      Q9: Does water-level calibration require a standard solution?
      No. Water-level calibration requires a container at least 1.5 m deep and a tape measure or another tool capable of measuring water depth.

      Q10: Can ammonium nitrogen and potassium ions be calibrated together?
      Yes. Since a mixed calibration solution is used, ammonium nitrogen NH4-N and potassium ion K+ can be calibrated simultaneously. It is also possible to calibrate only one of them.

      Q11: What should be noted during ammonium nitrogen and potassium ion calibration?
      Ion-selective electrodes are sensitive to temperature and environmental conditions, so calibration is recommended in a stable environment. During calibration, the electrode must be immersed at least 10 cm below the liquid surface, and air bubbles at the bottom of the electrode must be removed.

      Q12: What should be noted when calibrating fluorescence analyzers?
      Before calibration, thoroughly clean the inside of the MAX protective cap, the MAX body surface, the gaps around each sensor, and the calibration cup to prevent residue from affecting calibration results.

      Q13: Does drift in optical analyzer readings always require recalibration?
      Not necessarily. Many optical measurement drifts are caused by contamination on the optical window. It is recommended to clean the optical window first before determining whether recalibration is required.

      Q14: What is trend calibration?
      Trend calibration compares MAX real-time on-site readings with laboratory reference values over time. Linear regression is then used to adjust the slope and intercept so that instrument readings more closely reflect the true values.

      Q15: Which parameters are more suitable for trend calibration?
      Ion-selective electrodes ISE and optical analyzers are more susceptible to external interference, such as interfering substances or changes in sample optical properties. Therefore, trend calibration is more strongly recommended for these parameters.

      Q16: Do pH, EC, and dissolved oxygen require trend calibration?
      In general, pH, EC, and dissolved oxygen have relatively high measurement consistency and stability, so trend calibration is usually not required.
  • D
    1. Abnormal Readings and Troubleshooting
      Q1: What should I do if readings are unstable?
      First check whether the sensor is clean, whether air bubbles are attached, whether the electrode is fully immersed in water, whether the installation angle is too close to horizontal, whether the water flow is stable, and whether the power supply and communication are normal.

      Q2: If readings differ from laboratory data, does that mean the device is faulty?
      Not necessarily. Real-time monitoring instruments measure directly in the field, while laboratory samples may undergo pretreatment and standardized testing. Differences between the two are common. Consistency can be improved through trend monitoring and trend calibration.

      Q3: What should I do if the PLC cannot connect to MAX or cannot read values?
      Please check the following in order:
      1. Connect to a PC using CAB06 and confirm whether AQCFG can read the device
      2. Confirm whether the communication settings of MAX and the PLC are consistent
      3. Check whether there is a Modbus ID conflict
      4. Check whether the PLC Modbus address is correct
      5. Confirm whether the power supply is sufficient
      6. Check whether the RS485 wiring is correct


      Q4: Can MAX be powered by a computer’s 5V supply?
      In most cases, yes. However, because MAX may be configured with various optional sensors, an external power supply is recommended to ensure proper device operation, especially when multiple sensors are installed.

      Q5: What should I do if absorbance analyzer readings are abnormal or unchanged?
      Read PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type set to floating point.

      Q6: What should I do if optical analyzer readings are abnormal?
      First check whether the optical window is dirty, whether air bubbles are attached, whether there are scratches, or whether moisture has entered. Then confirm whether recalibration is required.

      Q7: What should I do if absorbance analyzer readings are abnormal or unchanged?
      Read PLC address 10, HEX address 0x0009, Length 6 Words, Function Code 03, with the data type set to floating point.

      Q8: What should I do if readings are abnormal after replacing a sensor?
      Confirm whether the sensor has been reinstalled in its original designated port, whether the locking points are correct, and whether the connector is dry and securely installed. Recalibration is recommended after replacement.

      Q9: Can I change the port order myself when replacing electrodes?
      No. The MAX body has numbered ports, and electrodes and analyzers must be installed according to the factory-defined order. The order must not be changed. If the original position is unknown, contact AQUAS for confirmation.
  • F
    1. Maintenance Questions
      Q1: Is regular maintenance required?
      Yes. Although MAX is equipped with an automatic cleaning brush system to reduce routine maintenance needs, regular manual cleaning and inspection are still required.

      Q2: How often should routine maintenance be performed?
      The maintenance frequency should be adjusted according to site conditions. If oil, sludge, algae, or other contaminants are present on site, maintenance frequency should be increased. Periodic maintenance is recommended at least once per month.

      Q3: What should be included in routine maintenance?
      Clean the MAX body and sensor surfaces
      Check whether the cleaning brush is functioning properly
      Check whether electrodes or optical windows are dirty
      Check cables and mounting fixtures
      Check whether readings show abnormal drift

      Q4: What is included in periodic maintenance?
      Periodic maintenance mainly includes trend calibration and standard calibration to verify instrument performance and measurement accuracy.

      Q5: How should pH, ORP, and Ref electrodes be stored when not in use?
      When pH, ORP, and Ref electrodes are temporarily not in use, they must be stored in 3M KCl solution. Improper storage may damage the electrodes.

      Q6: How should sensors without special storage instructions be stored?
      Sensors without special storage requirements can be stored at room temperature in an environment with moderate humidity.

      Q7: Is calibration required after replacing an electrode?
      Yes, it is recommended. After replacing an electrode, confirm that all connections are correctly installed and perform calibration again to obtain optimal measurement results.

      Q8: Which electrodes are consumables that require regular replacement?
      Consumables that require regular inspection or replacement include pH, ORP, Ref, ammonium nitrogen, nitrate nitrogen, potassium ion, chloride ion, EC, and optical dissolved oxygen caps.

      Q9: When should an electrode be replaced?
      Replacement is recommended when the electrode response becomes slow, readings are unstable, the slope is too low, or any physical damage is observed.

      Q10: What is the recommended electrode replacement interval?
      Reference intervals are as follows:
      pH: approximately 12 months
      ORP: approximately 12 months
      Ref: approximately 12 months
      Ammonium nitrogen: approximately 18 months
      Nitrate nitrogen: approximately 18 months
      Potassium ion: approximately 18 months
      Chloride ion: approximately 18 months
      The actual replacement timing should still be determined based on site conditions and maintenance results.